Accuracy of segmented MR velocity mapping to measure small vessel pulsatile flow in a phantom simulating cardiac motion

The purpose of this study was to investigate the accuracy of conventional, segmented, and echo‐shared MR velocity mapping sequences to measure pulsatile flow in small moving vessels using a phantom with simulated cardiac motion. The phantom moved either cyclically in‐plane, through‐plane, in‐ and through‐plane, or was stationary. The mean error in average flow was –2% ± 3% (mean ± SD) for all sequences under all conditions, with or without background correction, as long as the region of interest (ROI) size was equal to the vessel cross‐sectional size. Overestimation of flow as a result of an oversized ROI was less than 20%, and independent of field of view (FOV) and matrix, as long as the offset in angle between the imaging plane and flow direction was less than 10 degrees. Segmented velocity mapping sequences are surprisingly accurate in measuring average flow and render flow profiles in small moving vessels despite the blurring in the images due to vessel motion. J. Magn. Reson. Imaging 2001;13:722–728. © 2001 Wiley‐Liss, Inc.

[1]  A. Sawyer-Glover,et al.  Renal blood flow: measurement in vivo with rapid spiral MR imaging. , 1998, Radiology.

[2]  T. Grist,et al.  Effect of and correction for in‐plane myocardial motion on estimates of coronary‐volume flow rates , 1997, Journal of magnetic resonance imaging : JMRI.

[3]  A. Haase,et al.  ECG-triggered arterial FLASH-MR flow measurement using an external standard. , 1987, Magnetic resonance imaging.

[4]  B. Carroll,et al.  Flow quantification using fast cine phase-contrast MR imaging, conventional cine phase-contrast MR imaging, and Doppler sonography: in vitro and in vivo validation. , 1997, AJR. American journal of roentgenology.

[5]  Nico Westerhof,et al.  In Vivo Validation of Magnetic Resonance Blood Volume Flow Measurements with Limited Spatial Resolution in Small Vessels , 1995, Magnetic resonance in medicine.

[6]  K.,et al.  Blood flow imaging with MR: spin-phase phenomena. , 1985, Radiology.

[7]  René M. Botnar,et al.  Automatic vessel segmentation using active contours in cine phase contrast flow measurements , 1999, Journal of magnetic resonance imaging : JMRI.

[8]  P. V. van Dijk Direct cardiac NMR imaging of heart wall and blood flow velocity. , 1984, Journal of computer assisted tomography.

[9]  René M. Botnar,et al.  Plow quantitation with echo‐planar phase‐contrast velocity mapping: In vitro and in vivo evaluation , 1995 .

[10]  T. Scalea,et al.  Blunt splenic injuries: nonsurgical treatment with CT, arteriography, and transcatheter arterial embolization of the splenic artery. , 1991, Radiology.

[11]  T. Carrel,et al.  Evaluation of coronary bypass flow with color-Doppler and magnetic resonance imaging techniques: comparison with intraoperative flow measurements. , 1999, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[12]  J. Kuijer,et al.  Flow profiles in the left anterior descending and the right coronary artery assessed by MR velocity quantification: effects of through-plane and in-plane motion of the heart. , 1999, Journal of computer assisted tomography.

[13]  T. Grist,et al.  Measurement of coronary blood flow and flow reserve using magnetic resonance imaging. , 1997, Cardiology.

[14]  A. Häkkinen,et al.  Multiphase segmented k-space velocity mapping in pulsatile flow waveforms. , 1998, Magnetic resonance imaging.

[15]  C. Higgins,et al.  Coronary flow reserve: noninvasive measurement in humans with breath-hold velocity-encoded cine MR imaging. , 1996, Radiology.

[16]  R. Edelman,et al.  Coronary arteries: breath-hold MR angiography. , 1991, Radiology.

[17]  N. Westerhof,et al.  Assessment of flow in the right human coronary artery by magnetic resonance phase contrast velocity measurement: Effects of cardiac and respiratory motion , 1996, Magnetic resonance in medicine.

[18]  René M. Botnar,et al.  Flow quantitation with echo-planar phase-contrast velocity mapping: in vitro and in vivo evaluation. , 1995, Journal of magnetic resonance imaging : JMRI.

[19]  D L Parker,et al.  Correction of partial‐volume effects in phase‐contrast flow measurements , 1995, Journal of magnetic resonance imaging : JMRI.

[20]  E M Haacke,et al.  Retrospective cardiac gating: a review of technical aspects and future directions. , 1989, Magnetic resonance imaging.

[21]  E. Wellnhofer,et al.  Noninvasive determination of coronary blood flow velocity with magnetic resonance imaging: Comparison of breath‐hold and navigator techniques with intravascular ultrasound , 1999, Magnetic resonance in medicine.

[22]  C. Nilsson,et al.  Use of k‐space segmentation in MR velocity mapping for rapid quantification of CSF flow , 1997, Journal of magnetic resonance imaging : JMRI.

[23]  C Thomsen,et al.  Comparison between retrospective gating and ECG triggering in magnetic resonance velocity mapping. , 1993, Magnetic resonance imaging.

[24]  C. Higgins,et al.  Heart disease: functional evaluation with MR imaging. , 1996, Radiology.

[25]  D N Firmin,et al.  Blood flow imaging by cine magnetic resonance. , 1986, Journal of computer assisted tomography.

[26]  Yi Wang,et al.  A complex‐difference phase‐contrast technique for measurement of volume flow rates , 1995, Journal of magnetic resonance imaging : JMRI.

[27]  C Thomsen,et al.  A segmented k‐space velocity mapping protocol for quantification of renal artery blood plow during breath‐holding , 1995, Journal of magnetic resonance imaging : JMRI.

[28]  C. Higgins,et al.  Quantification of coronary artery volume flow rate using fast velocity-encoded cine MR imaging. , 1997, AJR. American journal of roentgenology.

[29]  C. Lorenz,et al.  Quantification of in‐plane motion of the coronary arteries during the cardiac cycle: Implications for acquisition window duration for MR flow quantification , 1998, Journal of magnetic resonance imaging : JMRI.

[30]  R L Ehman,et al.  Analysis of systematic and random error in MR volumetric flow measurements , 1993, Magnetic resonance in medicine.

[31]  O Henriksen,et al.  Mitral and aortic valvular flow: Quantification with MR phase mapping , 1992, Journal of magnetic resonance imaging : JMRI.

[32]  N J Pelc,et al.  Renal artery blood flow: quantitation with phase-contrast MR imaging with and without breath holding. , 1994, Radiology.

[33]  T K Foo,et al.  Improved ejection fraction and flow velocity estimates with use of view sharing and uniform repetition time excitation with fast cardiac techniques. , 1995, Radiology.