Hemodynamics of the mitral valve under edge-to-edge repair: an in vitro steady flow study.

Edge-to-edge repair (ETER) is a mitral valve repair technique that restores valvular competence by suturing together the free edges of two leaflets. This repair technique alters mitral valve inflow and thus left ventricle hemodynamics during diastole. Our objective was to investigate fluid mechanics immediately downstream of the mitral valve under ETER during diastole. Fresh porcine mitral valves of the annulus size M32 with chordae removed were installed into a steady flow loop simulating a peak diastolic inflow through the mitral valve. Digital particle image velocimetry was used to measure the velocity field immediately downstream of the mitral valve under normal and ETER conditions. First, to study the suture length effect, suture was applied in the central position of the leaflet edge with suture lengths of 3 mm, 6 mm, and 9 mm, respectively. Then, 3 mm suture was set in the central, lateral, and commissural positions of the leaflet edge to study the suture position effect. Flow rate was 15 l/min. Velocity, Reynolds shear stress (RSS), and effective orifice area were assessed. A total of five mitral valves were tested. The normal mitral valve without the ETER had one jet downstream of the valve, but the mitral valve with the central or lateral sutures under the ETER had two jets downstream of the valve with a recirculation region downstream of the suture. The maximum velocity, the maximum RSS in the jets, the pressure drop across the mitral valve, and the jet deflection angle increased with the increase in suture length in the central position. When the suture position effect was investigated with the 3 mm suture, the maximum velocity, the maximum RSS, and the pressure drop across the valve in the central suture position were greater than those of the lateral and the commissural suture positions. The lateral suture demonstrated major and minor jets with the greater maximum velocity and maximum RSS in the major jet. When the suture was in the commissural position, the flow field downstream of the mitral valve was similar to that of the normal mitral valve without the ETER. The effective orifice area was smallest when the suture was applied in the central position as compared with other suture positions. Both suture length and position have an important impact on fluid mechanics downstream of the mitral valve under the ETER in terms of flow pattern, maximum velocity, and RSS distribution. The altered hemodynamics of the mitral valve and thus of the left ventricle by the ETER may change mitral valve and left ventricle function.

[1]  S. Jones A relationship between reynolds stresses and viscous dissipation: Implications to red cell damage , 2006, Annals of Biomedical Engineering.

[2]  O. Alfieri,et al.  The edge-to-edge technique: a simplified method to correct mitral insufficiency. , 1998, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[3]  S. Sutera,et al.  Flow-induced trauma to blood cells. , 1977, Circulation research.

[4]  Ajit P Yoganathan,et al.  The material properties of the native porcine mitral valve chordae tendineae: an in vitro investigation. , 2006, Journal of biomechanics.

[5]  L. Cohn,et al.  Midterm results of the edge-to-edge technique for complex mitral valve repair. , 2006, The Annals of thoracic surgery.

[6]  O. Alfieri,et al.  The double-orifice technique in mitral valve repair: a simple solution for complex problems. , 2001, The Journal of thoracic and cardiovascular surgery.

[7]  N. Hwang,et al.  An Experimental Study of Steady Flow Patterns of a New Trileaflet Mechanical Aortic Valve , 2004, ASAIO journal.

[8]  O. Alfieri,et al.  The hemodynamic effects of double-orifice valve repair for mitral regurgitation: a 3D computational model. , 1999, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[9]  P. Oshkai,et al.  Quantitative Imaging of Flow through a Mechanical Heart Valve , 2005 .

[10]  O. Alfieri,et al.  The double-orifice technique as a standardized approach to treat mitral regurgitation due to severe myxomatous disease: surgical technique. , 2000, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[11]  D. Baudendistel Heart Disease A Textbook of Cardiovascular Medicine , 1993 .

[12]  Y. Fung,et al.  Mechanics of the Circulation , 2011, Developments in Cardiovascular Medicine.

[13]  E. Braunwald Heart Disease: A Textbook of Cardiovascular Medicine , 1992, Annals of Internal Medicine.

[14]  R GORLIN,et al.  Hydraulic formula for calculation of the area of the stenotic mitral valve, other cardiac valves, and central circulatory shunts. I. , 1951, American heart journal.

[15]  G Donzella,et al.  Structural effects of an innovative surgical technique to repair heart valve defects. , 2005, Journal of biomechanics.

[16]  R Fumero,et al.  A computational study of the hemodynamics after "edge-to-edge" mitral valve repair. , 2001, Journal of biomechanical engineering.

[17]  J. Yeo,et al.  Shear Stress Investigation Across Mechanical Heart Valve , 2007, ASAIO journal.

[18]  E L Yellin,et al.  Mechanisms of mitral valve motion during diastole. , 1981, The American journal of physiology.

[19]  Alberto Redaelli,et al.  3-D computational analysis of the stress distribution on the leaflets after edge-to-edge repair of mitral regurgitation. , 2002, The Journal of heart valve disease.