Cementless Total Hip Arthroplasty with a Tapered, Rectangular Titanium Stem and a Threaded Cup: A Minimum Ten-Year Follow-up

Background: We report the results of cementless total hip arthroplasty with a tapered, rectangular titanium stem that was introduced in 1979 and continues to be used today with only minor changes. The aim of the design is to achieve primary stability to resist rotational and axial forces through precision rasping and press-fit implantation of a tapered, rectangular femoral component.Methods: Between October 1986 and November 1987, 208 total hip arthroplasties with insertion of a tapered, rectangular titanium stem and a threaded cup without cement were performed in 200 consecutive patients (average age, sixty-one years; range, twenty-two to eighty-four years).Results: At the time of the latest follow-up, fifty-one patients (fifty-two hips) had died and sixteen patients had been lost to follow-up, leaving 133 patients. Twelve hips had been revised, two in patients who subsequently died, leaving 123 living patients without revision. The median follow-up time was 120.7 months. Five cups needed revision surgery because of aseptic loosening; two, because of massive polyethylene wear; one, because of posttraumatic migration; and one, because of breakage. Three femoral stems were revised: one because of malpositioning (the reoperation was done five days after implantation); one, because of infection; and the third, after multiple failed acetabular revisions. The mean Harris hip score for the patients who did not have revision was 85.4 points (range, 46 to 100 points) at the time of the latest follow-up. Four patients (3%) complained of thigh pain that was not associated with another disorder. According to the criteria of Engh et al., all femoral implants were graded as stable bone-ingrown. The probability of survival of both the femoral and the acetabular component at ten years, with any revision as the end point, was 0.92 (95% confidence interval, 0.88 to 0.97). The probability of survival of the cup was 0.93 (95% confidence interval, 0.89 to 0.97), and that of the stem was 0.99 (95% confidence interval, 0.97 to 1.00).Conclusions: The results of arthroplasty with a tapered, rectangular titanium stem combined with a conical threaded cup inserted without cement were excellent at a minimum of ten years. Our data suggest that femoral stem fixation continues to be secure, while the threaded cup is prone to aseptic loosening.

[1]  W. Harris,et al.  Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. , 1969, The Journal of bone and joint surgery. American volume.

[2]  C. Engh,et al.  Total Hip Arthroplasty with Porous-Coated Prostheses Fixed without Cement in Patients Who Are Sixty-five Years of Age or Older* , 1998, The Journal of bone and joint surgery. American volume.

[3]  F. Kerschbaumer,et al.  The Transgluteal Approach to the Hip Joint , 1979, Archives of orthopaedic and traumatic surgery.

[4]  R. Barrack,et al.  Thigh pain despite bone ingrowth into uncemented femoral stems. , 1992, The Journal of bone and joint surgery. British volume.

[5]  B. Love A femoral component inserted without cement in total hip arthroplasty. A study of the Tri-Lock component with an average ten-year duration of follow-up. , 1999, The Journal of bone and joint surgery. American volume.

[6]  J. Charnley,et al.  Radiological demarcation of cemented sockets in total hip replacement. , 1976, Clinical orthopaedics and related research.

[7]  R. Streicher,et al.  Joint replacement components made of hot-forged and surface-treated Ti-6Al-7Nb alloy. , 1992, Biomaterials.

[8]  M Schemper,et al.  A note on quantifying follow-up in studies of failure time. , 1996, Controlled clinical trials.

[9]  C. Engh,et al.  Long-Term Results of Use of the Anatomic Medullary Locking Prosthesis in Total Hip Arthroplasty* , 1997, The Journal of bone and joint surgery. American volume.

[10]  N. Mantel Evaluation of survival data and two new rank order statistics arising in its consideration. , 1966, Cancer chemotherapy reports.

[11]  C. Engh,et al.  Results of cementless revision for failed cemented total hip arthroplasty. , 1988, Clinical orthopaedics and related research.

[12]  David R. Cox,et al.  Regression models and life tables (with discussion , 1972 .

[13]  R Poss,et al.  Clinical and radiographic evaluation of total hip replacement. A standard system of terminology for reporting results. , 1990, The Journal of bone and joint surgery. American volume.

[14]  C. Engh,et al.  Femoral fracture during non-cemented total hip arthroplasty. , 1989, The Journal of bone and joint surgery. American volume.

[15]  J. Charnley,et al.  Rate of wear in total hip replacement. , 1975, Clinical orthopaedics and related research.

[16]  L. Riley,et al.  Ectopic ossification following total hip replacement. Incidence and a method of classification. , 1973, The Journal of bone and joint surgery. American volume.

[17]  R Deutman,et al.  [Total hip arthroplasty]. , 1977, Tijdschrift voor ziekenverpleging.

[18]  G. Böhm,et al.  Reactions of surrounding tissue to the cementless hip implant Ti-6Al-4V after an implantation period of several years , 1988, Archives of orthopaedic and traumatic surgery.

[19]  A. Kapandji,et al.  Cementless primary total hip replacement. Four to eight year results with the Zweymüller-Alloclassic prosthesis. , 1998, International orthopaedics.

[20]  F. Lintner,et al.  Biologic fixation of a press-fit titanium hip joint endoprosthesis. , 1988, Clinical orthopaedics and related research.

[21]  B. Espehaug,et al.  Early aseptic loosening of uncemented femoral components in primary total hip replacement. A review based on the Norwegian Arthroplasty Register. , 1995, The Journal of bone and joint surgery. British volume.

[22]  B. Morrey,et al.  Effect of femoral head size on wear of the polyethylene acetabular component. , 1990, The Journal of bone and joint surgery. American volume.

[23]  P. Zenz,et al.  [10 years of cementless implantation of total hip endoprosthesis using Zweymüller's stem]. , 1995, Zeitschrift für Orthopädie und ihre Grenzgebiete.

[24]  W. Capello,et al.  Omnifit cementless total hip arthroplasty. A 10-year average followup. , 1999, Clinical orthopaedics and related research.

[25]  J. McLaughlin,et al.  Total hip arthroplasty with an uncemented femoral component. Excellent results at ten-year follow-up. , 1997, The Journal of bone and joint surgery. British volume.

[26]  H. Amstutz,et al.  "Modes of failure" of cemented stem-type femoral components: a radiographic analysis of loosening. , 1979, Clinical orthopaedics and related research.

[27]  Total Hip Replacements Using the Ceramic Mittelmeier Prosthesis , 1996, Clinical orthopaedics and related research.

[28]  E. Kaplan,et al.  Nonparametric Estimation from Incomplete Observations , 1958 .

[29]  F Lintner,et al.  Tissue reactions to titanium endoprostheses. Autopsy studies in four cases. , 1986, The Journal of arthroplasty.

[30]  A C Campbell,et al.  Thigh pain after cementless hip arthroplasty. Annoyance or ill omen. , 1992, The Journal of bone and joint surgery. British volume.

[31]  M. Semlitsch,et al.  Concept and material properties of a cementless hip prosthesis system with Al2O3 ceramic ball heads and wrought Ti-6Al-4V Stems , 2004, Archives of orthopaedic and traumatic surgery.

[32]  C. Engh,et al.  Patterns of osteolysis around total hip components inserted with and without cement. , 1995, The Journal of bone and joint surgery. American volume.