An algorithm for PET tumor volume and activity quantification: without specifying camera's point spread function (PSF).

PURPOSE The authors have developed an algorithm for segmentation and removal of the partial volume effect (PVE) of tumors in positron emission tomography (PET) images. The algorithm accurately measures functional volume (FV) and activity concentration (AC) of tumors independent of the camera's full width half maximum (FWHM). METHODS A novel iterative histogram thresholding (HT) algorithm is developed to segment the tumors in PET images, which have low resolution and suffer from inherent noise in the image. The algorithm is initiated by manually drawing a region of interest (ROI). The segmented tumors are subjected to the iterative deconvolution thresholding segmentation (IDTS) algorithm, where the Van-Cittert's method of deconvolution is used for correcting PVE. The IDTS algorithm is fully automated and accurately measures the FV and AC, and stops once it reaches convergence. The convergence criteria or stopping conditions are developed in such a way that the algorithm does not rely on estimating the FWHM of the point spread function (PSF) to perform the deconvolution process. The algorithm described here was tested in phantom studies, where hollow spheres (0.5-16 ml) were used to represent tumors with a homogeneous activity distribution, and an irregular shaped volume was used to represent a tumor with a heterogeneous activity distribution. The phantom studies were performed with different signal to background ratios (SBR) and with different acquisition times (1 min, 3 min, and 5 min). The parameters in the algorithm were also changed (FWHM and matrix size of the Gaussian function) to check the accuracy of the algorithm. Simulated data were also used to test the algorithm with tumors having heterogeneous activity distribution. RESULTS The results show that changing the size and shape of the ROI during initiation of the algorithm had no significant impact on the FV. An average FV overestimation of 30% and an average AC underestimation of 35% were observed for the smallest tumor (0.5 ml) over the entire range of noise and SBR level. The difference in average FV and AC estimations from the actual volumes were less than 5% as the tumor size increased to 16 ml. For tumors with heterogeneous activity profile, the overall volume error was less than 10%. The average overestimation of FV was less than 10% and classification error was around 11%. CONCLUSIONS The algorithm developed herein was extensively tested and is not dependent on accurately quantifying the camera's PSF. This feature demonstrates the robustness of the algorithm and enables it to be applied on a wide range of noise and SBR within an image. The ultimate goal of the algorithm is to be able to be operated independent of the camera type used and the reconstruction algorithm deployed.

[1]  S. Nehmeh,et al.  An iterative technique to segment PET lesions using a Monte Carlo based mathematical model. , 2009, Medical physics.

[2]  S. Gulec,et al.  Safety and efficacy of Y-90 microsphere treatment in patients with primary and metastatic liver cancer: The tumor selectivity of the treatment as a function of tumor to liver flow ratio , 2007, Journal of Translational Medicine.

[3]  Yoshimi Anzai,et al.  FDG‐PET/CT–guided intensity modulated head and neck radiotherapy: A pilot investigation , 2005, Head & neck.

[4]  I. Buvat,et al.  Partial-Volume Effect in PET Tumor Imaging* , 2007, Journal of Nuclear Medicine.

[5]  Anne Bol,et al.  A gradient-based method for segmenting FDG-PET images: methodology and validation , 2007, European Journal of Nuclear Medicine and Molecular Imaging.

[6]  Christian Roux,et al.  A Fuzzy Locally Adaptive Bayesian Segmentation Approach for Volume Determination in PET , 2009, IEEE Transactions on Medical Imaging.

[7]  A. Kirov,et al.  Partial volume effect correction in PET using regularized iterative deconvolution with variance control based on local topology , 2008, Physics in medicine and biology.

[8]  Robert Jeraj,et al.  A method for partial volume correction of PET-imaged tumor heterogeneity using expectation maximization with a spatially varying point spread function , 2010, Physics in medicine and biology.

[9]  Habib Zaidi,et al.  A novel fuzzy C-means algorithm for unsupervised heterogeneous tumor quantification in PET. , 2010, Medical physics.

[10]  Andreas Bockisch,et al.  Segmentation of PET volumes by iterative image thresholding. , 2007, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[11]  Giuseppe Baselli,et al.  Lesion quantification in oncological positron emission tomography: a maximum likelihood partial volume correction strategy. , 2009, Medical physics.

[12]  P. Flamen,et al.  Multimodality imaging can predict the metabolic response of unresectable colorectal liver metastases to radioembolization therapy with Yttrium-90 labeled resin microspheres , 2008, Physics in medicine and biology.

[13]  Dimitris Visvikis,et al.  Accurate automatic delineation of heterogeneous functional volumes in positron emission tomography for oncology applications. , 2010, International journal of radiation oncology, biology, physics.

[14]  S. Libutti,et al.  Partial-Volume Correction in PET: Validation of an Iterative Postreconstruction Method with Phantom and Patient Data , 2007, Journal of Nuclear Medicine.

[15]  David Dagan Feng,et al.  Segmentation of VOI From Multidimensional Dynamic PET Images by Integrating Spatial and Temporal Features , 2006, IEEE Transactions on Information Technology in Biomedicine.

[16]  W. Roa,et al.  Iterative threshold segmentation for PET target volume delineation. , 2007, Medical physics.

[17]  J. Hirvonen,et al.  Comparison of 2D and 3D performance for FDG PET with different acquisition times in oncological patients , 2009, Nuclear medicine communications.

[18]  Benjamin M. W. Tsui,et al.  Modeling respiratory mechanics in the MCAT and spline-based MCAT phantoms , 1999 .

[19]  Daniel A Low,et al.  A novel PET tumor delineation method based on adaptive region-growing and dual-front active contours. , 2008, Medical physics.