SAR and B1 field distributions in a heterogeneous human head model within a birdcage coil

Calculations of radiofrequency magnetic (B1) field and specific energy absorption rate (SAR) distributions in a sphere of tissue and a multi‐tissue human head model in a 12‐element birdcage coil are presented. The coil model is driven in linear and quadrature modes at 63, 175, 200, and 300 MHz. Plots of B, field magnitude and SAR distributions, average SAR, maximum local SAR, and measures of B1 field homogeneity and signal‐to‐noise ratio are given. SAR levels for arbitrary pulse sequences can be estimated from the calculated data. Maximum local SAR levels are lower at lower frequencies, in quadrature rather than in linear coils, and in linear fields oriented posterior‐to‐anterior rather than left‐to‐right in the head. It should be possible to perform many experiments in the head at frequencies up to 300 MHz without exceeding standard limits for local or average SAR levels.

[1]  Jianming Jin,et al.  On the SAR and field inhomogeneity of birdcage coils loaded with the human head , 1997, Magnetic resonance in medicine.

[2]  Christopher M. Collins,et al.  A Method for Accurate Calculation ofB1Fields in Three Dimensions. Effects of Shield Geometry on Field Strength and Homogeneity in the Birdcage Coil , 1997 .

[3]  H. Wen,et al.  Simulation ofB1Field Distribution and Intrinsic Signal-to-Noise in Cardiac MRI as a Function of Static Magnetic Field , 1997 .

[4]  W. Chew,et al.  Computation of electromagnetic fields for high-frequency magnetic resonance imaging applications. , 1996, Physics in medicine and biology.

[5]  R. Stollberger,et al.  Spatial distribution of high-frequency electromagnetic energy in human head during MRI: numerical results and measurements , 1996, IEEE Transactions on Biomedical Engineering.

[6]  R. Mark Henkelman,et al.  Rotating Frame RF Current Density Imaging , 1995, Magnetic resonance in medicine.

[7]  R. Henkelman,et al.  RF Current Density Imaging in Homogeneous Media , 1992, Magnetic resonance in medicine.

[8]  P A Bottomley,et al.  Homogeneous Tissue Model Estimates of RF Power Deposition in Human NMR Studies , 1992, Annals of the New York Academy of Sciences.

[9]  J W Carlson,et al.  Electromagnetic fields of surface coil in vivo NMR at high frequencies , 1991, Magnetic resonance in medicine.

[10]  O. Gandhi,et al.  Magnetic resonance imaging: calculation of rates of energy absorption by a human-torso model. , 1990, Bioelectromagnetics.

[11]  Increased rf power absorption in MR imaging due to rf coupling between body coil and surface coil , 1989, Magnetic resonance in medicine.

[12]  J. Hornak,et al.  Magnetic field mapping , 1988, Magnetic resonance in medicine.

[13]  Michael Garwood,et al.  Amplitude- and frequency-modulated pulses to achieve 90° plane rotations with inhomogeneous B1 fields , 1987 .

[14]  D. Hoult,et al.  The field dependence of NMR imaging. I. Laboratory assessment of signal‐to‐noise ratio and power deposition , 1986, Magnetic resonance in medicine.

[15]  W. Barber,et al.  Comparison of linear and circular polarization for magnetic resonance imaging , 1985 .

[16]  J. Schenck,et al.  Estimating radiofrequency power deposition in body NMR imaging , 1985, Magnetic resonance in medicine.

[17]  P A Bottomley,et al.  RF magnetic field penetration, phase shift and power dissipation in biological tissue: implications for NMR imaging. , 1978, Physics in medicine and biology.

[18]  K. Yee Numerical solution of initial boundary value problems involving maxwell's equations in isotropic media , 1966 .