Residual motion of lung tumors in end-of-inhale respiratory gated radiotherapy based on external surrogates.

It has been noted that some lung tumors exhibit large periodic motion due to respiration. To limit the amount of dose to healthy lung tissues, many clinics have begun gating radiotherapy treatment using externally placed surrogates. It has been observed by several institutions that the end-of-exhale (EOE) tumor position is more reproducible than other phases of the breathing cycle, so the gating window is often set there. From a treatment planning perspective, end-of-inhale (EOI) phase might be preferred for gating because the expanded lungs will further decrease the healthy tissue within the treatment field. We simulate gated treatment at the EOI phase, using a set of recently measured internal/external anatomy patient data. This paper attempts to answer three questions: (1) How much is the tumor residual motion when we use an external surrogate gating window at EOI? (2) How could we reduce the residual motion in the EOI gating window? (3) Is there a preference for amplitude- versus phase-based gating at EOI? We found that under free breathing conditions the residual motion of the tumors is much larger for EOI phase than for EOE phase. The mean values of residual motion at EOI were found to be 2.2 and 2.7 mm for amplitude- and phase-based gating, respectively, and, at EOE, 1.0 and 1.2 mm for amplitude- and phase-based gating, respectively. However, we note that the residual motion in the EOI gating window is correlated well with the reproducibility of the external surface position in the EOI phase. Using the results of a published breath-coaching study, we deduce that the residual motion of a lung tumor at EOI would approach that at EOE, with the same duty cycle (30%), under breath-coaching conditions. Additionally, we found that under these same conditions, phase-based gating approaches the same residual motion as amplitude-based gating, going from a 28% difference to 11%, for the patient with the largest difference between the two gating modalities. We conclude that it is feasible to achieve the same reproducibility of tumor location at EOI as at EOE if breath coaching is implemented, enabling us to reap the benefits of the dosimetric advantage of EOI gating.

[1]  G J Kutcher,et al.  Deep inspiration breath-hold technique for lung tumors: the potential value of target immobilization and reduced lung density in dose escalation. , 1999, International journal of radiation oncology, biology, physics.

[2]  Tatsuya Fujisaki,et al.  Effects of density changes in the chest on lung stereotactic radiotherapy. , 2004, Radiation medicine.

[3]  Steve B Jiang,et al.  A technique for respiratory-gated radiotherapy treatment verification with an EPID in cine mode , 2005 .

[4]  Steve B Jiang,et al.  Synchronized moving aperture radiation therapy (SMART): average tumour trajectory for lung patients. , 2003, Physics in medicine and biology.

[5]  Steve B. Jiang,et al.  Towards fluoroscopic respiratory gating for lung tumours without radiopaque markers , 2005, Physics in medicine and biology.

[6]  Tomoki Kimura,et al.  Reproducibility of organ position using voluntary breath-hold method with spirometer for extracranial stereotactic radiotherapy. , 2003, International journal of radiation oncology, biology, physics.

[7]  Radhe Mohan,et al.  Patient training in respiratory-gated radiotherapy. , 2003, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[8]  E. Yorke,et al.  Deep inspiration breath hold and respiratory gating strategies for reducing organ motion in radiation treatment. , 2004, Seminars in radiation oncology.

[9]  C C Ling,et al.  The deep inspiration breath-hold technique in the treatment of inoperable non-small-cell lung cancer. , 2000, International journal of radiation oncology, biology, physics.

[10]  H. Shirato,et al.  Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor. , 2000, International journal of radiation oncology, biology, physics.

[11]  R. Mohan,et al.  Quantifying the predictability of diaphragm motion during respiration with a noninvasive external marker. , 2003, Medical physics.

[12]  M. V. van Herk,et al.  Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy. , 2002, International journal of radiation oncology, biology, physics.

[13]  Steve B. Jiang,et al.  Synchronized moving aperture radiation therapy (SMART): improvement of breathing pattern reproducibility using respiratory coaching , 2006, Physics in medicine and biology.

[14]  Koichi Yamazaki,et al.  Feasibility of insertion/implantation of 2.0-mm-diameter gold internal fiducial markers for precise setup and real-time tumor tracking in radiotherapy. , 2003, International journal of radiation oncology, biology, physics.

[15]  Steve B. Jiang,et al.  Residual motion of lung tumours in gated radiotherapy with external respiratory surrogates , 2005, Physics in medicine and biology.

[16]  R. T. Ten Haken,et al.  High-dose radiation improved local tumor control and overall survival in patients with inoperable/unresectable non-small-cell lung cancer: long-term results of a radiation dose escalation study. , 2005, International journal of radiation oncology, biology, physics.

[17]  M. V. van Herk,et al.  Physical aspects of a real-time tumor-tracking system for gated radiotherapy. , 2000, International journal of radiation oncology, biology, physics.

[18]  Gig S Mageras,et al.  Comparison of end normal inspiration and expiration for gated intensity modulated radiation therapy (IMRT) of lung cancer. , 2005, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[19]  Kenneth E Rosenzweig,et al.  Evaluation of deep inspiration breath-hold lung treatment plans with Monte Carlo dose calculation. , 2002, International journal of radiation oncology, biology, physics.