Torque Measurement During Body Cavity Entry Using A Threaded Visual Cannula

Port creation is one of the most common complications associated in endoscopic procedures. The focus of our study was to establish a reproducible method of determining the torque profile generated during body cavity entry using a Threaded Visual Cannula (TVC). A TVC was evaluated on 32 samples of foam media. A torque profile was generated by measuring the torque required to cannulate the foam media. Penetration of the media was captured by a digital imaging system. The magnitude of torque to required cannulate the foam media was dictated by the amount of friction present between the foam and the external threads of the TVC. Furthermore, the magnitude of friction was dictated by the material properties of the individual media. A reproducible method of measuring entry torque during body cavity entry was established. Index terms: Torque measurement, laparoscopic surgery, port creation A. Bakirtzian, A. Ternamian, G. Manson, L. Tremblay, and B. Benhabib, TORQUE MEASUREMENT DURING BODY CAVITY ENTRY USING A THREADED VISUAL CANNULA

[1]  Z. Sipos,et al.  [Open laparoscopy]. , 1988, Medicinski pregled.

[2]  L. Kohn,et al.  To Err Is Human : Building a Safer Health System , 2007 .

[3]  J. Fuller,et al.  Trocar-associated injuries and fatalities: an analysis of 1399 reports to the FDA. , 2005, Journal of minimally invasive gynecology.

[4]  G. Vilos,et al.  Laparoscopic peritoneal entry with the reusable threaded visual cannula. , 2010, Journal of minimally invasive gynecology.

[5]  B. Eisner,et al.  An ex-vivo evaluation of the application and strength of a novel laparoscopic knot substitute device. , 2010, Journal of endourology.

[6]  Olfa Boubaker,et al.  Robot-Assisted Therapy: Design, Control and Optimization , 2012 .

[7]  J. Samet,et al.  From the Food and Drug Administration. , 2002, JAMA.

[8]  Masahiro Ohka,et al.  Object handling precision using mouse-like haptic display generating tactile and force sensation , 2013 .

[9]  Bijan Shirinzadeh,et al.  An actuated force feedback‐enabled laparoscopic instrument for robotic‐assisted surgery , 2014, The international journal of medical robotics + computer assisted surgery : MRCAS.

[10]  Ryojun Ikeura,et al.  OPTIMIZING PERCEIVED HEAVINESS AND MOTION FOR LIFTING OBJECTS WITH A POWER ASSIST ROBOT SYSTEM CONSIDERING CHANGE IN TIME CONSTANT , 2012 .

[11]  J. Clarke Designing safety into the minimally invasive surgical revolution , 2008, Surgical Endoscopy.

[12]  Miguel Srougi,et al.  Safety profile of trocar and insufflation needle access systems in laparoscopic surgery. , 2009, Journal of the American College of Surgeons.

[13]  Miguel L Franco,et al.  An integrated pneumatic tactile feedback actuator array for robotic surgery , 2009, The international journal of medical robotics + computer assisted surgery : MRCAS.

[14]  Artin Ternamian,et al.  Laparoscopic entry: a review of techniques, technologies, and complications. , 2007, Journal of obstetrics and gynaecology Canada : JOGC = Journal d'obstetrique et gynecologie du Canada : JOGC.

[15]  C. Marano,et al.  To err is human. Building a safer health system , 2005 .

[16]  Artin Ternamian,et al.  How to improve laparoscopic access safety: ENDOTIP , 2001, Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy.

[17]  R. Garry,et al.  Towards evidence‐based laparoscopic entry techniques: clinical problems and dilemmas , 1999 .

[18]  A. Kheddar,et al.  A Combined Force and Thermal Feedback Interface for Minimally Invasive Procedures Simulation , 2013, IEEE/ASME Transactions on Mechatronics.

[19]  J. Bisley,et al.  The role of tactile feedback in grip force during laparoscopic training tasks , 2013, Surgical Endoscopy.