Diagnostic quality of multiplanar reformations obtained with a newly developed cone beam device for maxillofacial imaging.

OBJECTIVES To compare the diagnostic quality of multiplanar reformations generated by a newly developed cone beam device with the diagnostic quality of multiplanar reformations obtained using a state of the art multidetector row CT scanner. METHODS Digital volume tomograms (DVTs) were obtained with the prototype of GALILEOS (Sirona Dental Systems Inc., Bensheim, Germany), a newly developed compact-sized cone beam machine with a scan volume of 15 x 15 x 15 cm. CT scans were performed for comparison with two 16-detector row CT scanners. The study included 30 image pairs obtained from the same patients. The multiplanar reformations in axial, coronal and sagittal planes were subjectively assessed by 3 observers using a 1(excellent)-5(poor) scale on 35 weighted criteria including findings detection, image quality and visualization of several anatomical structures. RESULTS There was no statistically significant difference between DVT and CT in cumulative values expressing the diagnostic quality of the reformations (P = 0.071). The interobserver agreement was better with CT than with DVT. Findings detection and the visualization of the majority of anatomical structures within the mandibulomaxillary area were ranked at the same level for DVT as with CT. CONCLUSIONS The diagnostic quality of multiplanar reformations can be considered the same in both devices. Although sharpness, noise level and contrast resolution do not reach the level of CT, DVT images proved statistically not inferior for the detection of lesions which may be adequately depicted by both imaging modes.

[1]  J Brief,et al.  Clinical indications for digital volume tomography in oral and maxillofacial surgery. , 2002, Dento maxillo facial radiology.

[2]  David C Hatcher,et al.  Cone beam CT for pre-surgical assessment of implant sites. , 2003, Journal of the California Dental Association.

[3]  R. Schmelzle,et al.  Postoperative imaging of zygomaticomaxillary complex fractures using digital volume tomography. , 2004, Journal of oral and maxillofacial surgery : official journal of the American Association of Oral and Maxillofacial Surgeons.

[4]  S. Brooks,et al.  Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit. , 2003, Dento maxillo facial radiology.

[5]  A. Ahuja,et al.  Dentascan in oral imaging. , 2001, Clinical Radiology.

[6]  Y. Nakagawa,et al.  Preoperative application of limited cone beam computerized tomography as an assessment tool before minor oral surgery. , 2002, International journal of oral and maxillofacial surgery.

[7]  H. Stamatakis,et al.  Radiographic examination of the temporomandibular joint using cone beam computed tomography. , 2004, Dento maxillo facial radiology.

[8]  J. Mühling,et al.  [Digital volume tomography--an extension to the diagnostic procedures available for application before surgical removal of third molars]. , 2002, Mund-, Kiefer- und Gesichtschirurgie : MKG.

[9]  S. Brooks,et al.  Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CAT. , 2006, Dento maxillo facial radiology.

[10]  K. Ueda,et al.  Using a flat-panel detector in high resolution cone beam CT for dental imaging. , 2004, Dento maxillo facial radiology.

[11]  J. Mühling,et al.  Erweiterte Diagnostik im Rahmen der operativen Weisheitszahnentfernung mittels digitaler Volumentomographie , 2002, Mund-, Kiefer- und Gesichtschirurgie.

[12]  André Gahleitner,et al.  Dental CT: imaging technique, anatomy, and pathologic conditions of the jaws , 2003, European Radiology.

[13]  Masao Araki,et al.  Comparison of image performance between cone-beam computed tomography for dental use and four-row multidetector helical CT. , 2006, Journal of oral science.

[14]  Yoshinori Arai,et al.  A comparison of a new limited cone beam computed tomography machine for dental use with a multidetector row helical CT machine. , 2003, Oral surgery, oral medicine, oral pathology, oral radiology, and endodontics.

[15]  E Tammisalo,et al.  Development of a compact computed tomographic apparatus for dental use. , 1999, Dento maxillo facial radiology.

[16]  Masahiro Izumi,et al.  Effectiveness of dental computed tomography in diagnostic imaging of periradicular lesion of each root of a multirooted tooth: a case report. , 2006, Journal of endodontics.

[17]  R. Schmelzle,et al.  SHORT COMMUNICATION Intraoperative 3D imaging of the facial skeleton using the SIREMOBIL Iso-C 3D , 2004 .

[18]  J. Abrahams,et al.  Dental CT imaging: a look at the jaw. , 2001, Radiology.

[19]  George Hripcsak,et al.  Measuring agreement in medical informatics reliability studies , 2002, J. Biomed. Informatics.

[20]  P. Mozzo,et al.  A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results , 1998, European Radiology.