Design and Characterization of a Bidirectional, Low Cost Flowmeter for Neonatal Ventilation

In this paper, we propose a novel, low cost flowmeter suitable for application in disposable breathing circuits. The sensor consists of two nominally identical transistors employed as hot sensing elements, placed into a pipe where the fluid flows. The working principle is based on the convective heat transfer between the transistors, heated by Joule effect, and the colder hitting gas. The proposed design allows the sensor to discriminate flow direction. The sensor response has been numerically simulated, and the results validated by experimental trials varying the pipe diameter (i.e., 10 and 30 mm), the flowrate values (ranging from - 8 to 8 L/min), and the collector current (i.e., 100, 300, and 500 mA). Experimental results show that the configuration with a pipe diameter of 10 mm at the highest collector current guarantees the highest mean sensitivity (1143 mV/L· min-1) at low flowrate (i.e., ±1 L/min); in addition, this configuration ensures the minimum dead space (0.5 versus 5 mL for 30 mm of diameter). However, the 30-mm pipe diameter allows extending the range of measurement (up to ±8 versus ±3.5 L/min at 10 mm), and improving both the discrimination threshold (<;0.1 L/min) and the symmetry of response. The response time of the sensor is 340 ms. These characteristics together with the low dead space and low cost foster its application to neonatal ventilation.

[1]  A. Lay-Ekuakille,et al.  Spirometric Measurement Postprocessing: Expiration Data Recovery , 2010, IEEE Sensors Journal.

[2]  Sergio Silvestri,et al.  Linearity dependence on oxygen fraction and gas temperature of a novel Fleisch pneumotachograph for neonatal ventilation at low flow rates , 2012 .

[3]  Anton F. P. van Putten,et al.  Thermal flow measurements at Gr/Re2≫1 by silicon anemometry , 1999, IEEE Trans. Instrum. Meas..

[4]  D. G. Martin,et al.  Resistance to airflow in anaesthetic breathing systems. , 1989, British journal of anaesthesia.

[5]  M. Putten,et al.  Discovery of linear asymptotic behaviour at micro-Kelvin resolution in a modulated Rayleigh–Bénard chamber , 2007 .

[6]  Emiliano Schena,et al.  A new ecological method for the estimation of Nutritive Sucking Efficiency in newborns: Measurement principle and experimental assessment , 2013, 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).

[7]  Sergio Silvestri,et al.  A high sensitivity fiber optic macro-bend based gas flow rate transducer for low flow rates: theory, working principle, and static calibration. , 2013, The Review of scientific instruments.

[8]  Sergio Silvestri,et al.  Micromachined Flow Sensors in Biomedical Applications , 2012, Micromachines.

[9]  Emiliano Schena,et al.  Technological Solutions and Main Indices for the Assessment of Newborns' Nutritive Sucking: A Review , 2014, Sensors.

[10]  Walter Lang,et al.  Minimum Detectable Air Velocity by Thermal Flow Sensors , 2013, Sensors.

[11]  Sergio Silvestri,et al.  A transistor based air flow transducer for thermohygrometric control of neonatal ventilatory applications. , 2008, The Review of scientific instruments.

[12]  Anton F. P. van Putten,et al.  Silicon thermal anemometry : developments and applications , 1996 .

[13]  Sergio Silvestri,et al.  An orifice meter for bidirectional air flow measurements: Influence of gas thermo-hygrometric content on static response and bidirectionality , 2013 .

[14]  A. Coates,et al.  A very low dead space pneumotachograph for ventilatory measurements in newborns. , 1990, Journal of applied physiology.

[15]  Louis Vessot King,et al.  On the Convection of Heat from Small Cylinders in a Stream of Fluid: Determination of the Convection Constants of Small Platinum Wires with Applications to Hot-Wire Anemometry , 1914 .

[16]  F. Flack,et al.  Comparative Review of Techniques for Recording Respiratory Events at Rest and during Deglutition , 1997, Dysphagia.