Target-tracking deliveries using conventional multileaf collimators planned with 4D direct-aperture optimization.

Respiratory motion-induced degradation of intensity-modulated radiotherapy can be corrected by the dynamic target-tracking motion of multileaf collimator equipment on a conventional linear accelerator. This paper presents a new system by which the motion of the tissue and the delivery equipment can be incorporated into the treatment optimization using a 4D direct-aperture optimization method. The program can optimize a static or dynamic delivery with respect to a 4D patient model. The individualized patient model consists of a series of discrete phases and describes changes in tissue: deformation, geometry, attenuation and scatter properties over the breathing cycle. A set of treatment apertures is matched to the respiratory phases of the motion model, and motion of the apertures between phases is constrained by the maximum leaf velocity. Plans with dynamic and static deliveries optimized on 4D patient models were compared to static plans optimized on a single phase. This investigation was carried out on a 4D digital motion phantom and repeated on a 4D patient model. The effect of motion of the static plan on the 4D phantom was evaluated by recalculating dose from all phases of the 4D model. The plan cost was evaluated as a combination of the rms spread in tumour dose from the prescribed dose and the volume of normal lung receiving doses above 10 Gy with relative weightings of 5 and 1 respectively. The motion was found to degrade the static plan by 30 +/- 4% with respect to the 3D cost function value. In contrast, the motion did not cause significant degradation to a treatment if the treatment was optimized on the 4D phantom and the cost was improved by 16 +/- 3% by optimizing with dynamic leaf tracking motion. All results are relative to the static single-phase plan. In the 4D patient model the observed tissue motion was considerably less and the measured benefit of 4D planning was consequently reduced. For the 4D patient the plan cost was not significantly changed by the tissue motion. Optimizing on 4D patient conferred an improvement of 7.5 +/- 0.3%, and the 4D plan with dynamic leaf motion improved the plan cost by 8.3 +/- 0.1%.

[1]  P. Keall 4-dimensional computed tomography imaging and treatment planning. , 2004, Seminars in radiation oncology.

[2]  S. Webb,et al.  Direct-aperture optimization applied to selection of beam orientations in intensity-modulated radiation therapy , 2007, Physics in medicine and biology.

[3]  Randall K Ten Haken,et al.  A method for incorporating organ motion due to breathing into 3D dose calculations in the liver: sensitivity to variations in motion. , 2003, Medical physics.

[4]  Hualiang Zhong,et al.  Assessment of dose reconstruction errors in image-guided radiation therapy , 2008, Physics in medicine and biology.

[5]  Lech Papiez,et al.  DMLC leaf‐pair optimal control of IMRT delivery for a moving rigid target , 2004 .

[6]  P. Evans,et al.  Effects of averaging over motion and the resulting systematic errors in radiation therapy , 2006, Physics in medicine and biology.

[7]  S Webb,et al.  The effect on IMRT conformality of elastic tissue movement and a practical suggestion for movement compensation via the modified dynamic multileaf collimator (dMLC) technique , 2005, Physics in medicine and biology.

[8]  William Paul Segars,et al.  Development of a new dynamic NURBS-based cardiac-torso (NCAT) phantom , 2001 .

[9]  Uwe Oelfke,et al.  Real-time tracking of tumor motions and deformations along the leaf travel direction with the aid of a synchronized dynamic MLC leaf sequencer , 2007, Physics in medicine and biology.

[10]  M. V. van Herk,et al.  Physical aspects of a real-time tumor-tracking system for gated radiotherapy. , 2000, International journal of radiation oncology, biology, physics.

[11]  H. Shirato,et al.  Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor. , 2000, International journal of radiation oncology, biology, physics.

[12]  Lech Papiez,et al.  Dynamic-MLC leaf control utilizing on-flight intensity calculations: a robust method for real-time IMRT delivery over moving rigid targets. , 2007, Medical physics.

[13]  J. Wong,et al.  The use of active breathing control (ABC) to reduce margin for breathing motion. , 1999, International journal of radiation oncology, biology, physics.

[14]  Steve B Jiang,et al.  Synchronized moving aperture radiation therapy (SMART): average tumour trajectory for lung patients. , 2003, Physics in medicine and biology.

[15]  Radhe Mohan,et al.  Four-dimensional radiotherapy planning for DMLC-based respiratory motion tracking. , 2005, Medical physics.

[16]  S Webb,et al.  Does elastic tissue intrafraction motion with density changes forbid motion-compensated radiotherapy? , 2006, Physics in medicine and biology.

[17]  James L Bedford,et al.  Speed versus accuracy in a fast convolution photon dose calculation for conformal radiotherapy. , 2002, Physics in medicine and biology.

[18]  R. Mohan,et al.  Motion adaptive x-ray therapy: a feasibility study , 2001, Physics in medicine and biology.

[19]  Steve B. Jiang,et al.  Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulation. , 2002, Physics in medicine and biology.

[20]  S Webb,et al.  Tracking ‘differential organ motion’ with a ‘breathing’ multileaf collimator: magnitude of problem assessed using 4D CT data and a motion-compensation strategy , 2007, Physics in medicine and biology.

[21]  Steve B. Jiang,et al.  Temporo-spatial IMRT optimization: concepts, implementation and initial results , 2005, Physics in medicine and biology.

[22]  Lech Papiez,et al.  4D DMLC leaf sequencing to minimize organ at risk dose in moving anatomy. , 2007, Medical physics.

[23]  Steve B. Jiang,et al.  Synchronized moving aperture radiation therapy (SMART): improvement of breathing pattern reproducibility using respiratory coaching , 2006, Physics in medicine and biology.

[24]  Uwe Oelfke,et al.  Linac-integrated 4D cone beam CT: first experimental results , 2006, Physics in medicine and biology.

[25]  Dinesh Tewatia,et al.  Clinical implementation of target tracking by breathing synchronized delivery. , 2006, Medical physics.

[26]  Alan Nichol,et al.  Direct aperture optimization for online adaptive radiation therapy. , 2007, Medical physics.

[27]  Lech Papiez,et al.  DMLC leaf-pair optimal control for mobile, deforming target. , 2005, Medical physics.

[28]  Surrey Sm,et al.  Motion effects in (intensity modulated) radiation therapy: a review , 2006 .

[29]  S Webb,et al.  IMRT delivery to a moving target by dynamic MLC tracking: delivery for targets moving in two dimensions in the beam's eye view , 2006, Physics in medicine and biology.

[30]  Indrin J Chetty,et al.  Dose reconstruction in deforming lung anatomy: dose grid size effects and clinical implications. , 2005, Medical physics.

[31]  J Wong,et al.  Adaptive modification of treatment planning to minimize the deleterious effects of treatment setup errors. , 1997, International journal of radiation oncology, biology, physics.

[32]  S Webb,et al.  Adapting IMRT delivery fraction-by-fraction to cater for variable intrafraction motion , 2008, Physics in medicine and biology.

[33]  J. Deasy,et al.  Beyond bixels: generalizing the optimization parameters for intensity modulated radiation therapy. , 2002, Medical physics.

[34]  Steve B. Jiang,et al.  Estimation of the delivered patient dose in lung IMRT treatment based on deformable registration of 4D-CT data and Monte Carlo simulations , 2006, Physics in medicine and biology.

[35]  J. McClelland,et al.  A continuous 4D motion model from multiple respiratory cycles for use in lung radiotherapy. , 2006, Medical physics.

[36]  Lech Papiez,et al.  Synchronized delivery of DMLC intensity modulated radiation therapy for stationary and moving targets. , 2005, Medical physics.

[37]  Sergey Povzner,et al.  Management of three-dimensional intrafraction motion through real-time DMLC tracking. , 2008, Medical physics.

[38]  Jan-Jakob Sonke,et al.  Variability of four-dimensional computed tomography patient models. , 2008, International journal of radiation oncology, biology, physics.

[39]  Radhe Mohan,et al.  Patient training in respiratory-gated radiotherapy. , 2003, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[40]  S Webb,et al.  A strategy to minimize errors from differential intrafraction organ motion using a single configuration for a ‘breathing’ multileaf collimator , 2006, Physics in medicine and biology.

[41]  H. Kubo,et al.  Respiration gated radiotherapy treatment: a technical study. , 1996, Physics in medicine and biology.