Improved Performance of On-line Atom Trapping in Flame Furnace Atomic Absorption Spectrometry by Chemical Vapor Generation: Determination of Cadmium in High-Salinity Water Samples

ABSTRACT On-line atom trapping inside a nickel flame furnace using chemical vapor generation for sample introduction was proposed for the determination of trace cadmium by flame atomic absorption spectrometry (AAS). Cadmium volatile species was generated upon reaction with potassium borohydride and then flushed into a flame furnace for on-line trapping by a flow of nitrogen carrier gas. The middle part of the flame furnace, where the carrier gas impacts, is cooled by the gas flow, and this provides a fine strategy for on-line atom trapping for the purpose of preconcentration. A stainless steel plate is put on the top of the flame burner in the middle to form a flame-free zone, which also greatly lowers the temperature of the flame furnace and facilitates the atom-trapping process. Due to the introduction of chemical vapor generation, matrix effect was greatly alleviated compared with direct pneumatic nebulization for on-line atom trapping in flame furnace AAS. With trapping time of 35 s, the current approach achieved an excellent limit of detection of 20 ng L−1. The proposed method was successfully applied for the quantification of cadmium in high-salinity samples.

[1]  J. Kratzer,et al.  Stibine and bismuthine trapping in quartz tube atomizers for atomic absorption spectrometry — Method optimization and analytical applications☆ , 2008 .

[2]  O. Ataman Vapor generation and atom traps: Atomic absorption spectrometry at the ng/L level , 2008 .

[3]  M. Arruda,et al.  A metallic furnace atomizer in hydride generation atomic absorption spectrometry: Determination of bismuth and selenium☆ , 2008 .

[4]  X. Hou,et al.  Highly sensitive and interference-free determination of bismuth in environmental samples by electrothermal vaporization atomic fluorescence spectrometry after hydride trapping on iridium-coated tungsten coil , 2008 .

[5]  X. Hou,et al.  Highly sensitive pneumatic nebulization flame furnace atomic absorption spectrometry: complete sample aerosol introduction and on-line preconcentration of cadmium by atom trap , 2008 .

[6]  H. Matusiewicz,et al.  Determination of total antimony and inorganic antimony species by hydride generation in situ trapping flame atomic absorption spectrometry : a new way to (ultra)trace speciation analysis , 2008 .

[7]  M. Arruda,et al.  Metal furnace heated by flame as a hydride atomizer for atomic absorption spectrometry: Sb determination in environmental and pharmaceutical samples. , 2007, Talanta: The International Journal of Pure and Applied Analytical Chemistry.

[8]  M. Vieira,et al.  Coupling of ultrasonic nebulization to flame furnace atomic absorption spectrometry—new possibilities for trace element determination , 2007 .

[9]  H. Matusiewicz,et al.  Hydride generation-in situ trapping-flame atomic absorption spectrometry hybridization for indium and thallium determination , 2007 .

[10]  H. Matusiewicz,et al.  Determination of tellurium by hydride generation with in situ trapping flame atomic absorption spectrometry , 2007 .

[11]  J. Kratzer,et al.  Arsine and selenium hydride trapping in a novel quartz device for atomic-absorption spectrometry , 2007, Analytical and bioanalytical chemistry.

[12]  Deniz Korkmaz,et al.  Determination of antimony by using a quartz atom trap and electrochemical hydride generation atomic absorption spectrometry , 2007 .

[13]  H. Matusiewicz,et al.  Determination of cadmium and lead in reference materials by volatile species generation with in situ trapping flame atomic absorption spectrometry , 2006 .

[14]  H. Berndt,et al.  Air driven on-line separation and preconcentration on a C18 column coupled with thermospray flame furnace AAS for the determination of cadmium and lead at μg l−1 levels , 2004 .

[15]  X. Hou,et al.  Chemical vapor generation by reaction of cadmium with potassium tetrahydroborate and sodium iodate in acidic aqueous solution for atomic fluorescence spectrometric application , 2004 .

[16]  M. Arruda,et al.  Determination of cadmium and lead at low levels by using preconcentration at fullerene coupled to thermospray flame furnace atomic absorption spectrometry , 2004 .

[17]  Z. Mester,et al.  Analytical Applications of Volatile Metal Derivatives , 2002 .

[18]  N. Ertaş,et al.  A novel silica trap for lead determination by hydride generation atomic absorption spectrometry , 2002 .

[19]  B. Dočekal,et al.  Investigation of in situ trapping of selenium and arsenic hydrides within a tungsten tube atomiser , 2001 .

[20]  H. Berndt,et al.  Thermospray flame furnace atomic absorption spectrometry (TS-FF-AAS) — a simple method for trace element determination with microsamples in the μg/l concentration range , 2000 .

[21]  H. Berndt,et al.  Beam injection flame furnace atomic absorption spectrometry: a new flame method. , 2000, Analytical chemistry.

[22]  H. Matusiewicz Atom trapping and in situ preconcentration techniques for flame atomic absorption spectrometry , 1997 .

[23]  M. Liu,et al.  The determination of trace cadmium by flow injection cold vapor generation AAS , 1997 .

[24]  H. Matusiewicz,et al.  Atomic spectrometric detection of hydride forming elements following in situ trapping within a graphite furnace , 1996 .

[25]  A. Sanz-Medel,et al.  Generation of cadmium atoms at room temperature using vesicles and its application to cadmium determination by cold vapor atomic spectrometry , 1995 .

[26]  Xiao-wei Guo,et al.  Studies on the reaction between cadmium and potassium tetrahydroborate in aqueous solution and its application in atomic fluorescence spectrometry , 1995 .

[27]  Guo Xiao-wei,et al.  Determination of cadmium at ultratrace levels by cold vapour atomic absorption spectrometry , 1995 .