Powered Stapler Malfunction During Laparoscopic Nephrectomy

Powered laparoscopic staplers have entered the market in the last few years. We report the first case of a malfunction of a powered laparoscopic stapler during a laparoscopic nephrectomy for a nonfunctioning kidney. The complication was salvaged with further laparoscopic maneuvers by the surgeon without conversion or any further intraoperative complication. Powered laparoscopic staplers, like the nonpowered versions, are equally prone to device malfunction despite manual safety override mechanisms. Laparoscopic surgeons using these devices must be aware of this rare complication and possess the necessary skills to troubleshoot and overcome the problem without endangering the patient.

[1]  A. ElGeidie,et al.  Stapler's malfunction during laparoscopic sleeve gastrectomy: an unusual but correctable complication. , 2013, Surgery for obesity and related diseases : official journal of the American Society for Bariatric Surgery.

[2]  K. Stahlfeld,et al.  The unacknowledged incidence of laparoscopic stapler malfunction , 2012, Surgical Endoscopy.

[3]  R. Hsi,et al.  Analysis of techniques to secure the renal hilum during laparoscopic donor nephrectomy: review of the FDA database. , 2009, Urology.

[4]  U. Boggi,et al.  Robotic suture of a large caval injury caused by endo-GIA stapler malfunction during laparoscopic wedge resection of liver segments VII and VIII en-bloc with the right hepatic vein , 2009, Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy.

[5]  T. Tzai,et al.  En bloc stapling of renal hilum during hand-assisted retroperitoneoscopic nephroureterectomy in dialysis patients. , 2008, Urology.

[6]  S. Ravichandran,et al.  EN BLOC STAPLER LIGATION OF THE RENAL VASCULAR PEDICLE DURING LAPAROSCOPIC NEPHRECTOMY , 2008, BJU international.

[7]  E. Wallen,et al.  Efficacy and safety of en bloc ligation of renal hilum during laparoscopic nephrectomy. , 2005, Urology.

[8]  A. Shalhav,et al.  En bloc stapling of renal hilum during laparoscopic nephrectomy and nephroureterectomy. , 2004, Urology.

[9]  M. Meng,et al.  Laparoscopic linear cutting stapler failure. , 2002, Urology.