Lens Exposure during Brain Scans Using Multidetector Row CT Scanners: Methods for Estimation of Lens Dose

BACKGROUND AND PURPOSE: Some recent studies on radiation lens injuries have indicated much lower dose thresholds than specified by the current radiation protection guidelines. The purpose of this research was to measure the lens dose during brain CT scans with multidetector row CT and to assess methods for estimating the lens dose. MATERIALS AND METHODS: With 8 types of multidetector row CT scanners, both axial and helical scans were obtained for the head part of a human-shaped phantom by using normal clinical settings with the orbitomeatal line as the baseline. We measured the doses on both eyelids by using an RPLGD during whole-brain scans including the orbit with the starting point at the level of the inferior orbital rim. To assess the effect of the starting points on the lens doses, we measured the lens doses by using 2 other starting points for scanning (the orbitomeatal line and the superior orbital rim). RESULTS: The CTDIvols and the lens doses during whole-brain CT including the orbit were 50.9–113.3 mGy and 42.6–103.5 mGy, respectively. The ratios of lens dose to CTDIvol were 80.6%–103.4%. The lens doses decreased as the starting points were set more superiorly. The lens doses during scans from the superior orbital rim were 11.8%–20.9% of the doses during the scans from the inferior orbital rim. CONCLUSIONS: CTDIvol can be used to estimate the lens dose during whole-brain CT when the orbit is included in the scanning range.

[1]  Icrp 1990 Recommendations of the International Commission on Radiological Protection , 1991 .

[2]  Bruce H Alexander,et al.  Risk of cataract after exposure to low doses of ionizing radiation: a 20-year prospective cohort study among US radiologic technologists. , 2008, American journal of epidemiology.

[3]  Cynthia H McCollough,et al.  CT dosimetry: comparison of measurement techniques and devices. , 2008, Radiographics : a review publication of the Radiological Society of North America, Inc.

[4]  Atsushi Minamoto,et al.  A REANALYSIS OF ATOMIC-BOMB CATARACT DATA, 2000–2002: A THRESHOLD ANALYSIS , 2006, Health physics.

[5]  D McLean,et al.  Paediatric and adult computed tomography practice and patient dose in Australia. , 2006, Australasian radiology.

[6]  Managing patient dose in computed tomography. A report of the International Commission on Radiological Protection. , 2000, Annals of the ICRP.

[7]  R. Shore,et al.  Postoperative Cataract Cases among Atomic Bomb Survivors: Radiation Dose Response and Threshold , 2007, Radiation research.

[8]  B Lane,et al.  Single- versus multi-detector row CT of the brain: quality assessment. , 2001, Radiology.

[9]  N. Theocharopoulos,et al.  Reduction of eye lens radiation dose by orbital bismuth shielding in pediatric patients undergoing CT of the head: a Monte Carlo study. , 2005, Medical physics.

[10]  K. Hopper,et al.  Radioprotection to the eye during CT scanning. , 2001, AJNR. American journal of neuroradiology.

[11]  E. Tsui,et al.  Lens Exclusion in Computed Tomography Scans of the Brain — The Local Practice , 2001 .

[12]  Michael F McNitt-Gray,et al.  AAPM/RSNA Physics Tutorial for Residents: Topics in CT. Radiation dose in CT. , 2002, Radiographics : a review publication of the Radiological Society of North America, Inc.

[13]  Icrp Recommendations of the International Commission on Radiological Protection Publication 60 , 1991 .

[14]  W. Schull,et al.  Cataract in atomic bomb survivors based on a threshold model and the occurrence of severe epilation. , 1996, Radiation research.

[15]  D. Dance,et al.  Analysis of surface dose variation in CT procedures. , 2001, The British journal of radiology.

[16]  J. Sjöstrand,et al.  A clinical study of radiation cataract formation in adult life following γ irradiation of the lens in early childhood , 1997 .

[17]  F A Cucinotta,et al.  Space Radiation and Cataracts in Astronauts , 2001, Radiation research.

[18]  Jack Valentin,et al.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. , 2007, Annals of the ICRP.

[19]  S. Demura,et al.  Prediction equation for head volume of Japanese young adults , 2005, Journal of sports sciences.

[20]  J. Valentin,et al.  Abstract: Avoidance of radiation injuries from medical interventional procedures, ICRP Publication 85 , 2000 .

[21]  R. Sievert,et al.  Book Reviews : Recommendations of the International Commission on Radiological Protection (as amended 1959 and revised 1962). I.C.R.P. Publication 6. 70 pp. PERGAMON PRESS. Oxford, London and New York, 1964. £1 5s. 0d. [TB/54] , 1964 .

[22]  W. P. Chang,et al.  Lenticular Opacities in Populations Exposed to Chronic Low-Dose-Rate Gamma Radiation from Radiocontaminated Buildings in Taiwan , 2001, Radiation research.

[23]  R. Shore,et al.  Cataracts among Chernobyl Clean-up Workers: Implications Regarding Permissible Eye Exposures , 2007, Radiation research.

[24]  A. Hida,et al.  Cataract in atomic bomb survivors , 2004, International journal of radiation biology.

[25]  T. Kron,et al.  Variation of patient dose in head CT. , 1998, The British journal of radiology.

[26]  L. Yeoman,et al.  Gantry angulation in brain CT: dosage implications, effect on posterior fossa artifacts, and current international practice. , 1992, Radiology.

[27]  Vilhjalmur Rafnsson,et al.  Cosmic radiation increases the risk of nuclear cataract in airline pilots: a population-based case-control study. , 2005, Archives of ophthalmology.

[28]  J. Strohmaier,et al.  Diagnostic x-ray exposure and lens opacities: the Beaver Dam Eye Study. , 1993, American journal of public health.