Temperature distribution effects on micro-CFPCR performance

Continuous flow polymerase chain reactors (CFPCRs) are BioMEMS devices that offer unique capabilities for the ultra-fast amplification of target DNA fragments using repeated thermal cycling, typically over the following temperature ranges: 90°C–95°C for denaturation, 50°C–70°C for renaturation, and 70°C–75°C for extension. In CFPCR, DNA cocktail is pumped through the constant temperature zones and reaches thermal equilibrium with the channel walls quickly due to its low thermal capacitance. In previous work, a polycarbonate CFPCR was designed with microchannels 150 μm deep, 50 μm wide, and 1.78 m long—including preheating and post-heating zones, fabricated with LIGA, and demonstrated. The high thermal resistance of the polycarbonate led to a high temperature gradient in the micro-device at steady-state and was partly responsible for the low amplification yield. Several steps were taken to ensure that there were three discrete, uniform temperature zones on the polycarbonate CFPCR device including: reducing the thickness of the CFPCR substrate to decrease thermal capacitance, using copper plates as heating elements to ensure a uniform temperature input, and making grooves between temperature zones to increase the resistance to lateral heat conduction between zones. Finite element analyses (FEA) were used to evaluate the macro temperature distribution in the CFPCR device and the micro temperature distribution along a single microchannel. At steady-state, the simulated CFPCR device had three discrete temperature zones, each with a uniform temperature distribution with a variation of ±0.3°C. An infrared (IR) camera was used to measure the steady-state temperature distribution in the prototype CFPCR and validated the simulation results. The temperature distributions along a microchannel at flow velocities from 0 mm/s to 6 mm/s were used to estimate the resulting temperatures of the DNA reagents in a single microchannel. A 500 bp DNA fragment was generated from a bacteriophage λ-DNA target using 20 cycles of PCR. The amplification efficiencies compared to a commercial thermal cycler were 72.7% (2 mm/s), 44% (3 mm/s), and 29.4% (4 mm/s). The amplification efficiency with the modified CFPCR device increased by 363% at 2 mm/s and 440% at 3 mm/s compared to amplification obtained using a CFPCR device with the same fluidic layout, (Hashimoto et al., Lab Chip 4:638, 2004) strictly due to the improved temperature distribution.

[1]  S. Senturia Microsystem Design , 2000 .

[2]  Ming Yang,et al.  Cost-effective thermal isolation techniques for use on microfabricated DNA amplification and analysis devices , 2005 .

[3]  J. El-Ali,et al.  Simulation and experimental validation of a SU-8 based PCR thermocycler chip with integrated heaters and temperature sensor , 2004 .

[4]  S. Terry,et al.  A gas chromatographic air analyzer fabricated on a silicon wafer , 1979, IEEE Transactions on Electron Devices.

[5]  A. Ricco,et al.  Peer Reviewed: Plastic Advances Microfluidic Devices , 2002 .

[6]  Christopher R. Lowe,et al.  Silicon microchambers for DNA amplification , 1998 .

[7]  J P Landers,et al.  Polymerase chain reaction in polymeric microchips: DNA amplification in less than 240 seconds. , 2001, Analytical biochemistry.

[8]  K. Myambo,et al.  DNA sequencing with Thermus aquaticus DNA polymerase and direct sequencing of polymerase chain reaction-amplified DNA. , 1988, Proceedings of the National Academy of Sciences of the United States of America.

[9]  Theodore K. Christopoulos,et al.  Continuous-flow DNA and RNA amplification chip combined with laser-induced fluorescence detection , 2003 .

[10]  Gwo-Bin Lee,et al.  Micromachined polymerase chain reaction system for multiple DNA amplification of upper respiratory tract infectious diseases. , 2005, Biosensors & bioelectronics.

[11]  Dong-Chul Han,et al.  PDMS-based micro PCR chip with Parylene coating , 2003 .

[12]  Yu Zhou,et al.  In situ hybridization to chitosan/magnetite nanocomposite induced by the magnetic field , 2006 .

[13]  Q. Hao Analytical heat-transfer modeling of multilayered microdevices , 2004 .

[14]  K. Mullis,et al.  Specific synthesis of DNA in vitro via a polymerase-catalyzed chain reaction. , 1987, Methods in enzymology.

[15]  Pin-Chuan Chen,et al.  Rapid PCR in a continuous flow device. , 2004, Lab on a chip.

[16]  Robin Fortt,et al.  Thermal optimisation of the Reimer-Tiemann reaction using thermochromic liquid crystals on a microfluidic reactor. , 2005, Lab on a chip.

[17]  D. Ehrlich,et al.  A closed-cycle capillary polymerase chain reaction machine. , 2001, Analytical chemistry.

[18]  Daniel J. Sadler,et al.  Thermal management of BioMEMS: temperature control for ceramic-based PCR and DNA detection devices , 2003 .

[19]  Frank P. Incropera,et al.  Fundamentals of Heat and Mass Transfer , 1981 .

[20]  Zhixin Li,et al.  Droplet-based micro oscillating-flow PCR chip , 2005 .

[21]  Steven A Soper,et al.  Electrokinetically synchronized polymerase chain reaction microchip fabricated in polycarbonate. , 2005, Analytical chemistry.

[22]  Ming-Yuan Huang,et al.  Simulation and experimental validation of micro polymerase chain reaction chips , 2000 .

[23]  M. Morris,et al.  Raman spectroscopic measurement of spatial and temporal temperature gradients in operating electrophoresis capillaries. , 1994, Analytical chemistry.

[24]  Pieter Telleman,et al.  Microsystem Engineering of Lab-on-a-chip devices. , 2003 .

[25]  A Manz,et al.  Chemical amplification: continuous-flow PCR on a chip. , 1998, Science.

[26]  Hye Jung Cho,et al.  Precise temperature control and rapid thermal cycling in a micromachined DNA polymerase chain reaction chip , 2002 .

[27]  Luke P. Lee,et al.  In situ thermal diagnostics of the micro-PCR system using liquid crystals , 2005 .

[28]  James S. Wilkinson,et al.  Design and theoretical evaluation of a novel microfluidic device to be used for PCR , 2003 .

[29]  Johan Roeraade,et al.  Continuous segmented-flow polymerase chain reaction for high-throughput miniaturized DNA amplification. , 2003, Analytical chemistry.

[30]  M. Morris,et al.  Spatially resolved temperature measurements in electrophoresis capillaries by Raman thermometry. , 1993, Analytical chemistry.

[31]  M. Gaitan,et al.  Temperature measurement in microfluidic systems using a temperature-dependent fluorescent dye. , 2001, Analytical chemistry.

[32]  Y. Murakami,et al.  Development of a microchamber array for picoliter PCR. , 2001, Analytical chemistry.