The presence of air bubbles in the aorta of a patient with a HeartMate II left ventricular assist device: a novel sign of outflow graft obstruction.

Microbubbles formed by cavitation have been frequently reported in patients with mechanical heart valves, but have not previously been described in patients with continuous-flow left-ventricular assist devices (LVADs). Obstruction of the LVAD circuit by thrombus, outflow graft kinking, or cannula malposition may cause high-pressure gradients and turbulent flow, often resulting in hemolysis. The article is the first in the field to describe the presence of microbubbles in the ascending aorta of a patient with severe LVAD-associated hemolysis. It is proposed that these bubbles were formed by cavitation in the presence of a high-pressure gradient due to LVAD outflow graft obstruction and suggested that this may be a novel imaging sign of LVAD obstruction in patients with continuous-flow LVADs.

[1]  J. Elefteriades,et al.  Avoiding technical pitfalls in left ventricular assist device placement. , 2011, Cardiology clinics.

[2]  D. Burstow,et al.  Systemic air embolization originating from a pleural air leak via a left ventricular assist device cannula anastomosis site. , 2010, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[3]  S. Raman,et al.  Noninvasive assessment of left ventricular assist devices with cardiovascular computed tomography and impact on management. , 2010, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[4]  T. Myers,et al.  Management of air embolism during HeartMate XVE exchange. , 2007, Texas Heart Institute journal.

[5]  L. Kappenberger,et al.  Cavitation versus Degassing: In Vitro Study of the Microbubble Phenomenon Observed During Echocardiography in Patients with Mechanical Prosthetic Cardiac Valves , 2002, Echocardiography.

[6]  M. Elkind,et al.  Massive air embolism with left ventricular assist device. , 2002, Neurology.

[7]  R. Bolman,et al.  Transcatheter closure of patent foramen ovale for hypoxemia during left ventricular assist device support. , 1999, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[8]  G Rosenberg,et al.  In vivo observation of cavitation on prosthetic heart valves. , 1996, ASAIO journal.