Spectrophotometric determination of aqueous sulfide on a pneumatically enhanced centrifugal microfluidic platform.

A pneumatically enhanced centrifugal microfluidic platform was developed for rapid spectrophotometric determination of aqueous sulfide. This platform performs an automated analysis based on the reaction between hydrogen sulfide and N,N-dimethyl-p-phenylenediamine in the presence of iron(III) chloride to form Methylene Blue. The platform design minimizes the number of integrated valves required, compared to other centrifugal systems, significantly improving the ease of fabrication. The sequential analytical procedure and spectrophotometric analyses were performed directly on-disk, demonstrating significant advantages in portability and cost over conventional analytical methods. This method allows for rapid and precise determination of aqueous sulfide in the concentration range of 0.4-2.0 mg L(-1), which can be extended by a pneumatically induced, on-disk serial dilution to 6.0 mg L(-1). A detection limit of 0.4 mg L(-1) was calculated for this pneumatically enhanced method.

[1]  Eric D Salin,et al.  A valveless pneumatic fluid transfer technique applied to standard additions on a centrifugal microfluidic platform. , 2011, Analytical chemistry.

[2]  M. Miró,et al.  A multisyringe flow injection method for the automated determination of sulfide in waters using a miniaturised optical fiber spectrophotometer. , 2004, Talanta.

[3]  J. K. Fogo,et al.  Spectrophotometric Determination of Hydrogen Sulfide , 1949 .

[4]  R. Santelli,et al.  On Line Dissolution of ZnS For Sulfide Determination in Stabilized Water Samples with Zinc Acetate, Using Spectrophotometry by Methylene Blue Method , 1999 .

[5]  J. R. Kramer,et al.  Trace sulfide determination in oxic freshwaters , 2003 .

[6]  L. Gustafsson Determination of ultramicro amounts of sulphate as methylene blue—I: The colour reaction , 1960 .

[7]  J. Morse,et al.  Examination and Refinement of the Determination of Aqueous Hydrogen Sulfide by the Methylene Blue Method , 2011 .

[8]  R. Santelli,et al.  Determination of sulfide in waters by flow-injection solid phase spectrophotometry , 2000 .

[9]  Joel D. Cline,et al.  SPECTROPHOTOMETRIC DETERMINATION OF HYDROGEN SULFIDE IN NATURAL WATERS1 , 1969 .

[10]  Eric D Salin,et al.  Pneumatically pumping fluids radially inward on centrifugal microfluidic platforms in motion. , 2010, Analytical chemistry.

[11]  Andreas Manz,et al.  Latest developments in micro total analysis systems. , 2010, Analytical chemistry.

[12]  Jintae Kim,et al.  Centrifugal microfluidics for biomedical applications. , 2010, Lab on a chip.

[13]  A centrifugal microanalysis system for the determination of nitrite and hexavalent chromium. , 2009, Talanta.

[14]  Marc Madou,et al.  Lab on a CD. , 2006, Annual review of biomedical engineering.

[15]  R. Zengerle,et al.  Microfluidic lab-on-a-chip platforms: requirements, characteristics and applications. , 2010, Chemical Society reviews.

[16]  J. Masini,et al.  Application of sequential injection-monosegmented flow analysis (SI-MSFA) to spectrophotometric determination of sulfide in simulated waters samples. , 2003, Talanta.

[17]  Eric D Salin,et al.  A rapid prototyping technique for valves and filters in centrifugal microfluidic devices. , 2009, Lab on a chip.

[18]  Eric D Salin,et al.  Pneumatic flow switching on centrifugal microfluidic platforms in motion. , 2011, Analytical chemistry.

[19]  J. Masini,et al.  Spectrophotometric determination of acid volatile sulfide in river sediments by sequential injection analysis exploiting the methylene blue reaction. , 2001, Talanta.

[20]  Eric D Salin,et al.  Magnetically driven solid sample preparation for centrifugal microfluidic devices. , 2009, Analytical chemistry.

[21]  Alexei Kazarine,et al.  Automated liquid-liquid extraction by pneumatic recirculation on a centrifugal microfluidic platform. , 2012, Analytical chemistry.

[22]  Marc Madou,et al.  Infrared controlled waxes for liquid handling and storage on a CD-microfluidic platform. , 2011, Lab on a chip.

[23]  H. Nishita,et al.  Microestimation of Sulfur in Plant Materials, Soils, and Irrigation Waters , 1952 .