Pulse oximetry during lumbar epidural anesthesia: reliability of values measured at the hand and the foot.

Pulse oximetry is dependent upon the presence of a pulsating vascular bed. The signal detection will be impaired in the presence of vasoconstriction or venous congestion, conditions which may occur readily in clinical practice. We compared the oximetric measurements (SpO2) at the hand and the foot with arterial hemoglobin saturation (SaO2) during lumbar epidural anesthesia. After administration of a crystalloid solution (20 mL/kg body weight lactated Ringer's solution), 40 adult male patients, scheduled for inguinal hernioplasty, received 15 mL of 0.50% plain bupivacaine into the lumbar epidural space. Two pulse oximeter probes were applied to the index finger and toe of the patients, and the SpO2 values were recorded continuously. Arterial hemoglobin saturation (SaO2) was measured using a co-oximeter 5 min before and 30 min after the onset of sensory block. No significant differences were detected between SaO2 (97.7%, SD 0.4%) and SpO2 basal values recorded from the hand (97.8%, SD 0.8%) and the foot (98.1%, SD 0.4%). After the onset of epidural anesthesia, a progressive decrease of SpO2 value recorded from the hand was observed: at 30 min it was 92.3% +/- 1.3% (P < 0.01 compared with baseline). At the same time, the SaO2 value was 97.5% +/- 0.9% (P < 0.01 compared with SpO2 from the hand). On the contrary, no significant difference from both basal value and SaO2 was detected in SpO2 measurements from the toe during the epidural block. In all patients intraoperative decrease of heart rate and arterial blood pressure was < or = 15% from baseline.(ABSTRACT TRUNCATED AT 250 WORDS)

[1]  W B Runciman,et al.  Potential errors in pulse oximetry , 1991, Anaesthesia.

[2]  R. K. Webb,et al.  A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion , 1991, Anaesthesia.

[3]  B. Robinson,et al.  Comparison of pulse oximeters: accuracy at low arterial pressure in volunteers. , 1990, British journal of anaesthesia.

[4]  J. Severinghaus,et al.  Pulse oximeter failure thresholds in hypotension and vasoconstriction. , 1990, Anesthesiology.

[5]  C. Hanning,et al.  Comparison of four pulse oximeters: effects of venous occlusion and cold-induced peripheral vasoconstriction. , 1990, British journal of anaesthesia.

[6]  A. Coe,et al.  Is crystalloid preloading useful in spinal anaesthesia in the elderly? , 1990, Anaesthesia.

[7]  Epidural Block for Caesarean Section and Circulatory Changes , 1989, The Lancet.

[8]  C. Hanning,et al.  Comparison of pulse oximeters: effects of vasoconstriction and venous engorgement. , 1989, British journal of anaesthesia.

[9]  J B Gross,et al.  Principles of Pulse Oximetry: Theoretical and Practical Considerations , 1989, Anesthesia and analgesia.

[10]  R. Loeb,et al.  Blood flow limits and pulse oximeter signal detection. , 1987, Anesthesiology.

[11]  D. Paulus,et al.  Pulse oximetry for vascular monitoring in upper extremity replantation surgery. , 1986, The Journal of hand surgery.

[12]  J. Arndt,et al.  Peridural Anesthesia and the Distribution of Blood in Supine Humans , 1985, Anesthesiology.

[13]  S. Sjögren,et al.  Circulatory Changes during Continuous Epidural Blockade , 1972, Acta anaesthesiologica Scandinavica. Supplementum.

[14]  J. Bonica,et al.  Circulatory Effects of Peridural Block: I. Effects of Level of Analgesia and Dose of Lidocaine , 1970, Anesthesiology.

[15]  M. Harmel,et al.  Physiology of Spinal Anesthesia , 1970, The Yale Journal of Biology and Medicine.

[16]  S. Shimosato,et al.  The Role of the Venous System in Cardiocirculatory Dynamics during Spinal and Epidural Anesthesia in Man , 1969, Anesthesiology.