Comparison of 250 MHz electron spin echo and continuous wave oxygen EPR imaging methods for in vivo applications.

PURPOSE The authors compare two electron paramagnetic resonance imaging modalities at 250 MHz to determine advantages and disadvantages of those modalities for in vivo oxygen imaging. METHODS Electron spin echo (ESE) and continuous wave (CW) methodologies were used to obtain three-dimensional images of a narrow linewidth, water soluble, nontoxic oxygen-sensitive trityl molecule OX063 in vitro and in vivo. The authors also examined sequential images obtained from the same animal injected intravenously with trityl spin probe to determine temporal stability of methodologies. RESULTS A study of phantoms with different oxygen concentrations revealed a threefold advantage of the ESE methodology in terms of reduced imaging time and more precise oxygen resolution for samples with less than 70 torr oxygen partial pressure. Above 100 torr, CW performed better. The images produced by both methodologies showed pO2 distributions with similar mean values. However, ESE images demonstrated superior performance in low pO2 regions while missing voxels in high pO2 regions. CONCLUSIONS ESE and CW have different areas of applicability. ESE is superior for hypoxia studies in tumors.

[1]  Murali C Krishna,et al.  Pharmacokinetics of a triarylmethyl‐type paramagnetic spin probe used in EPR oximetry , 2004, Magnetic resonance in medicine.

[2]  H. Halpern,et al.  Retractable loop-gap resonators for electron paramagnetic resonance imaging with in situ irradiation capabilities , 2011 .

[3]  Colin Mailer,et al.  Spectral fitting: The extraction of crucial information from a spectrum and a spectral image , 2003, Magnetic resonance in medicine.

[4]  Thomas J Pohida,et al.  A new strategy for fast radiofrequency CW EPR imaging: direct detection with rapid scan and rotating gradients. , 2007, Journal of magnetic resonance.

[5]  Erwin G. Van Meir Hypoxia-mediated selection of cells with diminished apoptotic potential to solid tumours. , 1996, Neurosurgery.

[6]  Bruce Klitzman,et al.  Direct demonstration of instabilities in oxygen concentrations within the extravascular compartment of an experimental tumor. , 2006, Cancer research.

[7]  H J Halpern,et al.  Oxymetry deep in tissues with low-frequency electron paramagnetic resonance. , 1994, Proceedings of the National Academy of Sciences of the United States of America.

[8]  David E. Housman,et al.  Hypoxia-mediated selection of cells with diminished apoptotic potential in solid tumours , 1996, Nature.

[9]  Martyna Elas,et al.  Electron paramagnetic resonance oxygen image hypoxic fraction plus radiation dose strongly correlates with tumor cure in FSa fibrosarcomas. , 2008, International journal of radiation oncology, biology, physics.

[10]  Xiaochuan Pan,et al.  Image reconstruction in regions of interest from truncated Radon transforms of even dimensions , 2005 .

[11]  H. Halpern,et al.  In vivo EPR Imaging , 2005 .

[12]  Colin Mailer,et al.  Spin echo spectroscopic electron paramagnetic resonance imaging , 2006, Magnetic resonance in medicine.

[13]  Howard J Halpern,et al.  Simulation of 4D spectral-spatial EPR images. , 2007, Journal of magnetic resonance.

[14]  Howard J Halpern,et al.  Comparison of local and global angular interpolation applied to spectral-spatial EPR image reconstruction. , 2007, Medical physics.

[15]  Colin Mailer,et al.  A passive dual‐circulator based transmit/receive switch for use with reflection resonators in pulse electron paramagnetic resonance , 2009 .

[16]  James B. Mitchell,et al.  Overhauser enhanced magnetic resonance imaging for tumor oximetry: Coregistration of tumor anatomy and tissue oxygen concentration , 2002, Proceedings of the National Academy of Sciences of the United States of America.

[17]  Richard W Quine,et al.  A Very Fast Switched Attenuator Circuit for Microwave and R.F. Applications. , 2010, Concepts in magnetic resonance. Part B, Magnetic resonance engineering.

[18]  Howard J Halpern,et al.  Characterization of response to radiation mediated gene therapy by means of multimodality imaging , 2009, Magnetic resonance in medicine.

[19]  Gareth R. Eaton,et al.  Spectral-spatial two-dimensional EPR imaging , 1987 .

[20]  M. Dewhirst,et al.  Fourier analysis of fluctuations of oxygen tension and blood flow in R3230Ac tumors and muscle in rats. , 1999, American journal of physiology. Heart and circulatory physiology.

[21]  Xiaochuan Pan,et al.  Targeted-ROI imaging in electron paramagnetic resonance imaging. , 2007, Journal of magnetic resonance (San Diego, Calif. 1997 : Print).

[22]  J S Petersson,et al.  EPR and DNP properties of certain novel single electron contrast agents intended for oximetric imaging. , 1998, Journal of magnetic resonance.

[23]  R. Quine,et al.  Fast-Response VHF Pulsed 2 KW Power Amplifiers , 2006 .

[24]  Howard J. Halpern,et al.  Imaging radio frequency electron‐spin‐resonance spectrometer with high resolution and sensitivity for in vivo measurements , 1989 .

[25]  Paul C. Lauterbur,et al.  Theory and simulation of NMR spectroscopic imaging and field plotting by projection reconstruction involving an intrinsic frequency dimension , 1984 .

[26]  G. Rinard,et al.  Frequency dependence of EPR signal intensity, 250 MHz to 9.1 GHz. , 2002, Journal of magnetic resonance.

[27]  Andrew G. Taube,et al.  Single‐point (constant‐time) imaging in radiofrequency Fourier transform electron paramagnetic resonance † , 2002, Magnetic resonance in medicine.

[28]  Boris Epel,et al.  Spectrometer manager: A versatile control software for pulse EPR spectrometers , 2005 .

[29]  James B. Mitchell,et al.  Evaluation and comparison of pulsed and continuous wave radiofrequency electron paramagnetic resonance techniques for in vivo detection and imaging of free radicals. , 2002, Journal of magnetic resonance.

[30]  Howard J Halpern,et al.  Spatially uniform sampling in 4-D EPR spectral-spatial imaging. , 2007, Journal of magnetic resonance.

[31]  Deepti S Vikram,et al.  Methods for noninvasive imaging of tissue hypoxia. , 2007, Antioxidants & redox signaling.

[32]  Howard J Halpern,et al.  Immobilization Using Dental Material Casts Facilitates Accurate Serial and Multimodality Small Animal Imaging. , 2008, Concepts in magnetic resonance. Part B, Magnetic resonance engineering.

[33]  S. Eaton,et al.  Electron Paramagnetic Resonance Imaging , 2011 .

[34]  J. Zweier,et al.  Three-dimensional spectral-spatial EPR imaging of free radicals in the heart: a technique for imaging tissue metabolism and oxygenation. , 1994, Proceedings of the National Academy of Sciences of the United States of America.

[35]  C. Pelizzari,et al.  Reduction of Image Artifacts in Mice by Bladder Flushing with a Novel Double-Lumen Urethral Catheter , 2006, Molecular imaging.

[36]  B H Robinson,et al.  Linewidth analysis of spin labels in liquids. I. Theory and data analysis. , 1999, Journal of magnetic resonance.

[37]  Murali C Krishna,et al.  Electron paramagnetic resonance imaging of tumor hypoxia: Enhanced spatial and temporal resolution for in vivo pO2 determination , 2006, Magnetic resonance in medicine.

[38]  J. Berliner,et al.  Magnetic resonance imaging of biological specimens by electron paramagnetic resonance of nitroxide spin labels. , 1985, Science.

[39]  Colin Mailer,et al.  A Versatile High Speed 250 MHz Pulse Imager for Biomedical Applications. , 2008, Concepts in magnetic resonance. Part B, Magnetic resonance engineering.

[40]  M. Dewhirst,et al.  Microvascular studies on the origins of perfusion-limited hypoxia. , 1996, The British journal of cancer. Supplement.