The reliability and concurrent validity of scapular plane shoulder elevation measurements using a digital inclinometer and goniometer

This study investigated the reliability and concurrent validity of active shoulder elevation in the scapular plane (scaption) using a digital inclinometer and goniometer. Two investigators used a goniometer and digital inclinometer to measure scaption on 30 asymptomatic participants in a blinded repeated measures design. Good reliability was present with intraclass correlation coefficients (ICCs) for intrarater reliability of goniometry = 0.87, intrarater digital inclinometry = 0.88, interrater goniometry = 0.92, and interrater digital inclinometry = 0.89. The minimal detectable change (MDC95) for the interrater analysis indicated that a change equal to or greater than 8 degrees for goniometry and 9 degrees for inclinometry is required to be 95% certain that the change is not due to intertrial variability or measurement error. The concurrent validity between goniometry and digital inclinometry was excellent with an ICC value of 0.94 for both raters. The 95% limits of agreement suggest that the difference between these two measurement instruments can be expected to vary by up to ±11 degrees. The results support the interchangeable use of goniometry and digital inclinometer for measuring scaption. Clinicians and researchers should consider the MDC values presented when interpreting change during subsequent measurement sessions.

[1]  C. Taber,et al.  Taber's Cyclopedic Medical Dictionary , 1963 .

[2]  S. Azen,et al.  Reliability of goniometric measurements. , 1978, Physical therapy.

[3]  J. Fleiss,et al.  Intraclass correlations: uses in assessing rater reliability. , 1979, Psychological bulletin.

[4]  J. M. Rothstein,et al.  Goniometric Reliability in a Clinical Setting , 1983 .

[5]  M. Province,et al.  Reliability of goniometric measurements in patients with Duchenne muscular dystrophy. , 1985, Physical therapy.

[6]  Richard W. Bohannon,et al.  Clinical measurement of range of motion. Review of goniometry emphasizing reliability and validity. , 1987, Physical therapy.

[7]  J M Rothstein,et al.  Goniometric reliability in a clinical setting. Shoulder measurements. , 1987, Physical therapy.

[8]  F. Verstappen,et al.  Variability and reliability of joint measurements , 1990, The American journal of sports medicine.

[9]  W Kowsuwon,et al.  Axial Rotation Gravity Goniometer: A Simple Design of Instrument and a Controlled Reliability Study , 1990, Clinical orthopaedics and related research.

[10]  T. R. Garrett,et al.  Reliability of goniometric measurements of active arm elevation in the scapular plane obtained in a clinical setting. , 1994, Archives of physical medicine and rehabilitation.

[11]  D. Lubarsky,et al.  Goniometric assessment of shoulder range of motion: comparison of testing in supine and sitting positions. , 1998, Archives of physical medicine and rehabilitation.

[12]  R. P. Di Fabio,et al.  Physical therapy and health-related outcomes for patients with common orthopaedic diagnoses. , 1998, The Journal of orthopaedic and sports physical therapy.

[13]  Valerie Parham,et al.  Taber's Cyclopedic Medical Dictionary (Book) , 1998 .

[14]  The reproducibility of measurement of shoulder movement. , 1999, Acta orthopaedica Scandinavica.

[15]  Blackburn Ta,et al.  Rehabilitation after ligamentous and labral surgery of the shoulder: guiding concepts. , 2000 .

[16]  T. A. Blackburn,et al.  Rehabilitation after ligamentous and labral surgery of the shoulder: guiding concepts. , 2000, Journal of athletic training.

[17]  R. Buchbinder,et al.  How reliably do rheumatologists measure shoulder movement? , 2002, Annals of the rheumatic diseases.

[18]  C. Terwee,et al.  Inter-observer reproducibility of measurements of range of motion in patients with shoulder pain using a digital inclinometer , 2004, BMC musculoskeletal disorders.

[19]  J. Weir Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. , 2005, Journal of strength and conditioning research.

[20]  Intraobserver reliability of 4 physiologic movements of the shoulder in subjects with and without symptoms. , 2006, Archives of physical medicine and rehabilitation.

[21]  Alice Nieuwboer,et al.  The hemiplegic arm: Interrater reliability and concurrent validity of passive range of motion measurements , 2007, Disability and rehabilitation.

[22]  L. Higgins,et al.  Rehabilitation following reverse total shoulder arthroplasty. , 2007, The Journal of orthopaedic and sports physical therapy.