Thermodynamic Temperature Amplification by Means of Two Coupled Gas‐Controlled Heat‐Pipes
暂无分享,去创建一个
Two gas‐controlled heat‐pipes, one working with mercury and the other with sodium, have been connected to the same helium pressure line. At any given pressure, the high temperature of the sodium heat‐pipe is thermodynamically related to the low temperature of the mercury heat‐pipe. Pressure control well within 1 ppm is obtained in both devices by using a 25 Ω SPRT inserted in the thermometer well of the mercury heat‐pipe. A computer controls the helium pressure by operating input/output electrovalves and a variable volume. With such a system, helium pressures between 8 kPa and 200 kPa are controlled with a sensitivity well beyond the resolution of currently available pressure gauges, thus reaching a temperature stability in the mercury heat‐pipe at a level of about 0.2 mK. The very high reproducibility of the 25 Ω SPRT used in the mercury heat‐pipe between 240 °C and 400 °C, within 2⋅10−7, is transferred with this amplification system to higher temperatures, thus obtaining in the sodium heat‐pipe between ...