Comparison of Monte Carlo calculations around a Fletcher Suit Delclos ovoid with radiochromic film and normoxic polymer gel dosimetry.

The Fletcher Suit Delclos (FSD) ovoids employed in intracavitary brachytherapy (ICB) for cervical cancer contain shields to reduce dose to the bladder and rectum. Many treatment planning systems (TPS) do not include the shields and other ovoid structures in the dose calculation. Instead, TPSs calculate dose by summing the dose contributions from the individual sources and ignoring ovoid structures such as the shields. The goal of this work was to calculate the dose distribution with Monte Carlo around a Selectron FSD ovoid and compare these calculations with radiochromic film (RCF) and normoxic polymer gel dosimetry. Monte Carlo calculations were performed with MCNPX 2.5.c for a single Selectron FSD ovoid with and without shields. RCF measurements were performed in a plane parallel to and displaced laterally 1.25 cm from the long axis of the ovoid. MAGIC gel measurements were performed in a polymethylmethacrylate phantom. RCF and MAGIC gel were irradiated with four 33μGym2h-1 Cs-137 pellets for a period of 24 h. Results indicated that MCNPX calculated dose to within ±2% or 2 mm for 98% of points compared with RCF measurements and to within ±3% or 3 mm for 98% of points compared with MAGIC gel measurements. It is concluded that MCNPX 2.5.c can calculate dose accurately in the presence of the ovoid shields, that RCF and MAGIC gel can demonstrate the effect of ovoid shields on the dose distribution and the ovoid shields reduce the dose by as much as 50%.

[1]  J F Williamson,et al.  On the validity of the superposition principle in dose calculations for intracavitary implants with shielded vaginal colpostats. , 2001, Medical physics.

[2]  R. Lane,et al.  Contrast effects on dosimetry of a partial breast irradiation system. , 2004, Medical physics.

[3]  J. Galvin,et al.  Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. American Association of Physicists in Medicine. , 1998, Medical physics.

[4]  J. Williamson,et al.  Template-guided interstitial implants: Cs-137 reusable sources as a substitute for Ir-192. , 1987, Radiology.

[5]  L. Anderson,et al.  Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43 , 1995 .

[6]  D Baltas,et al.  The effect of patient inhomogeneities in oesophageal 192Ir HDR brachytherapy: a Monte Carlo and analytical dosimetry study. , 2004, Physics in medicine and biology.

[7]  J C Gore,et al.  Radiation therapy dosimetry using magnetic resonance imaging of polymer gels. , 1996, Medical physics.

[8]  J. Gore,et al.  NMR relaxation enhancement in gels polymerized and cross-linked by ionizing radiation: a new approach to 3D dosimetry by MRI. , 1993, Magnetic resonance imaging.

[9]  L. Anderson,et al.  Computation of radiation dose distributions for shielded cervical applicators. , 1985, International journal of radiation oncology, biology, physics.

[10]  D Baltas,et al.  A monte carlo dosimetry study of vaginal 192Ir brachytherapy applications with a shielded cylindrical applicator set. , 2004, Medical physics.

[11]  M. Lindstrom,et al.  The dose distribution surrounding 192Ir and 137Cs seed sources. , 1991, Physics in medicine and biology.

[12]  C. S. Poon,et al.  Practical T2 quantitation for clinical applications , 1992, Journal of magnetic resonance imaging : JMRI.

[13]  S. C. Prasad,et al.  Clinical electron-beam dosimetry: report of AAPM Radiation Therapy Committee Task Group No. 25. , 1991, Medical physics.

[14]  L. Anderson,et al.  Measurements of radiation dose distributions for shielded cervical applicators. , 1985, International Journal of Radiation Oncology, Biology, Physics.

[15]  J. Gore,et al.  Polymer gels for magnetic resonance imaging of radiation dose distributions at normal room atmosphere. , 2001, Physics in medicine and biology.

[16]  F. Mourtada,et al.  A segmented 32P source Monte Carlo model to derive AAPM TG-60 dosimetric parameters used for intravascular brachytherapy. , 2004, Medical physics.

[17]  C Baldock,et al.  A basic study of some normoxic polymer gel dosimeters. , 2002, Physics in medicine and biology.

[18]  J F Williamson,et al.  Quantitative evaluation of radiochromic film response for two-dimensional dosimetry. , 1997, Medical physics.

[19]  H. Meertens,et al.  Screens in ovoids of a Selectron cervix applicator. , 1985, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[20]  M Bamberg,et al.  Accuracy of field alignment in radiotherapy of head and neck cancer utilizing individualized face mask immobilization: a retrospective analysis of clinical practice. , 1995, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[21]  R. Sloboda,et al.  Combined experimental and Monte Carlo verification of brachytherapy plans for vaginal applicators , 1998, Physics in medicine and biology.

[22]  J F Williamson,et al.  Dose calculations about shielded gynecological colpostats. , 1990, International journal of radiation oncology, biology, physics.