Theoretical framework for the dual-energy cone-beam CT noise-power spectrum, NEQ, and tasked-based detectability index

The optimization of dual-energy computed tomography (DE-CT) is challenged by the lack of a theoretical foundation for image quality. This work reports a cascaded systems analysis model that was used to derive signal and noise propagation in DE-CBCT in prevalent Fourier metrics such as the noise-power spectrum (NPS) and noise-equivalent quanta (NEQ). The model was validated in comparison to measurements of the 3D NPS and NEQ in DE-CBCT images acquired using an experimental imaging bench. Task-based detectability index was derived using DE-NPS and NEQ as an objective function in optimizing DE imaging parameters such as the dose allocation factor (DA) and kVp pair. The resulting dose allocation optimization is in agreement with the practice of assigning more dose to the high-energy image (DA < 0.5), and the model provides a quantitative basis for examining the optimal dose allocation as a function of total dose, kVp pair, the presence of electronics noise, and the imaging task. An example optimization is shown for a breast tumor detection task. Using DE decomposition to cancel fibroglandular tissue (rendering a DE-CBCT image of breast tumor against an adipose tissue background) and assuming a total dose of 15mGy, the optimal kVp pair is identified at [45, 105]kVp with DA=0.46. The model is sufficiently general for applications beyond this example, demonstrating utility in the optimization in a broad range of imaging parameters. The model provides a new, valuable framework for understanding the theoretical limits of DE-CBCT imaging performance and maximizing image quality while minimizing radiation dose.

[1]  Goran Ristic,et al.  X-ray imaging performance of structured cesium iodide scintillators. , 2004, Medical physics.

[2]  A. Macovski,et al.  Energy-selective reconstructions in X-ray computerised tomography , 1976, Physics in medicine and biology.

[3]  Jeffrey H Siewerdsen,et al.  Cascaded systems analysis of the 3D noise transfer characteristics of flat-panel cone-beam CT. , 2008, Medical physics.

[4]  Cynthia H McCollough,et al.  Image quality optimization and evaluation of linearly mixed images in dual-source, dual-energy CT. , 2009, Medical physics.

[5]  J H Siewerdsen,et al.  Generalized DQE analysis of radiographic and dual-energy imaging using flat-panel detectors. , 2005, Medical physics.

[6]  J Yorkston,et al.  Empirical and theoretical investigation of the noise performance of indirect detection, active matrix flat-panel imagers (AMFPIs) for diagnostic radiology. , 1997, Medical physics.

[7]  Grace J Gang,et al.  Analysis of Fourier-domain task-based detectability index in tomosynthesis and cone-beam CT in relation to human observer performance. , 2011, Medical physics.

[8]  D. Yang,et al.  A dedicated cone-beam CT system for musculoskeletal extremities imaging: design, optimization, and initial performance characterization. , 2011, Medical physics.

[9]  M. Karçaaltıncaba,et al.  Dual-energy CT revisited with multidetector CT: review of principles and clinical applications. , 2011, Diagnostic and interventional radiology.

[10]  C. McCollough,et al.  Quantitative imaging of element composition and mass fraction using dual-energy CT: three-material decomposition. , 2009, Medical physics.