Simulation study of a novel target oriented SPECT design using a variable pinhole collimator

Purpose: In the past decade, demands for organ specific (target oriented) single‐photon emission computed tomography (SPECT) is increasing, and several groups have conducted studies on developing clinical dedicated SPECT with pinhole collimator to improve the spatial resolution. However, acceptance angle of the collimator cannot be adjusted to fit the different ROIs of target objects because the shape of pinhole could not be changed, and the magnifying factor cannot be maximized as the collimator‐to‐detector distance is fixed. Furthermore, those dedicated pinhole SPECTs are typically made for a single purpose and therefore possess a drawback in that it cannot be utilized for any other purpose. In this study, we propose a novel SPECT system using variable pinhole collimator (VP SPECT) whose parameters are flexible. Methods: The proposed variable pinhole collimator is modeled on conventional pinhole by piling several tungsten layers of different apertures. Depending on the combination of the holes in each layer, a variety of hole diameters and acceptance angles of the pinhole can be made. In addition, VP SPECT system allows attaching the collimator to the object as close as possible to maximize the sensitivity and adjust the distance of the pinhole from the scintillation detector to optimize the system resolution for each rotation angle, automatically. For quantitative measurement, we compared the sensitivity and spatial resolution of VP SPECT with those of conventional pinhole SPECT. To determine the possibility of the clinical and preclinical use of proposed system, a digital mouse whole‐body (MOBY) phantom is used for simulating the live mouse model. Results: The result of simulation using ultra‐micro hot spot phantom shows that the sensitivity of the proposed VP SPECT system is about 297% of that of the conventional system. While hot rods of diameter 0.6 mm can be distinguished in the image with the proposed VP SPECT system, 1.2‐mm hot rods are barely discernible in the conventional pinhole SPECT image. According to the result of MOBY phantom simulation, heart walls separated by 3 mm were not distinguished in conventional pinhole SPECT images, but were clearly discerned in VP SPECT images. Conclusions: In this study, we designed a novel pinhole collimator for SPECT and presented preliminary results of target oriented imaging with a simulation study. Currently, we are pursuing strategies to realize the proposed system, with the goal to apply the technology into a high‐sensitivity and high‐resolution preclinical SPECT. Should VP SPECT be applied to the clinical setting, we anticipate a high‐sensitivity, high‐resolution system for applications such as heart dedicated SPECT or related fields.

[1]  N. Schramm,et al.  High-resolution SPECT using multi-pinhole collimation , 2002, IEEE Nuclear Science Symposium Conference Record.

[2]  D Paix,et al.  Pinhole imaging of gamma rays. , 1967, Physics in medicine and biology.

[3]  Heather L. Durko,et al.  High-Resolution Anamorphic SPECT Imaging , 2014, IEEE Transactions on Nuclear Science.

[4]  H. Anger SCINTILLATION CAMERA WITH MULTICHANNEL COLLIMATORS. , 1964, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[5]  Max A. Viergever,et al.  Small field-of-view dedicated cardiac SPECT systems: impact of projection truncation , 2007, European Journal of Nuclear Medicine and Molecular Imaging.

[6]  B. Hasegawa,et al.  A novel approach to multipinhole SPECT for myocardial perfusion imaging. , 2006, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[7]  Jim Patton,et al.  A Novel High-Sensitivity Rapid-Acquisition Single-Photon Cardiac Imaging Camera , 2009, Journal of Nuclear Medicine.

[8]  L. Shepp,et al.  Maximum Likelihood Reconstruction for Emission Tomography , 1983, IEEE Transactions on Medical Imaging.

[9]  H. Wieczorek,et al.  Analytical model for SPECT detector concepts , 2006, IEEE Transactions on Nuclear Science.

[10]  Benjamin M. W. Tsui,et al.  MCAT to XCAT: The Evolution of 4-D Computerized Phantoms for Imaging Research , 2009, Proceedings of the IEEE.

[11]  Yusheng Li,et al.  C-SPECT—A Clinical Cardiac SPECT/Tct Platform: Design Concepts and Performance Potential , 2009, IEEE Transactions on Nuclear Science.

[12]  Vincent Breton,et al.  GATE (geant4 application for tomographic emission): a PET/SPECT general-purpose simulation platform , 2003 .

[13]  F. Beekman,et al.  U-SPECT-II: An Ultra-High-Resolution Device for Molecular Small-Animal Imaging , 2009, Journal of Nuclear Medicine.

[14]  E. Depuey,et al.  Very low-activity stress/high-activity rest, single-day myocardial perfusion SPECT with a conventional sodium iodide camera and wide beam reconstruction processing , 2012, Journal of Nuclear Cardiology.

[15]  C. Vanhove,et al.  Review of SPECT collimator selection, optimization, and fabrication for clinical and preclinical imaging. , 2015, Medical physics.

[16]  Scott D. Metzler,et al.  Analytic determination of the resolution-equivalent effective diameter of a pinhole collimator , 2004, IEEE Transactions on Medical Imaging.

[17]  Frans van der Have,et al.  The pinhole: gateway to ultra-high-resolution three-dimensional radionuclide imaging , 2007, European Journal of Nuclear Medicine and Molecular Imaging.

[18]  S. Cherry,et al.  chapter 17 – Single Photon Emission Computed Tomography , 2012 .

[19]  C. S. Levin,et al.  First Performance Results of Ce:GAGG Scintillation Crystals With Silicon Photomultipliers , 2013, IEEE Transactions on Nuclear Science.

[20]  Y. Qi,et al.  Optimized Collimator Designs for Small Animal SPECT Imaging With a Compact Gamma Camera , 2005, 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference.

[21]  J. Zeintl,et al.  A method for improving the efficiency of myocardial perfusion imaging using conventional SPECT and SPECT/CT imaging systems , 2010, IEEE Nuclear Science Symposuim & Medical Imaging Conference.

[22]  Pinhole aperture design for /sup 131/I tumor imaging , 1996 .

[23]  Donald W. Wilson,et al.  A prototype instrument for single pinhole small animal adaptive SPECT imaging. , 2008, Medical physics.

[24]  S. Vandenberghe,et al.  Design and simulation of a full-ring multi-lofthole collimator for brain SPECT , 2013, Physics in medicine and biology.

[25]  J. Dey Improvement of Performance of Cardiac SPECT Camera Using Curved Detectors With Pinholes , 2012, IEEE Transactions on Nuclear Science.

[26]  Peter Nillius,et al.  Theoretical Bounds and System Design for Multipinhole SPECT , 2010, IEEE Transactions on Medical Imaging.

[27]  P C Hawman,et al.  The cardiofocal collimator: a variable-focus collimator for cardiac SPECT. , 1994, Physics in medicine and biology.

[28]  H. B. Barber,et al.  System integration of FastSPECT III, a dedicated SPECT rodent-brain imager based on BazookaSPECT detector technology , 2009, 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC).

[29]  R. Buechel,et al.  3 – Single-photon emission computed tomography , 2015 .

[30]  B. De Man,et al.  Distance-driven projection and backprojection in three dimensions. , 2004, Physics in medicine and biology.

[31]  Ronald J. Jaszczak,et al.  Analytic determination of pinhole collimator sensitivity with penetration , 2001, IEEE Transactions on Medical Imaging.

[32]  Matthew A. Kupinski,et al.  Adaptive SPECT , 2008, IEEE Transactions on Medical Imaging.