Force-based Safe Vein Cannulation in Robot-assisted Retinal Surgery: A Preliminary Study

Retinal vein cannulation (RVC) is a potential treatment for retinal vein occlusion (RVO). Manual surgery has limitations in RVC due to extremely small vessels and instruments involved, as well as the presence of physiological hand tremor. Robot-assisted retinal surgery may be a better approach to smooth and accurate instrument manipulation during this procedure. Motion of the retina and cornea related to heartbeat may be associated with unexpected forces between the tool and eyeball. In this paper, we propose a force-based control strategy to automatically compensate for the movement of the retina maintaining the tip force and sclera force in a predetermined small range. A dual force-sensing tool is used to monitor the tip force, sclera force and tool insertion depth, which will be used to derive a desired joint velocity for the robot via a modified admittance controller. Then the tool is manipulated to compensate for the movement of the retina as well as reduce the tip force and sclera force. Quantitative experiments are conducted to verify the efficacy of the control strategy and a user study is also conducted by a retinal surgeon to demonstrate the advantages of our automatic compensation approach.

[1]  Yang Yang,et al.  Toward Safe Retinal Microsurgery: Development and Evaluation of an RNN-Based Active Interventional Control Framework , 2020, IEEE Transactions on Biomedical Engineering.

[2]  J. Weiss Treatment of central retinal vein occlusion by injection of tissue plasminogen activator into a retinal vein. , 1998, American journal of ophthalmology.

[3]  Iulian Iordachita,et al.  Towards Bimanual Vein Cannulation: Preliminary Study of a Bimanual Robotic System With a Dual Force Constraint Controller , 2020, 2020 IEEE International Conference on Robotics and Automation (ICRA).

[4]  Russell H. Taylor,et al.  New steady-hand Eye Robot with micro-force sensing for vitreoretinal surgery , 2010, 2010 3rd IEEE RAS & EMBS International Conference on Biomedical Robotics and Biomechatronics.

[5]  Russell H. Taylor,et al.  Toward Improving Patient Safety and Surgeon Comfort in a Synergic Robot-Assisted Eye Surgery: A Comparative Study , 2019, 2019 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS).

[6]  Neil M Bressler,et al.  Reading speed improvements in retinal vein occlusion after ranibizumab treatment. , 2013, JAMA ophthalmology.

[7]  Tien Yin Wong,et al.  The prevalence of retinal vein occlusion: pooled data from population studies from the United States, Europe, Asia, and Australia. , 2010, Ophthalmology.

[8]  Jin U. Kang,et al.  Robotic Retinal Surgery , 2020, Handbook of Robotic and Image-Guided Surgery.

[9]  Dominiek Reynaerts,et al.  In-Human Robot-Assisted Retinal Vein Cannulation, A World First , 2018, Annals of Biomedical Engineering.

[10]  Cameron N. Riviere,et al.  Physiological tremor amplitude during retinal microsurgery , 2002, Proceedings of the IEEE 28th Annual Northeast Bioengineering Conference (IEEE Cat. No.02CH37342).

[11]  Iulian Iordachita,et al.  Real-Time Sclera Force Feedback for Enabling Safe Robot-Assisted Vitreoretinal Surgery , 2018, 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).

[12]  M Maarten Steinbuch,et al.  Robot assistance for micrometer precision in vitreoretinal surgery , 2013 .

[13]  Cameron N. Riviere,et al.  Micron: An Actively Stabilized Handheld Tool for Microsurgery , 2012, IEEE Transactions on Robotics.

[14]  Russell H. Taylor,et al.  Auditory force feedback substitution improves surgical precision during simulated ophthalmic surgery. , 2013, Investigative ophthalmology & visual science.

[15]  Santiago Costantino,et al.  Measurement of ocular fundus pulsation in healthy subjects using a novel Fourier-domain optical coherence tomography. , 2011, Investigative ophthalmology & visual science.

[16]  Russell H. Taylor,et al.  A multi-function force sensing instrument for variable admittance robot control in retinal microsurgery , 2014, 2014 IEEE International Conference on Robotics and Automation (ICRA).

[17]  Marin Kobilarov,et al.  Adaptive Control of Sclera Force and Insertion Depth for Safe Robot-Assisted Retinal Surgery , 2019, 2019 International Conference on Robotics and Automation (ICRA).

[18]  Alois Knoll,et al.  The introduction of a new robot for assistance in ophthalmic surgery , 2013, 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).

[19]  Cameron N. Riviere,et al.  Manipulator Design and Operation of a Six-Degree-of-Freedom Handheld Tremor-Canceling Microsurgical Instrument , 2015, IEEE/ASME Transactions on Mechatronics.

[20]  Alois Knoll,et al.  Kinematics and dynamics analysis of a hybrid parallel-serial micromanipulator designed for biomedical applications , 2013, 2013 IEEE/ASME International Conference on Advanced Intelligent Mechatronics.

[21]  Russell H. Taylor,et al.  Toward Clinically Applicable Steady-Hand Eye Robot for Vitreoretinal Surgery , 2012 .

[22]  T. Rim,et al.  Retinal vessel structure measurement using spectral-domain optical coherence tomography , 2016, Eye.

[23]  A. Guerrouad,et al.  SMOS: stereotaxical microtelemanipulator for ocular surgery , 1989, Images of the Twenty-First Century. Proceedings of the Annual International Engineering in Medicine and Biology Society,.