Closed-loop adaptive optics in the human eye.

We have developed a prototype apparatus for real-time closed-loop measurement and correction of aberrations in the human eye. The apparatus uses infrared light to measure the wave-front aberration at 25 Hz with a Hartmann-Shack sensor. Defocus is removed by a motorized optometer, and higher-order aberrations are corrected by a membrane deformable mirror. The device was first tested with an artificial eye. Correction of static aberrations takes approximately five iterations, making the system capable of following aberration changes at 5 Hz. This capability allows one to track most of the aberration dynamics in the eye. Results in living eyes showed effective closed-loop correction of aberrations, with a residual uncorrected wave front of 0.1microm for a 4.3-mm pupil diameter. Retinal images of a point source in different subjects with and without adaptive correction of aberrations were estimated in real time. The results demonstrate real-time closed-loop correction of aberration in the living eye. An application of this device is as electro-optic "spectacles" to improve vision.

[1]  Y Fainman,et al.  Adaptive control of a micromachined continuous-membrane deformable mirror for aberration compensation. , 1999, Applied optics.

[2]  Sergio R. Restaino,et al.  Laboratory and field demonstration of a low cost membrane mirror adaptive optics system , 2000 .

[3]  Junzhong Liang,et al.  Aberrations and retinal image quality of the normal human eye. , 1997, Journal of the Optical Society of America. A, Optics, image science, and vision.

[4]  P Artal,et al.  Analysis of the performance of the Hartmann-Shack sensor in the human eye. , 2000, Journal of the Optical Society of America. A, Optics, image science, and vision.

[5]  Junzhong Liang,et al.  Objective measurement of wave aberrations of the human eye with the use of a Hartmann-Shack wave-front sensor. , 1994, Journal of the Optical Society of America. A, Optics, image science, and vision.

[6]  J. C. Dainty,et al.  A low cost adaptive optics system using a membrane mirror. , 2000, Optics express.

[7]  P Artal,et al.  High-resolution imaging of the living human fovea: measurement of the intercenter cone distance by speckle interferometry. , 1989, Optics letters.

[8]  R Navarro,et al.  Phase plates for wave-aberration compensation in the human eye. , 2000, Optics letters.

[9]  Robert K. Tyson Principles of Adaptive Optics , 1991 .

[10]  D R Williams,et al.  Supernormal vision and high-resolution retinal imaging through adaptive optics. , 1997, Journal of the Optical Society of America. A, Optics, image science, and vision.

[11]  P. Sarro,et al.  Flexible mirror micromachined in silicon. , 1995, Applied optics.

[12]  R. Weinreb,et al.  Active optical depth resolution improvement of the laser tomographic scanner. , 1989, Applied optics.

[13]  P Artal,et al.  Correction of the aberrations in the human eye with a liquid-crystal spatial light modulator: limits to performance. , 1998, Journal of the Optical Society of America. A, Optics, image science, and vision.