Clinical management of electromagnetic interferences in patients with pacemakers and implantable cardioverter-defibrillators: review of the literature and focus on magnetic resonance conditional devices

The number of cardiac implantable electronic devices (CIEDs) has greatly increased in the last 10 years. Many electronic devices used in daily activities generate electromagnetic interferences (EMIs) that can interact with CIEDs. In clinical practice, it is very important to know the potential sources of EMIs and their effect on CIEDs in order to understand how to manage or mitigate them. A very important source of EMI is magnetic resonance (MR), which is considered nowadays the diagnostic gold standard for different anatomical districts. In this review, we focused on the effects of EMI on CIEDs and on the clinical management. Moreover, we made a clarification about MR and CIEDs. In patients with CIEDs, EMIs may cause potentially serious and even life-threatening complications (inappropriate shocks, device malfunctions, inhibition of pacing in pacemaker-dependent patients) and may rarely dictate device replacement. The association of inappropriate shocks with increased mortality highlights the importance of minimizing the occurrence of EMI. Adequate advice and recommendations about the correct management of EMIs in patients with CIEDs are required to avoid all complications during hospitalization and in daily life. Furthermore, the article focused on actual management about MR and CIEDs.

[1]  H. Halperin,et al.  Clinical Utility and Safety of a Protocol for Noncardiac and Cardiac Magnetic Resonance Imaging of Patients With Permanent Pacemakers and Implantable-Cardioverter Defibrillators at 1.5 Tesla , 2006, Circulation.

[2]  Peter Hunold,et al.  Magnetic resonance imaging in patients with a pacemaker system designed for the magnetic resonance environment. , 2011, Heart rhythm.

[3]  H. Halperin,et al.  Effects of Surgical and Endoscopic Electrocautery on Modern‐Day Permanent Pacemaker and Implantable Cardioverter‐Defibrillator Systems , 2008, Pacing and clinical electrophysiology : PACE.

[4]  P. Marino,et al.  Lack of interference of electromagnetic navigation bronchoscopy to implanted cardioverter-defibrillator: in-vivo study. , 2014, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[5]  D. Albers,et al.  Shockwave lithotripsy and pacemakers: experience with 20 cases. , 1995, Journal of endourology.

[6]  S. Ryf,et al.  Potential interference of small neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators. , 2007, Heart rhythm.

[7]  A. Curnis,et al.  Assessing access to MRI of patients with magnetic resonance-conditional pacemaker and implantable cardioverter defibrillator systems: the Really ProMRI study design , 2015, Journal of cardiovascular medicine.

[8]  P. Reddy,et al.  Effects of Electromagnetic Interference on Implanted Cardiac Devices and Their Management , 2007, Cardiology in review.

[9]  L. Oddershede,et al.  The effect of radiotherapy beam energy on modern cardiac devices: an in vitro study. , 2014, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[10]  J. Nielsen,et al.  ESC Guidelines on cardiac pacing and cardiac resynchronization therapy , 2014 .

[11]  G. Trevi,et al.  Implantable Cardioverter Defibrillators and Cellular Telephones: Is There Any Interference? , 1999, Pacing and clinical electrophysiology : PACE.

[12]  J. Triedman,et al.  Digital music players cause interference with interrogation telemetry for pacemakers and implantable cardioverter-defibrillators without affecting device function. , 2008, Heart rhythm.

[13]  M. Chester,et al.  The simultaneous use of a biventricular implantable cardioverter defibrillator (ICD) and transcutaneous electrical nerve stimulation (TENS) unit: implications for device interaction. , 2003, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[14]  G. Boriani,et al.  Cardiac resynchronization therapy: implant rates, temporal trends and relationships with heart failure epidemiology , 2014, Journal of cardiovascular medicine.

[15]  P. Macdonald,et al.  Left Ventricular Mechanical Assist Devices and Cardiac Device Interactions: An Observational Case Series , 2009, Pacing and clinical electrophysiology : PACE.

[16]  M. McIvor,et al.  Study of Pacemaker and Implantable Cardioverter Defibrillator Triggering by Electronic Article Surveillance Devices (SPICED TEAS) , 1998, Pacing and clinical electrophysiology : PACE.

[17]  E. Kanal,et al.  ACR guidance document on MR safe practices: 2013 , 2013, Journal of magnetic resonance imaging : JMRI.

[18]  David L. Hayes,et al.  Randomized trial of pacemaker and lead system for safe scanning at 1.5 Tesla. , 2013, Heart rhythm.

[19]  D P Zipes,et al.  Interactions between electronic article surveillance systems and implantable cardioverter-defibrillators. , 1999, Circulation.

[20]  C. Chiu,et al.  A Prospective Pediatric Clinical Trial of Digital Music Players: Do They Interfere with Pacemakers? , 2009, Journal of cardiovascular electrophysiology.

[21]  I. Dell'Oca,et al.  Radiotherapy and implanted cardioverter defibrillators: novel techniques make it feasible. , 2017, Journal of cardiovascular medicine.

[22]  M. Lazebnik,et al.  Safe magnetic resonance imaging scanning of patients with cardiac rhythm devices: a role for computer modeling. , 2013, Heart rhythm.

[23]  Nikolaus Marx,et al.  Electromagnetic Interference With Implantable Cardioverter-Defibrillators at Power Frequency: An In Vivo Study , 2014, Circulation.

[24]  M. Chung,et al.  The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) Expert Consensus Statement on the perioperative management of patients with implantable defibrillators, pacemakers and arrhythmia monitors: facilities and patient management this document was developed as a joint project with , 2011, Heart rhythm.

[25]  J Rod Gimbel,et al.  Electronic article surveillance systems and interactions with implantable cardiac devices: risk of adverse interactions in public and commercial spaces. , 2007, Mayo Clinic proceedings.

[26]  Nora Goldschlager,et al.  Electromagnetic Interference and Implanted Cardiac Devices: The Nonmedical Environment (Part I) , 2012, Clinical cardiology.

[27]  S. Saksena,et al.  Interaction Between Electronic Article Surveillance Systems and Implantable Defibrillators: Insights from a Fourth Generation ICD , 1997, Pacing and clinical electrophysiology : PACE.

[28]  G. Boriani,et al.  Statistics on the use of cardiac electronic devices and electrophysiological procedures in 55 ESC countries: 2013 report from the European Heart Rhythm Association (EHRA). , 2014, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[29]  Eorge,et al.  INTERFERENCE WITH CARDIAC PACEMAKERS BY CELLULAR TELEPHONES , 2000 .

[30]  R. Fimmers,et al.  Safety, feasibility, and diagnostic value of cardiac magnetic resonance imaging in patients with cardiac pacemakers and implantable cardioverters/defibrillators at 1.5 T. , 2011, American heart journal.

[31]  M. Sweesy Understanding electromagnetic interference. , 2004, Heart rhythm.

[32]  Henry R. Halperin,et al.  Modern Pacemaker and Implantable Cardioverter/Defibrillator Systems Can Be Magnetic Resonance Imaging Safe: In Vitro and In Vivo Assessment of Safety and Function at 1.5 T , 2004, Circulation.

[33]  M. Chung,et al.  Safety and efficacy of radiofrequency energy catheter ablation of atrial fibrillation in patients with pacemakers and implantable cardiac defibrillators. , 2005, Heart rhythm.

[34]  D. Bluemke,et al.  Safety of Magnetic Resonance Imaging in Patients With Cardiovascular Devices: An American Heart Association Scientific Statement From the Committee on Diagnostic and Interventional Cardiac Catheterization, Council on Clinical Cardiology, and the Council on Cardiovascular Radiology and Intervention , 2007, Circulation.

[35]  A. Gami,et al.  Randomized Controlled Trial of Perioperative ICD Management: Magnet Application versus Reprogramming , 2014, Pacing and clinical electrophysiology : PACE.

[36]  Mehmet Ileri,et al.  The effects of mobile phones on pacemaker function. , 2005, International journal of cardiology.

[37]  Nora Goldschlager,et al.  Electromagnetic Interference and Implanted Cardiac Devices: The Medical Environment (Part II) , 2012, Clinical cardiology.

[38]  C. Hurkmans,et al.  Influence of radiotherapy on the latest generation of pacemakers. , 2005, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[39]  J. Stockman,et al.  A Prospective Evaluation of a Protocol for Magnetic Resonance Imaging of Patients With Implanted Cardiac Devices , 2013 .

[40]  Lluís Mont,et al.  2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). , 2013, European heart journal.

[41]  Eraldo Occhetta,et al.  Inappropriate implantable cardioverter-defibrillator discharges unrelated to supraventricular tachyarrhythmias. , 2006, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[42]  H. Mond,et al.  The 11th World Survey of Cardiac Pacing and Implantable Cardioverter‐Defibrillators: Calendar Year 2009–A World Society of Arrhythmia's Project , 2011, Pacing and clinical electrophysiology : PACE.

[43]  S. Laniado,et al.  The Effect of Metal Detector Gates on Implanted Permanent Pacemakers , 1988, Pacing and clinical electrophysiology : PACE.