3D printed phantoms mimicking cortical bone for the assessment of ultrashort echo time magnetic resonance imaging

PURPOSE Human cortical bone has a rapid T2∗ decay, and it can be visualized using ultrashort echo time (UTE) techniques in magnetic resonance imaging (MRI). These sequences operate at the limits of gradient and transmit-receive signal performance. Development of multicompartment anthropomorphic phantoms that can mimic human cortical bone can assist with quality assurance and optimization of UTE sequences. The aims of this study were to (a) characterize the MRI signal properties of a photopolymer resin that can be 3D printed, (b) develop multicompartment phantoms based on the resin, and (c) demonstrate the feasibility of using these phantoms to mimic human anatomy in the assessment of UTE sequences. METHODS A photopolymer resin (Prismlab China Ltd, Shanghai, China) was imaged on a 3 Tesla MRI system (Siemens Skyra) to characterize its MRI properties with emphasis on T2∗ signal and longevity. Two anthropomorphic phantoms, using the 3D printed resin to simulate skeletal anatomy, were developed and imaged using UTE sequences. A skull phantom was developed and used to assess the feasibility of using the resin to develop a complex model with realistic morphological human characteristics. A tibia model was also developed to assess the suitability of the resin at mimicking a simple multicompartment anatomical model and imaged using a three-dimensional UTE sequence (PETRA). Image quality measurements of signal-to-noise ratio (SNR) and contrast factor were calculated and these were compared to in vivo values. RESULTS The T2∗ and T1 (mean ± standard deviation) of the photopolymer resin was found to be 411 ± 19 μs and 74.39 ± 13.88 ms, respectively, and demonstrated no statistically significant change during 4 months of monitoring. The resin had a similar T2∗ decay to human cortical bone; however, had lower T1 properties. The bone water concentration of the resin was 59% relative to an external water reference phantom, and this was higher than in vivo values reported for human cortical bone. The multicompartment anthropomorphic head phantom was successfully produced and able to simulate realistic air cavities, bony anatomy, and soft tissue. Image quality assessment in the tibia phantom using the PETRA sequence showed the suitability of the resin to mimic human anatomy with high SNR and contrast making it suitable for tissue segmentation. CONCLUSIONS A solid resin material, which can be 3D printed, has been found to have similar magnetic resonance signal properties to human cortical bone. Phantoms replicating skeletal anatomy were successfully produced using this resin and demonstrated their use for image quality and segmentation assessment of ultrashort echo time sequences.

[1]  Adam Johansson,et al.  CT substitute derived from MRI sequences with ultrashort echo time. , 2011, Medical physics.

[2]  Jiang Du,et al.  Qualitative and quantitative ultrashort‐TE MRI of cortical bone , 2013, NMR in biomedicine.

[3]  E D Pellegrino,et al.  The chemical anatomy of bone. I. A comparative study of bone composition in sixteen vertebrates. , 1969, The Journal of bone and joint surgery. American volume.

[4]  Mark Bydder,et al.  Qualitative and quantitative ultrashort echo time (UTE) imaging of cortical bone. , 2010, Journal of magnetic resonance.

[5]  R. D. Ray,et al.  Bone Density and Composition , 1966 .

[6]  Michael Bock,et al.  MRI compatible head phantom for ultrasound surgery. , 2015, Ultrasonics.

[7]  E. Chang,et al.  UTE imaging in the musculoskeletal system , 2015, Journal of magnetic resonance imaging : JMRI.

[8]  Kalaivani Thangavel,et al.  Aqueous paramagnetic solutions for MRI phantoms at 3 T: A detailed study on relaxivities , 2017, Turkish J. Electr. Eng. Comput. Sci..

[9]  Y. Yeni,et al.  Influence of bone composition and apparent density on fracture toughness of the human femur and tibia. , 1998, Bone.

[10]  Matthew D Robson,et al.  Magnetic resonance imaging of cortical bone with ultrashort TE pulse sequences. , 2005, Magnetic resonance imaging.

[11]  N. Otsu A threshold selection method from gray level histograms , 1979 .

[12]  Jiang Du,et al.  Measurement of T1 of the Ultrashort T2* Components in White Matter of the Brain at 3T , 2014, PloS one.

[13]  Stephan E Maier,et al.  Three‐dimensional printing of MRI‐visible phantoms and MR image‐guided therapy simulation , 2017, Magnetic resonance in medicine.

[14]  Tatiana Kelil,et al.  3D printing from MRI Data: Harnessing strengths and minimizing weaknesses , 2017, Journal of magnetic resonance imaging : JMRI.

[15]  David Grodzki,et al.  Quiet Submillimeter MR Imaging of the Lung Is Feasible with a PETRA Sequence at 1.5 T. , 2015, Radiology.

[16]  Koichi Masuda,et al.  Quantitative ultrashort echo time (UTE) MRI of human cortical bone: Correlation with porosity and biomechanical properties , 2012, Journal of bone and mineral research : the official journal of the American Society for Bone and Mineral Research.

[17]  Mary B Leonard,et al.  Cortical bone water: in vivo quantification with ultrashort echo-time MR imaging. , 2008, Radiology.

[18]  Kh. I. Amirkhanov,et al.  The Influence of , 1957 .

[19]  C. Kuhl,et al.  MRI-Based Attenuation Correction for Hybrid PET/MRI Systems: A 4-Class Tissue Segmentation Technique Using a Combined Ultrashort-Echo-Time/Dixon MRI Sequence , 2012, The Journal of Nuclear Medicine.

[20]  Soumya Ghose,et al.  Substitute CT generation from a single ultra short time echo MRI sequence: preliminary study , 2017, Physics in medicine and biology.

[21]  P. Barrie Characterization of porous media using NMR methods , 2000 .

[22]  J M Pauly,et al.  Lung parenchyma: projection reconstruction MR imaging. , 1991, Radiology.