A physical action potential generator: design, implementation and evaluation

The objective was to develop a physical action potential generator (Paxon) with the ability to generate a stable, repeatable, programmable, and physiological-like action potential. The Paxon has an equivalent of 40 nodes of Ranvier that were mimicked using resin embedded gold wires (Ø = 20 μm). These nodes were software controlled and the action potentials were initiated by a start trigger. Clinically used Ag-AgCl electrodes were coupled to the Paxon for functional testing. The Paxon's action potential parameters were tunable using a second order mathematical equation to generate physiologically relevant output, which was accomplished by varying the number of nodes involved (1–40 in incremental steps of 1) and the node drive potential (0–2.8 V in 0.7 mV steps), while keeping a fixed inter-nodal timing and test electrode configuration. A system noise floor of 0.07 ± 0.01 μV was calculated over 50 runs. A differential test electrode recorded a peak positive amplitude of 1.5 ± 0.05 mV (gain of 40x) at time 196.4 ± 0.06 ms, including a post trigger delay. The Paxon's programmable action potential like signal has the possibility to be used as a validation test platform for medical surface electrodes and their attached systems.

[1]  W. Rushton A theory of the effects of fibre size in medullated nerve , 1951, The Journal of physiology.

[2]  Matthew D. Johnson,et al.  Spatial steering of deep brain stimulation volumes using a novel lead design , 2011, Clinical Neurophysiology.

[3]  Karin Wårdell,et al.  Relationship between Neural Activation and Electric Field Distribution during Deep Brain Stimulation , 2015, IEEE Transactions on Biomedical Engineering.

[4]  Nigel H. Lovell,et al.  Organic electrode coatings for next-generation neural interfaces , 2014, Front. Neuroeng..

[5]  N. P. Barradas,et al.  TiNx coated polycarbonate for bio-electrode applications , 2012 .

[6]  Chih-Wei Chang,et al.  A Low-Power Bio-Potential Acquisition System with Flexible PDMS Dry Electrodes for Portable Ubiquitous Healthcare Applications , 2013, Sensors.

[7]  J. Anthony Movshon,et al.  Comparison of Recordings from Microelectrode Arrays and Single Electrodes in the Visual Cortex , 2007, The Journal of Neuroscience.

[8]  Seeram Ramakrishna,et al.  Applications of conducting polymers and their issues in biomedical engineering , 2010, Journal of The Royal Society Interface.

[9]  Robert Rieger,et al.  A Device for Emulating Cuff Recordings of Action Potentials Propagating Along Peripheral Nerves , 2014, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[10]  J. Struijk,et al.  An artificial nerve fiber for evaluation of nerve cuff electrodes , 1997, Proceedings of the 19th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. 'Magnificent Milestones and Emerging Opportunities in Medical Engineering' (Cat. No.97CH36136).

[11]  Xiaodong Wang,et al.  An all printed wireless humidity sensor label , 2012 .

[12]  A. Hodgkin,et al.  A quantitative description of membrane current and its application to conduction and excitation in nerve , 1952, The Journal of physiology.

[13]  G Roy,et al.  A simple electronic analog of the squid axon membrane: the NEUROFET. , 1972, IEEE transactions on bio-medical engineering.

[14]  L. Kirkup,et al.  A direct comparison of wet, dry and insulating bioelectric recording electrodes. , 2000, Physiological measurement.

[15]  F. Solzbacher,et al.  Towards high aspect ratio tungsten Micro Electrode Array for neural recording and stimulation applications , 2008 .

[16]  R. Buschbacher,et al.  Manual of Nerve Conduction Studies , 1999 .

[17]  K. Yoshida,et al.  Experimental validation of the nerve conduction velocity selective recording technique using a multi-contact cuff electrode. , 2009, Medical engineering & physics.

[18]  E. Lewis Using electronic circuits to model simple neuroelectric interactions , 1968 .

[19]  R A Spaeth AN ARTIFICIAL NERVE. , 1921, Science.

[20]  C. McIntyre,et al.  Electric field and stimulating influence generated by deep brain stimulation of the subthalamic nucleus , 2004, Clinical Neurophysiology.

[21]  A. Petrėnas,et al.  A comparison of conductive textile-based and silver/silver chloride gel electrodes in exercise electrocardiogram recordings. , 2011, Journal of electrocardiology.

[22]  Claudio Pollo,et al.  Analysis of fractal electrodes for efficient neural stimulation , 2013, Front. Neuroeng..

[23]  Michael D Joseph,et al.  Poly(3,4-ethylenedioxythiophene) (PEDOT) polymer coatings facilitate smaller neural recording electrodes , 2011, Journal of neural engineering.

[24]  J. Webster,et al.  Dry electrodes for electrocardiography , 2013, Physiological measurement.

[25]  Karin Wårdell,et al.  Stereotactic implantation of deep brain stimulation electrodes: a review of technical systems, methods and emerging tools , 2010, Medical & Biological Engineering & Computing.

[26]  E. R. Lewis,et al.  An electronic model of neuroelectric point processes , 1968, Kybernetik.