Comparison Between Pre-Log and Post-Log Statistical Models in Ultra-Low-Dose CT Reconstruction

X-ray detectors in clinical computed tomography (CT) usually operate in current-integrating mode. Their complicated signal statistics often lead to intractable likelihood functions for practical use in model-based image reconstruction (MBIR). It is therefore desirable to design simplified statistical models without losing the essential factors. Depending on whether the CT transmission data are logarithmically transformed, pre-log and post-log models are two major categories of choices in CT MBIR. Both being approximations, it remains an open question whether one model can notably improve image quality over the other on real scanners. In this study, we develop and compare several pre-log and post-log MBIR algorithms under a unified framework. Their reconstruction accuracy based on simulation and clinical datasets are evaluated. The results show that pre-log MBIR can achieve notably better quantitative accuracy than post-log MBIR in ultra-low-dose CT, although in less extreme cases, post-log MBIR with handcrafted pre-processing remains a competitive alternative. Pre-log MBIR could play a growing role in emerging ultra-low-dose CT applications.

[1]  Laureline Berteloot,et al.  Model-based iterative reconstruction in pediatric chest CT: assessment of image quality in a prospective study of children with cystic fibrosis , 2013, Pediatric Radiology.

[2]  Bruno De Man,et al.  Analysis of statistical models for iterative reconstruction of extremely low-dose CT data , 2014, 2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC).

[3]  Jeffrey A. Fessler,et al.  Ieee Transactions on Image Processing: to Appear Globally Convergent Algorithms for Maximum a Posteriori Transmission Tomography , 2022 .

[4]  Hakan Erdogan,et al.  Monotonic algorithms for transmission tomography , 2002, 5th IEEE EMBS International Summer School on Biomedical Imaging, 2002..

[5]  J. Fessler,et al.  Modelling the physics in the iterative reconstruction for transmission computed tomography , 2013, Physics in medicine and biology.

[6]  Johan Nuyts,et al.  Reconstruction Scheme for Accelerated Maximum Likelihood Reconstruction: The Patchwork Structure , 2014, IEEE Transactions on Nuclear Science.

[7]  Jeffrey A. Fessler,et al.  Statistical image reconstruction for polyenergetic X-ray computed tomography , 2002, IEEE Transactions on Medical Imaging.

[8]  Jie Tang,et al.  Performance comparison between total variation (TV)-based compressed sensing and statistical iterative reconstruction algorithms , 2009, Physics in medicine and biology.

[9]  Lin Fu,et al.  Comparison between pre-log and post-log statistical models in low-dose CT iterative reconstruction , 2014, 2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC).

[10]  Patrick Dupont,et al.  An iterative maximum-likelihood polychromatic algorithm for CT , 2001, IEEE Transactions on Medical Imaging.

[11]  Norbert J Pelc,et al.  A limit on dose reduction possible with CT reconstruction algorithms without prior knowledge of the scan subject. , 2016, Medical physics.

[12]  David S. Lalush,et al.  MAP-EM And WLS-MAP-CG Reconstruction Methods For Transmission Imaging In Cardiac SPECT , 1993, 1993 IEEE Conference Record Nuclear Science Symposium and Medical Imaging Conference.

[13]  Zhou Yu,et al.  Fast Model-Based X-Ray CT Reconstruction Using Spatially Nonhomogeneous ICD Optimization , 2011, IEEE Transactions on Image Processing.

[14]  C. Bouman,et al.  A recursive filter for noise reduction in statistical iterative tomographic imaging , 2006, Electronic Imaging.

[15]  H. Malcolm Hudson,et al.  Accelerated image reconstruction using ordered subsets of projection data , 1994, IEEE Trans. Medical Imaging.

[16]  Jeffrey A. Fessler,et al.  Segmentation-free statistical image reconstruction for polyenergetic X-ray computed tomography , 2002, Proceedings IEEE International Symposium on Biomedical Imaging.

[17]  B. De Man,et al.  Distance-driven projection and backprojection in three dimensions. , 2004, Physics in medicine and biology.

[18]  Predrag Sukovic,et al.  Penalized weighted least-squares image reconstruction for dual energy X-ray transmission tomography , 2000, IEEE Transactions on Medical Imaging.

[19]  Bruno De Man,et al.  Ultra-low dose CT attenuation correction for PET/CT , 2011, Physics in medicine and biology.

[20]  David H. Kim,et al.  Abdominal CT with model-based iterative reconstruction (MBIR): initial results of a prospective trial comparing ultralow-dose with standard-dose imaging. , 2012, AJR. American journal of roentgenology.

[21]  Zhengrong Liang,et al.  Variance analysis of x-ray CT sinograms in the presence of electronic noise background. , 2012, Medical physics.

[22]  Sana Boudabbous,et al.  Computed tomography of the chest with model-based iterative reconstruction using a radiation exposure similar to chest X-ray examination: preliminary observations , 2013, European Radiology.

[23]  Ken D. Sauer,et al.  A local update strategy for iterative reconstruction from projections , 1993, IEEE Trans. Signal Process..

[24]  Jing Wang,et al.  Deriving adaptive MRF coefficients from previous normal-dose CT scan for low-dose image reconstruction via penalized weighted least-squares minimization. , 2014, Medical physics.

[25]  J. Fessler Statistical Image Reconstruction Methods for Transmission Tomography , 2000 .

[26]  Zhengrong Liang,et al.  An experimental study on the noise properties of x-ray CT sinogram data in Radon space , 2008, Physics in medicine and biology.

[27]  Jeffrey A. Fessler,et al.  Combining Ordered Subsets and Momentum for Accelerated X-Ray CT Image Reconstruction , 2015, IEEE Transactions on Medical Imaging.

[28]  Ken D. Sauer,et al.  A Model-Based Image Reconstruction Algorithm With Simultaneous Beam Hardening Correction for X-Ray CT , 2015, IEEE Transactions on Computational Imaging.

[29]  Wojciech Zbijewski,et al.  Statistical reconstruction for x-ray CT systems with non-continuous detectors. , 2007, Physics in medicine and biology.

[30]  Freek J. Beekman,et al.  Accelerated iterative transmission CT reconstruction using an ordered subsets convex algorithm , 1998, IEEE Transactions on Medical Imaging.

[31]  Jing Wang,et al.  Virtual Colonoscopy Screening with Ultra Low-Dose CT and Less-Stressful Bowel Preparation: A computer simulation study. , 2008, IEEE transactions on nuclear science.

[32]  Jing Wang,et al.  Penalized weighted least-squares approach for low-dose x-ray computed tomography , 2006, SPIE Medical Imaging.

[33]  Yoshito Otake,et al.  Accelerated statistical reconstruction for C-arm cone-beam CT using Nesterov's method. , 2015, Medical physics.

[34]  Hongbing Lu,et al.  Nonlinear sinogram smoothing for low-dose X-ray CT , 2004 .

[35]  Jolyon A. Browne,et al.  Developments with maximum likelihood X-ray computed tomography , 1992, IEEE Trans. Medical Imaging.

[36]  Dieter Hahn Statistical Iterative Reconstruction for X-ray Phase-Contrast Computed Tomography , 2014 .

[37]  M. Kachelriess,et al.  Quality of statistical reconstruction in medical CT , 2003, 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515).

[38]  P. L. La Riviere Penalized-likelihood sinogram smoothing for low-dose CT. , 2005, Medical physics.

[39]  Jeffrey A. Fessler,et al.  Grouped-coordinate ascent algorithms for penalized-likelihood transmission image reconstruction , 1997, IEEE Transactions on Medical Imaging.

[40]  J.A. O'Sullivan,et al.  Alternating minimization algorithms for transmission tomography using energy detectors , 2002, Conference Record of the Thirty-Sixth Asilomar Conference on Signals, Systems and Computers, 2002..

[41]  Jean-Baptiste Thibault,et al.  A three-dimensional statistical approach to improved image quality for multislice helical CT. , 2007, Medical physics.

[42]  Patrick J. La Rivière,et al.  Reduction of noise-induced streak artifacts in X-ray computed tomography through spline-based penalized-likelihood sinogram smoothing , 2005, IEEE Transactions on Medical Imaging.

[43]  Patrick J. La Riviere,et al.  An algorithm for modeling non-linear system effects in iterative CT reconstruction , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[44]  Jeffrey A. Fessler,et al.  Efficient and accurate likelihood for iterative image reconstruction in x-ray computed tomography , 2003, SPIE Medical Imaging.

[45]  Omer Demirkaya Reduction of noise and image artifacts in computed tomography by nonlinear filtration of projection images , 2001, SPIE Medical Imaging.

[46]  L. Xing,et al.  Iterative image reconstruction for CBCT using edge-preserving prior. , 2008, Medical physics.

[47]  Alfred O. Hero,et al.  Convergent incremental optimization transfer algorithms: application to tomography , 2006, IEEE Transactions on Medical Imaging.

[48]  Jingyan Xu,et al.  Electronic Noise Modeling in Statistical Iterative Reconstruction , 2009, IEEE Transactions on Image Processing.

[49]  K. Lange,et al.  EM reconstruction algorithms for emission and transmission tomography. , 1984, Journal of computer assisted tomography.

[50]  Ken D. Sauer,et al.  Statistical x-ray computed tomography imaging from photon-starved measurements , 2013, Electronic Imaging.

[51]  Joseph A. O'Sullivan,et al.  Alternating Minimization Algorithms for Transmission Tomography , 2007, IEEE Transactions on Medical Imaging.

[52]  Murali Rao,et al.  Weighted least-squares reconstruction methods for positron emission tomography , 1997, IEEE Transactions on Medical Imaging.

[53]  Jeffrey A. Fessler,et al.  Segmentation-free statistical image reconstruction for polyenergetic x-ray computed tomography with experimental validation , 2003 .

[54]  Jeffrey A. Fessler,et al.  Penalized-likelihood estimators and noise analysis for randoms-precorrected PET transmission scans , 1999, IEEE Transactions on Medical Imaging.

[55]  C. Helstrom,et al.  Compensation for readout noise in CCD images , 1995 .

[56]  J. O’Sullivan,et al.  Properties of preprocessed sinogram data in x-ray computed tomography. , 2006, Medical physics.

[57]  Jianhua Ma,et al.  Extracting Information From Previous Full-Dose CT Scan for Knowledge-Based Bayesian Reconstruction of Current Low-Dose CT Images , 2016, IEEE Transactions on Medical Imaging.

[58]  R Proksa,et al.  Noise and resolution in images reconstructed with FBP and OSC algorithms for CT. , 2007, Medical physics.

[59]  Bruce R. Whiting,et al.  Signal statistics in x-ray computed tomography , 2002, SPIE Medical Imaging.

[60]  Patrick Dupont,et al.  Reduction of metal streak artifacts in X-ray computed tomography using a transmission maximum a posteriori algorithm , 1999 .

[61]  Jean-Baptiste Thibault,et al.  Modeling and Pre-Treatment of Photon-Starved CT Data for Iterative Reconstruction , 2017, IEEE Trans. Medical Imaging.

[62]  Jiang Hsieh,et al.  Computed Tomography: Principles, Design, Artifacts, and Recent Advances, Fourth Edition , 2022 .

[63]  Jeffrey A. Fessler,et al.  Ieee Transactions on Image Processing: to Appear Hybrid Poisson/polynomial Objective Functions for Tomographic Image Reconstruction from Transmission Scans , 2022 .

[64]  W. Kalender,et al.  Generalized multi-dimensional adaptive filtering for conventional and spiral single-slice, multi-slice, and cone-beam CT. , 2001, Medical physics.

[65]  Zhengrong Liang,et al.  Adaptive nonlocal means-regularized iterative image reconstruction for sparse-view CT , 2014, 2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC).

[66]  Lei Zhang,et al.  Low-Dose X-ray CT Reconstruction via Dictionary Learning , 2012, IEEE Transactions on Medical Imaging.

[67]  Hakan Erdogan,et al.  Ordered subsets algorithms for transmission tomography. , 1999, Physics in medicine and biology.

[68]  R. White,et al.  Image recovery from data acquired with a charge-coupled-device camera. , 1993, Journal of the Optical Society of America. A, Optics and image science.

[69]  Jeffrey A. Fessler,et al.  Multi-Material Decomposition Using Statistical Image Reconstruction for Spectral CT , 2014, IEEE Transactions on Medical Imaging.

[70]  Jingyan Xu,et al.  Quantifying the Importance of the Statistical Assumption in Statistical X-ray CT Image Reconstruction , 2014, IEEE Transactions on Medical Imaging.

[71]  Bruce R Whiting,et al.  Statistical reconstruction for x-ray computed tomography using energy-integrating detectors , 2007, Physics in medicine and biology.

[72]  Xuanqin Mou,et al.  Statistical Interior Tomography , 2010, IEEE Transactions on Medical Imaging.

[73]  Patrick J. La Rivière,et al.  Penalized-likelihood sinogram restoration for computed tomography , 2006, IEEE Transactions on Medical Imaging.

[74]  Ken D. Sauer,et al.  Model-Based Iterative Reconstruction for Dual-Energy X-Ray CT Using a Joint Quadratic Likelihood Model , 2014, IEEE Transactions on Medical Imaging.

[75]  M. Shiung,et al.  Development and Validation of a Practical Lower-Dose-Simulation Tool for Optimizing Computed Tomography Scan Protocols , 2012, Journal of computer assisted tomography.

[76]  David S. Perlmutter,et al.  Statistical comparison of likelihood models for low dose x-ray CT , 2014, 2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC).

[77]  Ken D. Sauer,et al.  A unified approach to statistical tomography using coordinate descent optimization , 1996, IEEE Trans. Image Process..

[78]  J. Hsieh Adaptive streak artifact reduction in computed tomography resulting from excessive x-ray photon noise. , 1998, Medical physics.

[79]  Joseph A O'Sullivan,et al.  Prospects for quantitative computed tomography imaging in the presence of foreign metal bodies using statistical image reconstruction. , 2002, Medical physics.

[80]  M. Defrise,et al.  Iterative reconstruction for helical CT: a simulation study. , 1998, Physics in medicine and biology.

[81]  Jiang Hsieh,et al.  A study of four minimization approaches for iterative reconstruction in X-ray CT , 2005, IEEE Nuclear Science Symposium Conference Record, 2005.