BACKGROUND
The clinical relevance of elevated serum markers of hemolysis during left ventricular assist device (LVAD) support has not been fully ascertained.
METHODS
Lactate dehydrogenase (LDH) and serum free hemoglobin (sfHg) values were tallied monthly in 182 patients on HeartMate II (Thoratec, Pleasanton, CA) LVAD support. Peak values for each marker were identified, and 2 hemolysis definitions were applied to the cohort: Hemolysis according to Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) criteria (sfHg > 40 mg/dl with signs/symptoms) and/or hemolysis defined by an LDH ≥ 600 IU/liter (2.5-times the upper limit of laboratory normal). Kaplan-Meier survival free from death, urgent United Network of Organ Sharing 1A transplant for thrombosis, device exchange for thrombosis, and stroke/peripheral embolism was estimated, and Cox hazard ratios (HR) with the 95% confidence interval (95% CI) were calculated. Areas under the receiver-operating characteristic curves (AUCs) for predicting 1-year event-free survival were calculated.
RESULTS
Hemolysis occurred in 32 patients (18%) by INTERMACS criteria and in 68 (37%) patients by LDH criteria. Over a median (25(th), 75(th)) support of 427 days (245, 793 days), there were 78 events. One year event-free survival after the onset of INTERMACS-defined hemolysis was 16% ± 8.3% compared with 85% ± 3.2% in non-hemolyzers (HR, 14.7; 95% CI, 7.9-27; AUC 0.70 ± 0.05; p < 0.001; ). One year event-free survival after the onset of LDH-defined hemolysis was 32% ± 7.2% compared with 89% ± 3.2% in those with persistent LDH values < 600 IU/liter (HR, 8.0; 95% CI, 4.4-14; AUC 0.87 ± 0.04; p < 0.001). Patients who met the LDH hemolysis definition had longer times from hemolysis onset to clinical events and larger magnitudes of risk for embolism and device exchange for thrombosis than those with INTERMACS hemolysis.
CONCLUSIONS
Serum hemolysis marker elevations are associated with increased events in LVAD patients. LDH monitoring provides an earlier diagnosis of adverse events than sfHg, supporting need for a new INTERMACS definition of VAD-associated hemolysis.
[1]
V. Mehta,et al.
Diagnosis of hemolysis and device thrombosis with lactate dehydrogenase during left ventricular assist device support.
,
2014,
The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.
[2]
S. Russell,et al.
Advanced heart failure treated with continuous-flow left ventricular assist device.
,
2009,
The New England journal of medicine.
[3]
Nader Moazami,et al.
Extended mechanical circulatory support with a continuous-flow rotary left ventricular assist device.
,
2009,
Journal of the American College of Cardiology.
[4]
J. Hanley,et al.
A method of comparing the areas under receiver operating characteristic curves derived from the same cases.
,
1983,
Radiology.