Monochromatic breast computed tomography with synchrotron radiation: phase-contrast and phase-retrieved image comparison and full-volume reconstruction.

A program devoted to performing the first in vivo synchrotron radiation (SR) breast computed tomography (BCT) is ongoing at the Elettra facility. Using the high spatial coherence of SR, phase-contrast (PhC) imaging techniques can be used. The latest high-resolution BCT acquisitions of breast specimens, obtained with the propagation-based PhC approach, are herein presented as part of the SYRMA-3D collaboration effort toward the clinical exam. Images are acquired with a 60 - μ m   pixel dead-time-free single-photon-counting CdTe detector. The samples are imaged at 32 and 38 keV in a continuous rotating mode, delivering 5 to 20 mGy of mean glandular dose. Contrast-to-noise ratio (CNR) and spatial resolution performances are evaluated for both PhC and phase-retrieved images, showing that by applying the phase-retrieval algorithm a five-time CNR increase can be obtained with a minor loss in spatial resolution across soft tissue interfaces. It is shown that, despite having a poorer CNR, PhC images can provide a sharper visualization of microcalcifications, thus being complementary to phase-retrieved images. Furthermore, the first full-volume scan of a mastectomy sample ( 9 × 9 × 3    cm 3 ) is reported. This investigation into surgical specimens indicates that SR BCT in terms of CNR, spatial resolution, scan duration, and scan volume is feasible.

[1]  Martin P Tornai,et al.  Evaluation of the absorbed dose to the breast using radiochromic film in a dedicated CT mammotomography system employing a quasi-monochromatic x-ray beam. , 2011, Medical physics.

[2]  D J Crotty,et al.  Experimental spectral measurements of heavy K-edge filtered beams for x-ray computed mammotomography. , 2007, Physics in medicine and biology.

[3]  D. Dreossi,et al.  Measurement of the linear attenuation coefficients of breast tissues by synchrotron radiation computed tomography , 2010, Physics in medicine and biology.

[4]  Giovanni Mettivier,et al.  Imaging study of a phase-sensitive breast-CT system in continuous acquisition mode , 2017 .

[5]  V Fanti,et al.  Towards breast tomography with synchrotron radiation at Elettra: first images , 2016, Physics in medicine and biology.

[6]  D. Paganin,et al.  2D and 3D X-ray phase retrieval of multi-material objects using a single defocus distance. , 2010, Optics express.

[7]  Luigi Rigon,et al.  Clinical study in phase- contrast mammography: image-quality analysis , 2014, Philosophical Transactions of the Royal Society A: Mathematical, Physical and Engineering Sciences.

[8]  Piernicola Oliva,et al.  Optimization of the equalization procedure for a single-photon counting CdTe detector used for CT , 2017 .

[9]  Michaela C. C. Weigel,et al.  High-resolution spiral CT of the breast at very low dose: concept and feasibility considerations , 2011, European Radiology.

[10]  G Tromba,et al.  Glandular dose in breast computed tomography with synchrotron radiation , 2016, Physics in medicine and biology.

[11]  D. Dreossi,et al.  The effects of the imaging system on the validity limits of the ray-optical approach to phase contrast imaging. , 2005, Medical physics.

[12]  Franz Pfeiffer,et al.  Visualizing Typical Features of Breast Fibroadenomas Using Phase-Contrast CT: An Ex-Vivo Study , 2014, PloS one.

[13]  Piernicola Oliva,et al.  Characterization of Pixirad-1 photon counting detector for X-ray imaging , 2016 .

[14]  P. Cloetens,et al.  Phase objects in synchrotron radiation hard x-ray imaging , 1996 .

[15]  Anna Burvall,et al.  Phase Retrieval in X-ray Phase-contrast Imaging Suitable for Tomography , 2022 .

[16]  Yakov I Nesterets,et al.  Some simple rules for contrast, signal-to-noise and resolution in in-line x-ray phase-contrast imaging. , 2008, Optics express.

[17]  F Longo,et al.  GEANT4 for breast dosimetry: parameters optimization study , 2015, Physics in medicine and biology.

[18]  I. Sechopoulos A review of breast tomosynthesis. Part I. The image acquisition process. , 2013, Medical physics.

[19]  S. Wilkins,et al.  Simultaneous phase and amplitude extraction from a single defocused image of a homogeneous object , 2002, Journal of microscopy.

[20]  John M Boone,et al.  Technique factors and their relationship to radiation dose in pendant geometry breast CT. , 2005, Medical physics.

[21]  Akira Furukawa,et al.  The First Trial of Phase Contrast Imaging for Digital Full-Field Mammography Using a Practical Molybdenum X-Ray Tube , 2005, Investigative radiology.

[22]  I. Sechopoulos A review of breast tomosynthesis. Part II. Image reconstruction, processing and analysis, and advanced applications. , 2013, Medical physics.

[23]  Emilio Quaia,et al.  The SYRMEP Beamline of Elettra: Clinical Mammography and Bio‐medical Applications , 2010 .

[24]  Luca Brombal,et al.  A Proposal for a Quality Control Protocol in Breast CT with Synchrotron Radiation , 2018, Radiology and oncology.

[25]  P. C. Johns,et al.  X-ray characterisation of normal and neoplastic breast tissues. , 1987, Physics in medicine and biology.

[26]  Timur E. Gureyev,et al.  Noise propagation in x-ray phase-contrast imaging and computed tomography , 2014 .

[27]  M McCormack,et al.  Optimization of propagation-based x-ray phase-contrast tomography for breast cancer imaging , 2017, Physics in medicine and biology.

[28]  Piernicola Oliva,et al.  Large-area single-photon-counting CdTe detector for synchrotron radiation computed tomography: a dedicated pre-processing procedure. , 2018, Journal of synchrotron radiation.

[29]  Franz Pfeiffer,et al.  Phase-Contrast Hounsfield Units of Fixated and Non-Fixated Soft-Tissue Samples , 2015, PloS one.

[30]  Ruola Ning,et al.  Cone beam breast CT with multiplanar and three dimensional visualization in differentiating breast masses compared with mammography. , 2015, European journal of radiology.

[31]  Giovanni Mettivier,et al.  Dedicated breast computed tomography: Basic aspects. , 2015, Medical physics.

[32]  Roberto Pugliese,et al.  Enhanced and Flexible Software Tools for X-ray Computed Tomography at the Italian Synchrotron Radiation Facility Elettra , 2015, Fundam. Informaticae.

[33]  Emilio Quaia,et al.  Mammography with synchrotron radiation: first clinical experience with phase-detection technique. , 2011, Radiology.

[34]  Ehsan Samei,et al.  Simulation study of a quasi-monochromatic beam for x-ray computed mammotomography. , 2004, Medical physics.

[35]  K K W Siu,et al.  Interface-specific x-ray phase retrieval tomography of complex biological organs , 2011, Physics in medicine and biology.

[36]  Ioannis Sechopoulos,et al.  Clinical digital breast tomosynthesis system: dosimetric characterization. , 2012, Radiology.

[37]  R Longo,et al.  Imaging performance of phase-contrast breast computed tomography with synchrotron radiation and a CdTe photon-counting detector. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[38]  Paola Coan,et al.  X-ray phase-contrast imaging: from pre-clinical applications towards clinics , 2013, Physics in medicine and biology.

[39]  Samta Thacker,et al.  Evaluating the impact of X-ray spectral shape on image quality in flat-panel CT breast imaging. , 2007, Medical physics.

[40]  Fulvia Arfelli,et al.  Dose and diagnostic performance comparison between phase-contrast mammography with synchrotron radiation and digital mammography: a clinical study report , 2018, Journal of medical imaging.

[41]  Luca Brombal,et al.  Phase-Contrast Breast-CT: Optimization of Experimental Parameters and Reconstruction Algorithms , 2018, IFMBE Proceedings.

[42]  T. Gureyev,et al.  On the "unreasonable" effectiveness of transport of intensity imaging and optical deconvolution. , 2017, Journal of the Optical Society of America. A, Optics, image science, and vision.

[43]  R. Bellazzini,et al.  Chromatic X-ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC , 2013 .

[44]  Martin P. Tornai,et al.  Development and Initial Demonstration of a Low-Dose Dedicated Fully 3D Breast CT System , 2012, Digital Mammography / IWDM.