Specificity of noninvasive blood glucose sensing using optical coherence tomography technique: a pilot study.

Noninvasive monitoring of blood glucose concentration in diabetic patients would significantly reduce complications and mortality associated with this disease. In this paper, we experimentally and theoretically studied specificity of noninvasive blood glucose monitoring with the optical coherence tomography (OCT) technique. OCT images and signals were obtained from skin of Yucatan micropigs and New Zealand rabbits. Obtained results demonstrate that: (1) several body osmolytes may change the refractive index mismatch between the interstitial fluid (ISF) and scattering centres in tissue, however the effect of the glucose is approximately one to two orders of magnitude higher; (2) an increase of the ISF glucose concentration in the physiological range (3-30 mM) may decrease the scattering coefficient by 0.22% mM(-1) due to cell volume change; (3) stability of the OCT signal slope is dependent on tissue heterogeneity and motion artefacts; and (4) moderate skin temperature fluctuations (+/- 1 degree C) do not decrease accuracy and specificity of the OCT-based glucose sensor, however substantial skin heating or cooling (several degrees C) significantly change the OCT signal slope. These results suggest that the OCT technique may provide blood glucose concentration monitoring with sufficient specificity under normal physiological conditions.

[1]  G. W. Shaw,et al.  In vivo molecular sensing in diabetes mellitus: an implantable glucose sensor with direct electron transfer , 1989, Diabetologia.

[2]  W. March,et al.  Noninvasive Glucose Monitoring of the Aqueous Humor of the Eye: Part I. Measurement of Very Small Optical Rotations , 1982, Diabetes Care.

[3]  M. H. Koelink,et al.  Reduced light-scattering properties for mixtures of spherical particles: a simple approximation derived from Mie calculations. , 1992, Applied optics.

[4]  I. Gabriely,et al.  Transcutaneous glucose measurement using near-infrared spectroscopy during hypoglycemia. , 1999, Diabetes care.

[5]  N. Ray,et al.  Economic Consequences of Diabetes Mellitus in the U.S. in 1997 , 1998, Diabetes Care.

[6]  H. Orskov,et al.  Glucose content in human skin: relationship with blood glucose levels. , 1995, Scandinavian journal of clinical and laboratory investigation.

[7]  M Essenpreis,et al.  The influence of glucose concentration upon the transport of light in tissue-simulating phantoms. , 1995, Physics in medicine and biology.

[8]  A. Dunn,et al.  Light scattering from cells: finite-difference time-domain simulations and goniometric measurements. , 1999, Applied optics.

[9]  A. Kratz,et al.  Case records of the Massachusetts General Hospital. Weekly clinicopathological exercises. Normal reference laboratory values. , 1998, The New England journal of medicine.

[10]  V V Tuchin,et al.  Light propagation in tissues with controlled optical properties , 1996, European Conference on Biomedical Optics.

[11]  Yaochun Shen,et al.  Advances in photoacoustic noninvasive glucose testing. , 1999, Clinical chemistry.

[12]  Massoud Motamedi,et al.  Precision of measurement of tissue optical properties with optical coherence tomography. , 2003, Applied optics.

[13]  R. Esenaliev,et al.  Noninvasive blood glucose monitoring with optical coherence tomography: a pilot study in human subjects. , 2002, Diabetes care.

[14]  Gerard L. Cot,et al.  Application of a multivariate technique to Raman spectra for quantification of body chemicals , 1995, IEEE Transactions on Biomedical Engineering.

[15]  H. V. Hulst Light Scattering by Small Particles , 1957 .

[16]  Jody T. Bruulsema,et al.  Correlation between blood glucose concentration in diabetics and noninvasively measured tissue optical scattering coefficient. , 1997, Optics letters.

[17]  Proposed Strategies for Reducing User Error in Capillary Blood Glucose Monitoring , 1993, Diabetes Care.

[18]  D. Matthews,et al.  An Amperometric Needle‐type Glucose Sensor Tested in Rats and Man , 1988, Diabetic medicine : a journal of the British Diabetic Association.

[19]  R. Dunlay,et al.  Hyperkalemia in hospitalized patients , 2004, International Urology and Nephrology.

[20]  D. Häussinger,et al.  Cell volume in the regulation of hepatic function: a mechanism for metabolic control. , 1991, Biochimica et biophysica acta.

[21]  M A Arnold,et al.  Near-infrared spectroscopic measurement of physiological glucose levels in variable matrices of protein and triglycerides. , 1996, Analytical chemistry.

[22]  Gerard L. Coté,et al.  Noninvasive Optical Glucose Sensing — An Overview , 1997 .

[23]  A. Nettles User Error in Blood Glucose Monitoring: The National Steering Committee for Quality Assurance Report , 1993, Diabetes Care.

[24]  F. Duck Physical properties of tissue , 1990 .

[25]  W. Steen Absorption and Scattering of Light by Small Particles , 1999 .

[26]  Gracie Vargas Reduction of light scattering in biological tissue : implications for optical diagnostics and therapeutics , 2001 .

[27]  Z. Kam,et al.  Absorption and Scattering of Light by Small Particles , 1998 .

[28]  Valery V. Tuchin,et al.  Light propagation in tissues with controlled optical properties , 1996, European Conference on Biomedical Optics.

[29]  M. Fox,et al.  Noninvasive optical polarimetric glucose sensing using a true phase measurement technique , 1992, IEEE Transactions on Biomedical Engineering.

[30]  E Gratton,et al.  Possible correlation between blood glucose concentration and the reduced scattering coefficient of tissues in the near infrared. , 1994, Optics letters.

[31]  J. Fujimoto,et al.  Optical Coherence Tomography , 1991, LEOS '92 Conference Proceedings.

[32]  G. Ripandelli,et al.  Optical coherence tomography. , 1998, Seminars in ophthalmology.

[33]  T Kurihara-Bergstrom,et al.  Characterization of the Yucatan miniature pig skin and small intestine for pharmaceutical applications. , 1986, Laboratory animal science.

[34]  P J Stout,et al.  Comparison of glucose levels in dermal interstitial fluid and finger capillary blood. , 2001, Diabetes technology & therapeutics.

[35]  R. Esenaliev,et al.  Noninvasive monitoring of glucose concentration with optical coherence tomography , 2001 .

[36]  C. Baumgarten,et al.  Osmotic gradient-induced water permeation across the sarcolemma of rabbit ventricular myocytes , 1996, The Journal of general physiology.

[37]  A. Guyton,et al.  Textbook of Medical Physiology , 1961 .

[38]  M. Cope,et al.  Influence of glucose concentration on light scattering in tissue-simulating phantoms. , 1994, Optics letters.

[39]  David C. Klonoff,et al.  Noninvasive Blood Glucose Monitoring , 1997, Diabetes Care.

[40]  Ashley J. Welch,et al.  Effects of compression on soft tissue optical properties , 1996 .

[41]  M Matsumoto,et al.  Evaluation of Yucatan micropig skin for use as an in vitro model for skin permeation study. , 1997, Biological & pharmaceutical bulletin.

[42]  Y Zhang,et al.  Effects of solutes on optical properties of biological materials: models, cells, and tissues. , 1995, Analytical biochemistry.

[43]  E. V. Thomas,et al.  Noninvasive glucose monitoring in diabetic patients: a preliminary evaluation. , 1992, Clinical chemistry.