Enclosed Electronic System for Force Measurements in Knee Implants

Total knee arthroplasty is a widely performed surgical technique. Soft tissue force balancing during the operation relies strongly on the experience of the surgeon in equilibrating tension in the collateral ligaments. Little information on the forces in the implanted prosthesis is available during surgery and post-operative treatment. This paper presents the design, fabrication and testing of an instrumented insert performing force measurements in a knee prosthesis. The insert contains a closed structure composed of printed circuit boards and incorporates a microfabricated polyimide thin-film piezoresistive strain sensor for each condylar compartment. The sensor is tested in a mechanical knee simulator that mimics in-vivo conditions. For characterization purposes, static and dynamic load patterns are applied to the instrumented insert. Results show that the sensors are able to measure forces up to 1.5 times body weight with a sensitivity fitting the requirements for the proposed use. Dynamic testing of the insert shows a good tracking of slow and fast changing forces in the knee prosthesis by the sensors.

[1]  G. Bergmann,et al.  Loading of the knee joint during activities of daily living measured in vivo in five subjects. , 2010, Journal of biomechanics.

[2]  Georg Bergmann,et al.  The influence of footwear on knee joint loading during walking--in vivo load measurements with instrumented knee implants. , 2013, Journal of biomechanics.

[3]  Christopher Townsend,et al.  A multiaxial force-sensing implantable tibial prosthesis. , 2006, Journal of biomechanics.

[4]  Oguz Atasoy,et al.  Remote powering realization for smart orthopedic implants , 2012, 10th IEEE International NEWCAS Conference.

[5]  Rajesh Rajamani,et al.  Battery-Less Wireless Instrumented Knee Implant , 2013 .

[6]  R. McCarney,et al.  Knee pain and osteoarthritis in older adults: a review of community burden and current use of primary health care , 2001, Annals of the rheumatic diseases.

[7]  Marcus G Pandy,et al.  Grand challenge competition to predict in vivo knee loads , 2012, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[8]  D. D’Lima,et al.  An implantable telemetry device to measure intra-articular tibial forces. , 2005, Journal of biomechanics.

[9]  R. Hague,et al.  An empirical study into laser sintering of ultra-high molecular weight polyethylene (UHMWPE) , 2010 .

[10]  K. Aminian,et al.  Design and test of a MEMS strain-sensing device for monitoring artificial knee implants , 2013, Biomedical microdevices.

[11]  B. Rath,et al.  Evaluation of implant position and knee alignment after patient-specific unicompartmental knee arthroplasty. , 2011, The Knee.

[12]  C. Cooper,et al.  Risk factors for the incidence and progression of radiographic knee osteoarthritis. , 2000, Arthritis and rheumatism.

[13]  O. Harrysson,et al.  Custom-designed orthopedic implants evaluated using finite element analysis of patient-specific computed tomography data: femoral-component case study , 2007, BMC musculoskeletal disorders.

[14]  Swee Chuan Tjin,et al.  Pressure mapping at orthopaedic joint interfaces with fiber Bragg gratings , 2006 .

[15]  Arash Arami Kinematics Estimation and Loosening Detection in Smart Knee Prostheses , 2014 .

[16]  G. Bergmann,et al.  Design, calibration and pre-clinical testing of an instrumented tibial tray. , 2007, Journal of biomechanics.

[17]  Kamiar Aminian,et al.  Accurate internal-external rotation measurement in total knee prostheses: A magnetic solution. , 2012, Journal of biomechanics.

[18]  Mauro Serpelloni,et al.  Design and test of an autonomous sensor for force measurements in human knee implants , 2011 .

[19]  Pierre-André Farine,et al.  An Analog Front-End and ADC Integrated Circuit for Implantable Force and Orientation Measurements in Joint Prosthesis , 2012, MobiHealth.

[20]  R. Barrack,et al.  Do Patient-specific Guides Improve Coronal Alignment in Total Knee Arthroplasty? , 2012, Clinical orthopaedics and related research.

[21]  Oguz Atasoy,et al.  Instrumented Knee Prosthesis for Force and Kinematics Measurements , 2013, IEEE Transactions on Automation Science and Engineering.

[22]  Jiankang He,et al.  Custom fabrication of a composite hemi‐knee joint based on rapid prototyping , 2006 .

[23]  Oguz Atasoy,et al.  Instrumented prosthesis for knee implants monitoring , 2011, 2011 IEEE International Conference on Automation Science and Engineering.

[24]  J. Pritchett Heat Generated by Knee Prostheses , 2006, Clinical orthopaedics and related research.

[25]  L. Rothman Properties of Thin Polyimide Films , 1980 .

[26]  Lutz-Peter Nolte,et al.  Development of a force amplitude- and location-sensing device designed to improve the ligament balancing procedure in TKA , 2005, IEEE Transactions on Biomedical Engineering.

[27]  Brian R Wolf,et al.  Total knee arthroplasty volume, utilization, and outcomes among Medicare beneficiaries, 1991-2010. , 2012, JAMA.

[28]  P. Renaud,et al.  Partial release and detachment of microfabricated metal and polymer structures by anodic metal dissolution , 2005, Journal of Microelectromechanical Systems.