Extraction of a migrated coil from the Enterprise stent strut using a Solitaire AB stent

A 56-year-old woman was admitted to stent-assisted coiling for a 2-mm A1 aneurysm of the left anterior cerebral artery and a left 3-mm internal carotid artery aneurysm. While coiling the A1 aneurysm, the first 2 mm × 20 mm coil migrated through the 4.5 mm × 37 mm Enterprise stent struts, lodging at the distal anterior cerebral artery. A 4 mm × 15 mm Solitaire AB stent was used successfully in this case to remove the displaced coil. The A1 aneurysm was re-treated with a 2 mm × 40 mm coil after placement of the Enterprise stent, and the ophthalmic ICA aneurysm was also coiled through the stent struts. The patient was neurologically intact after treatment.

[1]  G. Schroth,et al.  Technical Feasibility and Application of Mechanical Thrombectomy with the Solitaire FR Revascularization Device in Acute Basilar Artery Occlusion , 2013, American Journal of Neuroradiology.

[2]  K. Kamada,et al.  Ruptured aneurysm with delayed distal coil migration requiring surgical treatment. Case report. , 2012, Neurologia medico-chirurgica.

[3]  M. Martínez‐Galdámez,et al.  Y-Stent-Assisted Coil Embolization of Anterior Circulation Aneurysms Using Two Solitaire AB Devices: A Single Center Experience , 2012, Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences.

[4]  Taylor J. Abel,et al.  Treatment of Endovascular Coil and Stent Migration Using the Merci Retriever: Report of Three Cases , 2012, Case reports in medicine.

[5]  X. Lv,et al.  Stent-assisted coil embolization of intracranial aneurysms using Solitaire stent. , 2012, Neurology India.

[6]  Seok-Mann Yoon,et al.  Waffle-Cone Technique Using Solitaire AB Stent , 2012, Journal of Korean Neurosurgical Society.

[7]  W. Jaschke,et al.  Endovascular Popliteal Thrombectomy Using the Self-Expanding, Retrievable Solitaire Recanalization Device , 2012, Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists.

[8]  H. Kim,et al.  Safety and Efficacy of Mechanical Thrombectomy with Solitaire Stent Retrieval for Acute Ischemic Stroke: A Systematic Review , 2012, Neurointervention.

[9]  H. Bäzner,et al.  Intracranial thrombectomy using the Solitaire stent: a historical vignette , 2011, Journal of NeuroInterventional Surgery.

[10]  M. Augustin,et al.  Early Single Center Experience with the Solitaire Thrombectomy Device for the Treatment of Acute Ischemic Stroke , 2011, Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences.

[11]  R. Chapot,et al.  Retrieval of Displaced Coil Using a Solitaire® Stent , 2010, Clinical Neuroradiology.

[12]  Michael Chen,et al.  Endovascular rescue of a misshapen intracranial stent: report of two cases , 2010, Journal of NeuroInterventional Surgery.

[13]  O. Witte,et al.  Solitaire stent for endovascular treatment of intracranial aneurysms: immediate and mid-term results in 15 patients with 17 aneurysms. , 2010, Journal of neuroradiology. Journal de neuroradiologie.

[14]  T. Liebig,et al.  Coil embolization of anterior circulation aneurysms supported by the Solitaire™ AB Neurovascular Remodeling Device , 2010, Neuroradiology.

[15]  J. Klisch,et al.  STENT‐ASSISTED COIL EMBOLIZATION OF POSTERIOR CIRCULATION ANEURYSMS USING SOLITAIRE AB: PRELIMINARY EXPERIENCE , 2009, Neurosurgery.

[16]  J. Thornton,et al.  Coil Migration through a Neuroform 3 Stent during Endovascular Coiling , 2009, Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences.

[17]  H. Feng,et al.  Surgical extraction of migrated coils via proximal segment of the anterior cerebral artery: an emergency alternative. , 2009, Neurology India.

[18]  C. Luo,et al.  Stent Management of Coil Herniation in Embolization of Internal Carotid Aneurysms , 2008, American Journal of Neuroradiology.

[19]  T. Jovin,et al.  Retrieval of a Displaced Detachable Coil and Intracranial Stent with an L5 Merci Retriever During Endovascular Embolization of an Intracranial Aneurysm , 2008, Journal of neuroimaging : official journal of the American Society of Neuroimaging.

[20]  H. Henkes,et al.  A new device for endovascular coil retrieval from intracranial vessels: alligator retrieval device. , 2006, AJNR. American journal of neuroradiology.

[21]  I. Saatci,et al.  Retrieval of an embolization coil from the internal carotid artery using the Amplatz Microsnare Retrieval System , 1995, CardioVascular and Interventional Radiology.