A Primary Study of Indirect ECG Monitor Embedded in a Bed for Home Health Care

A system for monitoring electrocardiogram (ECG) through clothes inserted between the measuring electrodes and the body surface of a subject when lying on a mattress has been proposed. The principle of the system is based on capacitive coupling involving the electrode, the clothes, and the skin. Validation of the system revealed the following: (1) In spite of the gain attenuation in the pass band of the system, distortion of the detected signal was subtle even when clothes thicker than 1mm were inserted, (2) The system was able to yield a stable ECG from a subject particularly during sound sleep, (3) The system succeeded in detecting ECG after changing the posture into any of supine, right lateral, or left lateral positions by adopting a newly devised electrode configuration. Therefore, the proposed system appears promising for application to bedding as a non-invasive and awareness-free system for ECG monitoring during sleep.

[1]  Vijay K. Jain,et al.  A New Technique for Monitoring Heart Signals-Part I: Instrumentation Design , 1986, IEEE Transactions on Biomedical Engineering.

[2]  Akinori Ueno,et al.  A System for Detecting Electrocardiographic Potential through Underwear Worn by an Infant from Its Dorsal Surface , 2007 .

[3]  Robert Puers,et al.  Towards the integration of textile sensors in a wireless monitoring suit , 2004 .

[4]  L. Geddes,et al.  Temporal changes in electrode impedance while recording the electrocardiogram with “Dry” electrodes , 1973, Annals of Biomedical Engineering.

[5]  Vijay K. Jain,et al.  A New Technique for Monitoring Heart Signals-Part II: Orthogonal Lead Extraction , 1986, IEEE Transactions on Biomedical Engineering.

[6]  Akinori Ueno,et al.  Detection of Electrocardiogram by Electrodes with Fabrics Using Capacitive Coupling , 2004 .

[7]  M. Ishijima,et al.  Cardiopulmonary monitoring by textile electrodes without subject-awareness of being monitored , 1997, Medical and Biological Engineering and Computing.

[8]  Tsuyoshi Kato,et al.  Capacitive Sensing of Electrocardiographic Potential Through Cloth From the Dorsal Surface of the Body in a Supine Position: A Preliminary Study , 2007, IEEE Transactions on Biomedical Engineering.

[9]  T Togawa,et al.  Observation of electrocardiograms through tap water. , 1989, Clinical physics and physiological measurement : an official journal of the Hospital Physicists' Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics.

[10]  Enzo Pasquale Scilingo,et al.  Performance evaluation of sensing fabrics for monitoring physiological and biomechanical variables , 2005, IEEE Transactions on Information Technology in Biomedicine.

[11]  Ilkka Korhonen,et al.  Health monitoring in the home of the future. , 2003, IEEE engineering in medicine and biology magazine : the quarterly magazine of the Engineering in Medicine & Biology Society.

[12]  Sungmee Park,et al.  Enhancing the quality of life through wearable technology , 2003, IEEE Engineering in Medicine and Biology Magazine.

[13]  T Togawa,et al.  Unconstrained heart-rate monitoring during bathing. , 1997, Biomedical instrumentation & technology.

[14]  M. Ishijima Monitoring of electrocardiograms in bed without utilizing body surface electrodes , 1993, IEEE Transactions on Biomedical Engineering.