Mechanical resistance of a new biomaterial, ostrich pericardium, and a new method of joining tissues combining suturing and a biological adhesive.

We studied the mechanical behavior in response to tensile stress of samples of ostrich pericardium bonded with a cyanoacrylate glue or sewn with a rectangular, overlapping suture that was subsequently sealed with the same bioadhesive. Seventy-two trials were performed in three series of 24 samples each: series AG, glued with an overlap of 1 cm2; series ASG, sewn with a rectangular, overlapping suture and sealed; and series AC, control samples that were left intact. The mean stress at rupture in series AG (glued) was 0.1 MPa, much lower than the working stress of a human valve leaflet, which is approximately 0.25 MPa. In the control series, this stress was 26.28 MPa. At rupture in series ASG (sutured/glued), the suture material was being subjected to a stress of 64.91 MPa, thus confirming the existence of an interaction between the suture and the shear stress exerted by the suture on the samples of pericardium. In series ASG, the mean value for the resistance to rupture when measured in machine kg was 8.83 kg, lower than but similar to that recorded in the control series AC (10.26 kg). The percentages of reversible deformation, or elongation, once the samples were torn were similar in series AC (19.15%) and ASG (21.93%). This phenomenon can only be explained by the damage to the collagen fibers in the area around the rupture, while other more distant regions work at a lower load within the elastic limit. We conclude that cyanocrylate adhesives alone are not suitable as bonding materials in cardiac bioprostheses. The results with the rectangular, overlapping suture, when subsequently sealed with an adhesive, can be considered good because, although this approach does not impede shear stress, it does maintain an excellent degree of resistance to rupture of the samples thus joined. We stress the need to take into account the concentration of the load in the design of bioprostheses.

[1]  Y. Ikada,et al.  Bioadhesion of gelatin films crosslinked with glutaraldehyde. , 1999, Journal of biomedical materials research.

[2]  W D White,et al.  Determinants of reoperation after 960 valve replacements with Carpentier-Edwards prostheses. , 1994, The Journal of thoracic and cardiovascular surgery.

[3]  Á. Cordon,et al.  Analysis of shearing stress in the limited durability of bovine pericardium used as a biomaterial , 1998, Journal of materials science. Materials in medicine.

[4]  D. Quintanar-Guerrero,et al.  In vitro evaluation of the bioadhesive properties of hydrophobic polybasic gels containing N,N-dimethylaminoethyl methacrylate-co-methyl methacrylate. , 2001, Biomaterials.

[5]  A. Carrera,et al.  Mechanical behavior of chemically treated ostrich pericardium subjected to uniaxial tensile testing: influence of the suture. , 2002, Journal of biomedical materials research.

[6]  J. M. Bernal,et al.  Durability of the Carpentier-Edwards porcine bioprosthesis: role of age and valve position. , 1995, The Annals of thoracic surgery.

[7]  J. G. García Páez,et al.  Resistance and elasticity of the suture threads employed in cardiac bioprostheses. , 1994, Biomaterials.

[8]  J. G. García Páez,et al.  Effect of the suture on the durability of bovine pericardium used in cardiac bioprostheses. , 1994, Biomaterials.

[9]  J. Monro,et al.  Biological valves beyond fifteen years: the Wessex experience. , 1995, The Annals of thoracic surgery.

[10]  Richard A. Lange,et al.  Prosthetic heart valves. , 1996, The New England journal of medicine.

[11]  W. Jamieson,et al.  Tricuspid valve replacement: porcine bioprostheses and mechanical prostheses. , 1995, The Annals of thoracic surgery.

[12]  Á. Cordon,et al.  Comparison of elasticities of components of a cardiac bioprosthesis leaflet. , 1996, Journal of biomedical materials research.

[13]  W. Jamieson,et al.  Reoperation in biological and mechanical valve populations: fate of the reoperative patient. , 1995, The Annals of thoracic surgery.

[14]  R. Goode,et al.  Efficacy of octyl-2-cyanoacrylate tissue glue in blepharoplasty. A prospective controlled study of wound-healing characteristics. , 1999, Archives of facial plastic surgery.

[15]  J. G. García Páez,et al.  Is cutting stress responsible for the limited durability of heart valve bioprostheses? , 1990, The Journal of thoracic and cardiovascular surgery.

[16]  Gouri Dhatt,et al.  Stress analysis of the human aortic valve , 1973 .

[17]  J. G. García Páez,et al.  Selection and interaction of biomaterials used in the construction of cardiac bioprostheses. , 1998, Journal of biomedical materials research.

[18]  G. Christie,et al.  Tears in bioprosthetic heart valve leaflets without calcific degeneration. , 1996, The Journal of heart valve disease.

[19]  A. Carrera,et al.  Ostrich pericardium, a biomaterial for the construction of valve leaflets for cardiac bioprostheses: mechanical behaviour, selection and interaction with suture materials. , 2001, Biomaterials.

[20]  M. Sacks,et al.  Collagen fiber architecture of bovine pericardium. , 1994, ASAIO journal.

[21]  W. White,et al.  In-hospital and long-term outcome after porcine tricuspid valve replacement. , 1995, The Journal of thoracic and cardiovascular surgery.

[22]  W. Jamieson,et al.  Porcine bioprostheses in the elderly: clinical performance by age groups and valve positions. , 1995, The Annals of thoracic surgery.

[23]  J Fisher,et al.  Anisotropic elasticity and strength of glutaraldehyde fixed bovine pericardium for use in pericardial bioprosthetic valves. , 1994, Journal of biomedical materials research.

[24]  P. Tesar,et al.  Late complications of tissue glues in aortic surgery. , 2000, The Annals of thoracic surgery.

[25]  W. Edwards,et al.  Temporal changes in the surgical pathology of prosthetic aortic valves: a study of 157 cases spanning 26 years (1970-1995). , 1998, Cardiovascular pathology : the official journal of the Society for Cardiovascular Pathology.