[Research on piezoelectric protein sensor array for rapid detecting HIV(1+2) antibody].
暂无分享,去创建一个
To research piezoelectric immunosensor array for rapid detecting HIV(1+2), piezoelectric immunosensor array matrix was designed. HIV(1+2)C1 antigen was immobilized onto the silver electrodes of quartz-crystal microbalance, which was modified with adsorption and cross-linked method. In the clean air flow and monitoring environment, standard quality control and clinical serum sample were detected. The linear range for the measurement of HIV(1+2) was 0.01-0.2 NCU/ml. The sensitivity, specificity and accuracy of HIV(1+2) piezoelectric protein sensor array were 91.7%, 93.3% and 92.7% respectively.