Using intraoperative pelvic landmarks for acetabular component placement in total hip arthroplasty.

Dislocation after total hip arthroplasty is frequently due to acetabular malpositioning. Positioning of the acetabular component using anatomical landmarks may reduce the incidence of dislocation from improper acetabular orientation. The pelvis provides 3 bony landmarks (ilium, superior pubic ramus, and superior acetabulum), which, when used to define a plane, allows cup orientation in abduction and version. Landmarks evaluated in 24 cadaveric acetabuli allowed slightly increased abduction and anteversion of the cup, compared with native acetabuli. Six hundred seventeen primary total hip arthroplasties were performed between 1996 and 2003 using this technique. Mean cup abduction was 44.4 degrees with 13.2 degrees of anteversion. This technique allows satisfactory reproducible cup orientation based on individual pelvic morphology. Review of patient outcome data suggest high patient satisfaction and lower dislocation rate without additional equipment, time, or cost.

[1]  H. Fredin,et al.  THE PROGNOSIS AND TREATMENT OF DISLOCATED TOTAL HIP ARTHROPLASTIES WITH A 22 MM HEAD , 1997 .

[2]  G. Selvik,et al.  Position, Orientation and Component Interaction in Dislocation of the Total Hip Prosthesis , 1988, Acta radiologica.

[3]  Karl-Heinz Widmer,et al.  A simplified method to determine acetabular cup anteversion from plain radiographs. , 2004, The Journal of arthroplasty.

[4]  D. Murray The definition and measurement of acetabular orientation. , 1993, The Journal of bone and joint surgery. British volume.

[5]  J. Olsen,et al.  Causes and Prevention , 1991, Scandinavian journal of social medicine.

[6]  J P Conaty,et al.  Classification and treatment of dislocations of total hip arthroplasty. , 1983, Clinical orthopaedics and related research.

[7]  B. Mckibbin,et al.  Anatomical factors in the stability of the hip joint in the newborn. , 1970, The Journal of bone and joint surgery. British volume.

[8]  B. Jaramaz,et al.  Computer Assisted Measurement of Cup Placement in Total Hip Replacement , 1998, Clinical orthopaedics and related research.

[9]  L.-j. Yuan,et al.  Dislocation after total hip arthroplasty , 1999, Archives of Orthopaedic and Trauma Surgery.

[10]  J. Charnley,et al.  Postoperative dislocation after Charnley low-friction arthroplasty. , 1978, Clinical orthopaedics and related research.

[11]  B Jaramaz,et al.  Image Guided Navigation System to Measure Intraoperatively Acetabular Implant Alignment , 1998, Clinical orthopaedics and related research.

[12]  J. Lewis,et al.  Dislocations after total hip-replacement arthroplasties. , 1978, The Journal of bone and joint surgery. American volume.

[13]  R S Turner,et al.  Postoperative total hip prosthetic femoral head dislocations. Incidence, etiologic factors, and management. , 1994, Clinical orthopaedics and related research.

[14]  G. Selvik,et al.  Space Orientation of Total Hip Prosthesis , 1986, Acta radiologica: diagnosis.

[15]  B F Morrey,et al.  Operative correction of an unstable total hip arthroplasty. , 1992, The Journal of bone and joint surgery. American volume.

[16]  D. McCollum,et al.  Dislocation after total hip arthroplasty. Causes and prevention. , 1990, Clinical orthopaedics and related research.

[17]  B F Morrey,et al.  Dislocations after total hip arthroplasty. , 1982, The Journal of bone and joint surgery. American volume.