A quantitative image quality comparison of four different image guided radiotherapy devices.

PURPOSE A study to quantitatively compare the image quality of four different image guided radiotherapy (IGRT) devices based on phantom measurements with respect to the additional dose delivered to the patient. METHODS Images of three different head-sized phantoms (diameter 16-18 cm) were acquired with the following four IGRT-CT solutions: (i) the Siemens Primatom single slice fan beam computed tomography (CT) scanner with an acceleration voltage of 130 kV, (ii) a Tomotherapy HI-ART II unit using a fan beam scanner with an energy of 3.5 MeV and (iii) the Siemens Artíste prototype, providing the possibility to perform kV (121 kV) and MV (6 MV) cone beam (CB) CTs. For each device three scan protocols (named low, normal, high) were selected to yield the same weighted computed tomography dose index (CTDI(w)). Based on the individual inserts of the different phantoms the image quality achieved with each device at a certain dose level was characterized in terms of homogeneity, spatial resolution, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and electron density-to-CT-number conversion. RESULTS Based on the current findings for head-sized phantoms all devices show an electron density-to-CT-number conversion almost independent of the imaging parameters and hence can be suited for treatment planning purposes. The evaluation of the image quality, however, points out clear differences due to the different energies and geometries. The Primatom standard CT scanner shows throughout the best performance, especially for soft tissue contrast and spatial resolution with low imaging doses. Reasonable soft tissue contrast can be obtained with slightly higher doses compared to the CT scanner with the kVCB and the Tomotherapy unit. In order to get similar results with the MVCB system a much higher dose needs to be applied to the patient. CONCLUSION Considering the entire investigations, especially in terms of contrast and spatial resolution, a rough tendency for decreasing image quality can be given: Primatom, Artíste prototype kVCB, Tomotherapy, Artíste prototype MVCB.

[1]  J O Deasy,et al.  Tomotherapy: a new concept for the delivery of dynamic conformal radiotherapy. , 1993, Medical physics.

[2]  Mark Oldham,et al.  Cone-beam-CT guided radiation therapy: A model for on-line application. , 2005, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[3]  Moyed Miften,et al.  Patient dose and image quality from mega-voltage cone beam computed tomography imaging. , 2007, Medical physics.

[4]  M. Oldham,et al.  Cone-beam-CT guided radiation therapy: technical implementation. , 2005, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[5]  J. Ehrhardt,et al.  Modulation transfer function of the EMI CT head scanner. , 1977, Medical physics.

[6]  Thomas Tücking Development and Realization of the IGRT Inline Concept , 2006 .

[7]  Rebecca Fahrig,et al.  Dose and image quality for a cone-beam C-arm CT system. , 2006, Medical physics.

[8]  Patrick A Kupelian,et al.  Performance characterization of megavoltage computed tomography imaging on a helical tomotherapy unit. , 2005, Medical physics.

[9]  S. Nill,et al.  Linac-integrated kV-cone beam CT: technical features and first applications. , 2006, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.