The Systematic Bias of Ingestible Core Temperature Sensors Requires a Correction by Linear Regression

An accurate measure of core body temperature is critical for monitoring individuals, groups and teams undertaking physical activity in situations of high heat stress or prolonged cold exposure. This study examined the range in systematic bias of ingestible temperature sensors compared to a certified and traceable reference thermometer. A total of 119 ingestible temperature sensors were immersed in a circulated water bath at five water temperatures (TEMP A: 35.12 ± 0.60°C, TEMP B: 37.33 ± 0.56°C, TEMP C: 39.48 ± 0.73°C, TEMP D: 41.58 ± 0.97°C, and TEMP E: 43.47 ± 1.07°C) along with a certified traceable reference thermometer. Thirteen sensors (10.9%) demonstrated a systematic bias > ±0.1°C, of which 4 (3.3%) were > ± 0.5°C. Limits of agreement (95%) indicated that systematic bias would likely fall in the range of −0.14 to 0.26°C, highlighting that it is possible for temperatures measured between sensors to differ by more than 0.4°C. The proportion of sensors with systematic bias > ±0.1°C (10.9%) confirms that ingestible temperature sensors require correction to ensure their accuracy. An individualized linear correction achieved a mean systematic bias of 0.00°C, and limits of agreement (95%) to 0.00–0.00°C, with 100% of sensors achieving ±0.1°C accuracy. Alternatively, a generalized linear function (Corrected Temperature (°C) = 1.00375 × Sensor Temperature (°C) − 0.205549), produced as the average slope and intercept of a sub-set of 51 sensors and excluding sensors with accuracy outside ±0.5°C, reduced the systematic bias to < ±0.1°C in 98.4% of the remaining sensors (n = 64). In conclusion, these data show that using an uncalibrated ingestible temperature sensor may provide inaccurate data that still appears to be statistically, physiologically, and clinically meaningful. Correction of sensor temperature to a reference thermometer by linear function eliminates this systematic bias (individualized functions) or ensures systematic bias is within ±0.1°C in 98% of the sensors (generalized function).

[1]  Aaron J. E. Bach,et al.  Intraocular Pressure Is a Poor Predictor of Hydration Status following Intermittent Exercise in the Heat , 2017, Front. Physiol..

[2]  A. Skurvydas,et al.  Brief Rewarming Blunts Hypothermia-Induced Alterations in Sensation, Motor Drive and Cognition , 2016, Front. Physiol..

[3]  Sébastien Racinais,et al.  Validation of an ingestible temperature data logging and telemetry system during exercise in the heat , 2016, Temperature.

[4]  Joanne N. Caldwell,et al.  Heat strain during military training activities: The dilemma of balancing force protection and operational capability , 2016, Temperature.

[5]  J. Castellani,et al.  Cold Stress Effects on Exposure Tolerance and Exercise Performance. , 2015, Comprehensive Physiology.

[6]  B. Aisbett,et al.  Simulated Firefighting Task Performance and Physiology Under Very Hot Conditions , 2015, Front. Physiol..

[7]  Ian B Stewart,et al.  Inside the 'Hurt Locker': the combined effects of explosive ordnance disposal and chemical protective clothing on physiological tolerance time in extreme environments , 2015, Extreme Physiology & Medicine.

[8]  Ben Rattray,et al.  Repeat work bouts increase thermal strain for Australian firefighters working in the heat , 2015, International journal of occupational and environmental health.

[9]  Andrew J. Pyke,et al.  Heat strain evaluation of overt and covert body armour in a hot and humid environment. , 2015, Applied ergonomics.

[10]  G. Kenny,et al.  Considerations for the measurement of core, skin and mean body temperatures. , 2014, Journal of thermal biology.

[11]  L. Taylor,et al.  Exposure to hot and cold environmental conditions does not affect the decision making ability of soccer referees following an intermittent sprint protocol , 2014, Front. Physiol..

[12]  G. Tew,et al.  Reliability of Intestinal Temperature Using an Ingestible Telemetry Pill System During Exercise in a Hot Environment , 2014, Journal of strength and conditioning research.

[13]  A. Gauthier,et al.  Calibration and performance assessment of a temperature sensor prototype using a 1-point calibration procedure. , 2012, The Review of scientific instruments.

[14]  Heather M. Logan-Sprenger,et al.  Effects of dehydration during cycling on skeletal muscle metabolism in females. , 2012, Medicine and science in sports and exercise.

[15]  Douglas J Casa,et al.  Validity of field expedient devices to assess core temperature during exercise in the cold. , 2011, Aviation, space, and environmental medicine.

[16]  Ian B Stewart,et al.  Negligible heat strain in armored vehicle officers wearing personal body armor , 2011, Journal of occupational medicine and toxicology.

[17]  Douglas J Casa,et al.  The American football uniform: uncompensable heat stress and hyperthermic exhaustion. , 2010, Journal of athletic training.

[18]  Douglas J Casa,et al.  Validity and reliability of devices that assess body temperature during indoor exercise in the heat. , 2009, Journal of athletic training.

[19]  Andrew P. Hunt,et al.  Calibration of an ingestible temperature sensor , 2008, Physiological measurement.

[20]  J. Castellani,et al.  National Athletic Trainers' Association position statement: environmental cold injuries. , 2008, Journal of athletic training.

[21]  G. Atkinson,et al.  The effectiveness of hand cooling at reducing exercise-induced hyperthermia and improving distance-race performance in wheelchair and able-bodied athletes. , 2008, Journal of applied physiology.

[22]  Victoria L Richmond,et al.  The effect of cool water ingestion on gastrointestinal pill temperature. , 2008, Medicine and science in sports and exercise.

[23]  D. Altman,et al.  Agreement Between Methods of Measurement with Multiple Observations Per Individual , 2007, Journal of biopharmaceutical statistics.

[24]  Douglas J Casa,et al.  Validity of devices that assess body temperature during outdoor exercise in the heat. , 2007, Journal of athletic training.

[25]  C. L. Lim,et al.  The ingestible telemetric body core temperature sensor: a review of validity and exercise applications , 2006, British Journal of Sports Medicine.

[26]  G. Atkinson,et al.  The validity and reliability of intestinal temperature during intermittent running. , 2006, Medicine and science in sports and exercise.

[27]  C. L. Lim,et al.  Continuous thermoregulatory responses to mass-participation distance running in heat. , 2006, Medicine and science in sports and exercise.

[28]  G Machin,et al.  Traceability and calibration in temperature measurement: a clinical necessity , 2006, Journal of medical engineering & technology.

[29]  Robert Carter,et al.  Epidemiology of hospitalizations and deaths from heat illness in soldiers. , 2005, Medicine and science in sports and exercise.

[30]  David W. DeGroot,et al.  Extremes of human heat tolerance: life at the precipice of thermoregulatory failure , 2004 .

[31]  J. M. Mckay,et al.  Physiological tolerance to uncompensable heat stress: effects of exercise intensity, protective clothing, and climate. , 1994, Journal of applied physiology.

[32]  Fang Zhao,et al.  A retrospective analysis of American football hyperthermia deaths in the United States , 2010, International Journal of Biometeorology.

[33]  A D Harper Smith,et al.  The validity of wireless iButtons® and thermistors for human skin temperature measurement , 2010, Physiological measurement.

[34]  R. Shepherd Validity of Devices That Assess Body Temperature During Outdoor Exercise in the Heat , 2008 .

[35]  R. Shephard Continuous Thermoregulatory Responses to Mass-Participation Distance Running in Heat , 2007 .

[36]  S. Schneider,et al.  Preliminary results: Effect of whole-body cooling in patients with myasthenia gravis. , 2006, Medicine and science in sports and exercise.