Computational and experimental evaluation of the fluid dynamics and hemocompatibility of the CentriMag blood pump.

The CentriMag centrifugal blood pump is a newly developed ventricular assist device based on magnetically levitated bearingless rotor technology. A combined computational and experimental study was conducted to characterize the hemodynamic and hemocompatibility performances of this novel blood pump. Both the three-dimensional flow features of the CentriMag blood pump and its hemolytic characteristics were analyzed using computational fluid dynamics (CFD)-based modeling. The hydraulic pump performance and hemolysis level were quantified experimentally. The CFD simulation demonstrated a clean and streamlined flow field in the main components of the CentriMag blood pump. The predicted results by hemolysis model indicated no significant high shear stress regions in the pump. A comparison of CFD predictions and experimental results showed good agreements. The relatively large gap passages (1.5 mm) between the outer rotor walls and the lower housing cavity walls provide a very good surface washing through a secondary flow path while the shear stresses in the secondary flow paths are reduced, resulting in a low rate of hemolysis ([Normalized Index of Hemolysis] NIH = 0.0029 +/- 0.006) without a decrease of the pump's hydrodynamic performance (pressure head: 352 mm Hg at a flow rate of 5.0 L/min and a rotational speed of 4,000 rpm).

[1]  M. Lachat,et al.  The CentriMag: a new optimized centrifugal blood pump with levitating impeller. , 2004, The heart surgery forum.

[2]  Natalie L James,et al.  Evaluation of hemolysis in the VentrAssist implantable rotary blood pump. , 2003, Artificial organs.

[3]  J F Antaki,et al.  Fluid dynamic characterization of operating conditions for continuous flow blood pumps. , 1999, ASAIO journal.

[4]  Reto Schöb Centrifugal pump without bearings or seals , 2002 .

[5]  M. Sato [Mechanical properties of living tissues]. , 1986, Iyo denshi to seitai kogaku. Japanese journal of medical electronics and biological engineering.

[6]  C Bludszuweit,et al.  Three-dimensional numerical prediction of stress loading of blood particles in a centrifugal pump. , 1995, Artificial organs.

[7]  Young I. Cho,et al.  A method of isolating surface tension and yield stress effects in a U-shaped scanning capillary-tube viscometer using a Casson model , 2002 .

[8]  Xinwei Song,et al.  A prototype HeartQuest ventricular assist device for particle image velocimetry measurements. , 2002, Artificial organs.

[9]  André Garon,et al.  Fast three-dimensional numerical hemolysis approximation. , 2004, Artificial organs.

[10]  R. Malinauskas,et al.  Plasma hemoglobin measurement techniques for the in vitro evaluation of blood damage caused by medical devices. , 2008, Artificial organs.

[11]  H. Reul,et al.  Estimation of Shear Stress-related Blood Damage in Heart Valve Prostheses - in Vitro Comparison of 25 Aortic Valves , 1990, The International journal of artificial organs.

[12]  Reto Schöb,et al.  Computational fluid dynamics analysis of a maglev centrifugal left ventricular assist device. , 2004, Artificial organs.

[13]  Y. Fung,et al.  Biomechanics: Mechanical Properties of Living Tissues , 1981 .