Experimental evaluation of two simple thermal models using transient temperature analysis.

Thermal models are used to predict temperature distributions of heated tissues during thermal therapies. Recent interest in short duration high temperature therapeutic procedures necessitates the accurate modelling of transient temperature profiles in heated tissues. Blood flow plays an important role in tissue heat transfer and the resultant temperature distribution. This work examines the transient predictions of two simple mathematical models of heat transfer by blood flow (the bioheat transfer equation model and the effective thermal conductivity equation model) and compares their predictions to measured transient temperature data. Large differences between the two models are predicted in the tissue temperature distribution as a function of blood flow for a short heat pulse. In the experiments a hot water needle, approximately 30 degrees C above ambient, delivered a 20 s heating pulse to an excised fixed porcine kidney that was used as a flow model. Temperature profiles of a thermocouple that primarily traversed the kidney cortex were examined. Kidney locations with large vessels were avoided in the temperature profile analysis by examination of the vessel geometry using high resolution computed tomography angiography and the detection of the characteristic large vessel localized cooling or heating patterns in steady-state temperature profiles. It was found that for regions without large vessels, predictions of the Pennes bioheat transfer equation were in much better agreement with the experimental data when compared to predictions of the scalar effective thermal conductivity equation model. For example, at a location r approximately 2 mm away from the source, the measured delay time was 10.6 +/- 0.5 s compared to predictions of 9.4 s and 5.4 s of the BHTE and ETCE models, respectively. However, for the majority of measured locations, localized cooling and heating effects were detected close to large vessels when the kidney was perfused. Finally, it is shown that increasing flow in regions without large vessels minimally perturbs temperature profiles for short exposure times; regions with large vessels still have a significant effect.

[1]  H. H. Penns Analysis of tissue and arterial blood temperatures in the resting human forearm , 1948 .

[2]  J Crezee,et al.  Experimental verification of bioheat transfer theories: measurement of temperature profiles around large artificial vessels in perfused tissue. , 1990, Physics in medicine and biology.

[3]  M. Amichetti,et al.  Report of long-term follow-up in a randomized trial comparing radiation therapy and radiation therapy plus hyperthermia to metastatic lymph nodes in stage IV head and neck patients. , 1994, International journal of radiation oncology, biology, physics.

[4]  D E Lemons,et al.  Theory and experiment for the effect of vascular microstructure on surface tissue heat transfer--Part I: Anatomical foundation and model conceptualization. , 1984, Journal of biomechanical engineering.

[5]  R C Preston,et al.  Perfusion corrections for ultrasonic heating in nonperfused media. , 1996, Ultrasound in medicine & biology.

[6]  G. Gazelle,et al.  Nanocrystalline computed tomography contrast agents for blood-pool and liver-spleen imaging. , 1994, Investigative radiology.

[7]  D. Haines,et al.  New horizons in catheter ablation. , 1995, Journal of interventional cardiology.

[8]  R B Roemer,et al.  Effects of physical parameters on high temperature ultrasound hyperthermia. , 1990, Ultrasound in medicine & biology.

[9]  H. Arkin,et al.  Recent developments in modeling heat transfer in blood perfused tissues , 1994, IEEE Transactions on Biomedical Engineering.

[10]  D Machin,et al.  Radiotherapy with or without hyperthermia in the treatment of superficial localized breast cancer: results from five randomized controlled trials. International Collaborative Hyperthermia Group. , 1996, International journal of radiation oncology, biology, physics.

[11]  J Werner,et al.  Efficiency function: improvement of classical bioheat approach. , 1994, Journal of applied physiology.

[12]  J. Valvano,et al.  Analysis of the Weinbaum-Jiji model of blood flow in the canine kidney cortex for self-heated thermistors. , 1994, Journal of biomechanical engineering.

[13]  J. Hunt,et al.  Blood flow cooling and ultrasonic lesion formation. , 1996, Medical physics.

[14]  R B Roemer,et al.  A counter current vascular network model of heat transfer in tissues. , 1996, Journal of biomechanical engineering.

[15]  Kullervo Hynynen,et al.  MR monitoring of focused ultrasonic surgery of renal cortex: Experimental and simulation studies , 1995, Journal of magnetic resonance imaging : JMRI.

[16]  K. Hynynen,et al.  Simultaneous magnetic resonance phase and magnitude temperature maps in muscle , 1996, Magnetic resonance in medicine.

[17]  A W Dutton,et al.  The simulation of discrete vessel effects in experimental hyperthermia. , 1994, Journal of biomechanical engineering.

[18]  Francis A. Duck,et al.  Physical properties of tissue : a comprehensive reference book , 1990 .

[19]  J.J.W. Lagendijk,et al.  A fast and simple algorithm for the calculation of convective heat transfer by large vessels in three-dimensional inhomogeneous tissues , 1991, IEEE Transactions on Biomedical Engineering.

[20]  R. Roemer Optimal power deposition in hyperthermia. I. The treatment goal: the ideal temperature distribution: the role of large blood vessels. , 1991, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[21]  H. H. Pennes Analysis of tissue and arterial blood temperatures in the resting human forearm. 1948. , 1948, Journal of applied physiology.

[22]  K. Hynynen,et al.  Experimental evaluation of two simple thermal models using hyperthermia in muscle in vivo. , 1993, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[23]  G. T. Martin,et al.  Thermal model for the local microwave hyperthermia treatment of benign prostatic hyperplasia , 1992, IEEE Transactions on Biomedical Engineering.

[24]  Michael C. Kolios,et al.  Magnetic resonance imaging of temperature changes during interstitial microwave heating: a phantom study. , 1997, Medical physics.

[25]  J Werner,et al.  Estimation of the thermal effect of blood flow in a branching countercurrent network using a three-dimensional vascular model. , 1994, Journal of biomechanical engineering.

[26]  P. Ask,et al.  Computer analysis of hyperthermia treatment of the prostate , 1997 .

[27]  J C Chato,et al.  Heat transfer to blood vessels. , 1980, Journal of biomechanical engineering.

[28]  H C Charles,et al.  1H MRI phase thermometry in vivo in canine brain, muscle, and tumor tissue. , 1996, Medical physics.

[29]  K. Hynynen,et al.  Magnetic resonance‐guided thermal surgery , 1993, Magnetic resonance in medicine.

[30]  J. Mooibroek,et al.  Interstitial heating: experiments in artificially perfused bovine tongues. , 1991, Physics in medicine and biology.

[31]  Andrew William. Dutton A finite element solution to conjugated heat transfer in tissue using magnetic resonance angiography to measure the in vitro velocity field. , 1993 .

[32]  J Lagendijk,et al.  A description of discrete vessel segments in thermal modelling of tissues. , 1996, Physics in medicine and biology.

[33]  J. Hunt,et al.  Rapid heating: critical theoretical assessment of thermal gradients found in hyperthermia treatments. , 1991, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[34]  O Mella,et al.  Hyperthermia as an adjuvant to radiation therapy of recurrent or metastatic malignant melanoma. A multicentre randomized trial by the European Society for Hyperthermic Oncology. , 2009, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[35]  Michael C. Kolios,et al.  Large blood vessel cooling in heated tissues: a numerical study. , 1995, Physics in medicine and biology.

[36]  D W Holdsworth,et al.  An investigation of the flow dependence of temperature gradients near large vessels during steady state and transient tissue heating. , 1999, Physics in medicine and biology.

[37]  Kenneth R. Holmes,et al.  MICROVASCULAR CONTRIBUTIONS IN TISSUE HEAT TRANSFER , 1980, Annals of the New York Academy of Sciences.

[38]  G. Haar,et al.  Ultrasound focal beam surgery. , 1995 .

[39]  S. Weinbaum,et al.  A new fundamental bioheat equation for muscle tissue: Part I--Blood perfusion term. , 1997, Journal of biomechanical engineering.

[40]  A Fenster,et al.  A high-resolution XRII-based quantitative volume CT scanner. , 1993, Medical physics.