Fiducial-based targeting accuracy for external-beam radiotherapy.

The accuracy of fiducial-based alignment of external radiotherapy beams is analyzed. The study considers three basic computational methods to determine the target position--the exact closed-form solution for three fiducials, the solution via singular value decomposition for four or more fiducials, and the iterative solution for any number of fiducials--and assesses their accuracy, robustness, and efficiency. Particular attention is paid to inaccuracies arising from the variability of fiducial positions in soft tissue. In nearly every test case it is found that all three solution methods, when properly implemented, yield the same result for the target position, but that the method of singular value decomposition must be modified to distinguish rotations from reflections. When an accurate measure of the rotation of the target site is needed, four fiducials give much better results than three, while more than five fiducials gain little further improvement.

[1]  H M Sandler,et al.  Localization of the prostatic apex for radiation therapy using implanted markers. , 1993, International journal of radiation oncology, biology, physics.

[2]  Kenichi Kanatani,et al.  Analysis of 3-D Rotation Fitting , 1994, IEEE Trans. Pattern Anal. Mach. Intell..

[3]  S. Umeyama,et al.  Least-Squares Estimation of Transformation Parameters Between Two Point Patterns , 1991, IEEE Trans. Pattern Anal. Mach. Intell..

[4]  P. Schönemann,et al.  A generalized solution of the orthogonal procrustes problem , 1966 .

[5]  M. Murphy An automatic six-degree-of-freedom image registration algorithm for image-guided frameless stereotaxic radiosurgery. , 1997, Medical physics.

[6]  Robert B. Fisher,et al.  Estimating 3-D rigid body transformations: a comparison of four major algorithms , 1997, Machine Vision and Applications.

[7]  K. Forster,et al.  Image-guided radiosurgery for the spine and pancreas. , 2000, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[8]  P Hofman,et al.  Detection of fiducial gold markers for automatic on-line megavoltage position verification using a marker extraction kernel (MEK). , 2000, International journal of radiation oncology, biology, physics.

[9]  J Szanto,et al.  Respiratory-induced prostate motion: quantification and characterization. , 2000, International journal of radiation oncology, biology, physics.

[10]  J. Crook,et al.  Prostate motion during standard radiotherapy as assessed by fiducial markers. , 1995, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[11]  L. Verhey,et al.  Computer-assisted positioning of radiotherapy patients using implanted radiopaque fiducials. , 1993, Medical physics.

[12]  S. Chandrasekhar Stochastic problems in Physics and Astronomy , 1943 .

[13]  K Lam,et al.  Measurement of prostate movement over the course of routine radiotherapy using implanted markers. , 1995, International journal of radiation oncology, biology, physics.