Multi-institutional comparison of computer-based independent dose calculation for intensity modulated radiation therapy and volumetric modulated arc therapy.

PURPOSE No multi-institutional studies of computer-based independent dose calculation have addressed the discrepancies among radiotherapy treatment planning systems (TPSs) and the verification programs for intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT). We conducted a multi-institutional study to investigate whether ±5% is a reasonable action level for independent dose calculation for IMRT/VMAT. METHODS In total, 477 IMRT/VMAT plans for prostate or head and neck (H&N) malignancies were retrospectively analyzed using a modified Clarkson-based commercial verification program. The doses from the TPSs and verification programs were compared using the mean ±1 standard deviation (SD). RESULTS In the TPS-calculated dose comparisons for prostate and H&N malignancies, the sliding window (SW) technique (-2.5 ± 1.8% and -5.3 ± 2.6%) showed greater negative systematic differences than the step-and-shoot (S&S) technique (-0.3 ± 2.2% and -0.8 ± 2.2%). The VMAT dose differences for prostate and H&N malignancies were 0.9 ± 1.8% and 1.1 ± 3.3%, respectively. The SDs were larger for the H&N plans than for the prostate plans in both IMRT and VMAT. Such plans including more out-of-field control points showed greater systematic differences and SDs. CONCLUSIONS This study will help individual institutions to establish an action level for agreement between primary calculations and verification for IMRT/VMAT. A local dose difference of ±5% at a point within the planning target volume (above -350 HU) may be a reasonable action level.

[1]  J. R. Clarkson,et al.  A Note on Depth Doses in Fields of Irregular Shape , 1941 .

[2]  Comparison of dose calculated by an intensity modulated radiotherapy treatment planning system and an independent monitor unit verification program , 2003, Journal of applied clinical medical physics.

[3]  Andrea L McNiven,et al.  A new metric for assessing IMRT modulation complexity and plan deliverability. , 2010, Medical physics.

[4]  F. Mata Colodro,et al.  Dosimetric validation of a redundant independent calculation software for VMAT fields , 2013 .

[5]  Quan Chen,et al.  Phantomless patient-specific TomoTherapy QA via delivery performance monitoring and a secondary Monte Carlo dose calculation. , 2014, Medical physics.

[6]  Mauro Iori,et al.  Testing of the analytical anisotropic algorithm for photon dose calculation. , 2006, Medical physics.

[7]  Fang-Fang Yin,et al.  Volumetric arc intensity-modulated therapy for spine body radiotherapy: comparison with static intensity-modulated treatment. , 2009, International journal of radiation oncology, biology, physics.

[8]  Rajat J Kudchadker,et al.  Quantification of beam complexity in intensity-modulated radiation therapy treatment plans. , 2014, Medical physics.

[9]  Chang Heon Choi,et al.  Confidence limits for patient‐specific IMRT dose QA: a multi‐institutional study in Korea , 2016, Journal of applied clinical medical physics.

[10]  Benedick A Fraass,et al.  Errors in radiotherapy: motivation for development of new radiotherapy quality assurance paradigms. , 2008, International journal of radiation oncology, biology, physics.

[11]  Jatinder R Palta,et al.  Quality assurance of intensity-modulated radiation therapy. , 2008, International journal of radiation oncology, biology, physics.

[12]  Lei Dong,et al.  Patient-specific point dose measurement for IMRT monitor unit verification. , 2003, International journal of radiation oncology, biology, physics.

[13]  Murty S. Goddu,et al.  Evaluation of the efficiency and effectiveness of independent dose calculation followed by machine log file analysis against conventional measurement based IMRT QA , 2012, Journal of applied clinical medical physics.

[14]  Elisabetta Cagni,et al.  Lung stereotactic ablative body radiotherapy: A large scale multi-institutional planning comparison for interpreting results of multi-institutional studies. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[15]  T. Zhu,et al.  Verification of monitor unit calculations for non-IMRT clinical radiotherapy: report of AAPM Task Group 114. , 2010, Medical physics.

[16]  Y Chen,et al.  Independent dosimetric calculation with inclusion of head scatter and MLC transmission for IMRT. , 2003, Medical physics.

[17]  M. Podgorsak,et al.  Evaluation of fluence‐based dose delivery incorporating the spatial variation of dosimetric leaf gap (DLG) , 2016, Journal of applied clinical medical physics.

[18]  J. Cadeddu,et al.  Safety and Efficacy of Stereotactic Ablative Radiation Therapy for Renal Cell Carcinoma Extracranial Metastases. , 2017, International journal of radiation oncology, biology, physics.

[19]  Ying Niu,et al.  Is RapidArc more susceptible to delivery uncertainties than dynamic IMRT? , 2012, Medical physics.

[20]  Y Watanabe Point dose calculations using an analytical pencil beam kernel for IMRT plan checking. , 2001, Physics in medicine and biology.

[21]  Luca Cozzi,et al.  On the dosimetric behaviour of photon dose calculation algorithms in the presence of simple geometric heterogeneities: comparison with Monte Carlo calculations , 2007, Physics in medicine and biology.

[22]  G. Rosner,et al.  Prospective Study of Hippocampal-Sparing Prophylactic Cranial Irradiation in Limited-Stage Small Cell Lung Cancer. , 2017, International journal of radiation oncology, biology, physics.

[23]  Y Chen,et al.  Monitor unit calculation for an intensity modulated photon field by a simple scatter-summation algorithm. , 2000, Physics in medicine and biology.

[24]  Prakash Jeevanandam,et al.  Comparison between In-house developed and Diamond commercial software for patient specific independent monitor unit calculation and verification with heterogeneity corrections. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[25]  W. Ulmer,et al.  A 3D photon superposition/convolution algorithm and its foundation on results of Monte Carlo calculations , 2005, Physics in medicine and biology.

[26]  Elisabetta Cagni,et al.  Multicentre treatment planning inter-comparison in a national context: The liver stereotactic ablative radiotherapy case. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[27]  John P. Gibbons,et al.  Khan's The Physics of Radiation Therapy , 2014 .