Analysis of the eddy-current induced artifacts and the temporal compensation in nuclear magnetic resonance imaging.

Reduction of eddy currents by a temporal compensation of the input current waveform to the gradient coil is studied with an analytic solution. The technique is the inverse filtering of the eddy-current affected field response, which is calculated from the diffusion equation. The limitation of the temporal compensation due to the spatially variant eddy currents is also investigated for whole-body diagnostic imaging systems and small-bore nuclear magnetic resonance (NMR) microscopy systems. Within a limited imaging volume of less than 60% of the gradient coil diameter, most of eddy-current problems can be solved by the technique.

[1]  Chang-Beom Ahn,et al.  Nuclear magnetic resonance microscopy with 4-μm resolution: Theoretical study and experimental results , 1988 .

[2]  Z H Cho,et al.  The effects of random directional distributed flow in nuclear magnetic resonance imaging. , 1987, Medical physics.

[3]  Zang-Hee Cho,et al.  Analytic solution of eddy currents and its temporal compensation in nuclear magnetic resonance imaging , 1990, Medical Imaging.

[4]  P. Mansfield,et al.  Active magnetic screening of gradient coils in NMR imaging , 1986 .

[5]  Z H Cho,et al.  Analysis of eddy currents in nuclear magnetic resonance imaging , 1991, Magnetic resonance in medicine.

[6]  W. Dannels,et al.  Exact temporal eddy current compensation in magnetic resonance imaging systems. , 1988, IEEE transactions on medical imaging.

[7]  R. Turner,et al.  Passive screening of switched magnetic field gradients , 1986 .

[8]  Eiichi Fukushima,et al.  Eddy current compensation by direct field detection and digital gradient modification , 1990 .

[9]  E. Kriezis,et al.  Eddy currents in linear conducting media , 1984 .

[10]  Z H Cho,et al.  An improved nuclear magnetic resonance diffusion coefficient imaging method using an optimized pulse sequence. , 1986, Medical physics.

[11]  A. Macovski,et al.  Selective Projection Imaging: Applications to Radiography and NMR , 1982, IEEE Transactions on Medical Imaging.

[12]  R. L. Stoll The analysis of eddy currents , 1974 .

[13]  C. Ahn,et al.  A generalized formulation of diffusion effects in micron resolution nuclear magnetic resonance imaging. , 1989, Medical physics.

[14]  P A Narayana,et al.  Reduction of pulsed gradient settling time in the superconducting magnet of a magnetic resonance instrument. , 1987, Medical physics.