Neonatal mannequin comparison of the Upright self-inflating bag and snap-fit mask versus standard resuscitators and masks: leak, applied load and tidal volumes

Objective Neonatal mask ventilation is a difficult skill to acquire and maintain. Mask leak is common and can lead to ineffective ventilation. The aim of this study was to determine whether newly available neonatal self-inflating bags and masks could reduce mask leak without additional load being applied to the face. Design Forty operators delivered 1 min episodes of mask ventilation to a mannequin using the Laerdal Upright Resuscitator, a standard Laerdal infant resuscitator (Laerdal Medical) and a T-Piece Resuscitator (Neopuff), using both the Laerdal snap-fit face mask and the standard Laerdal size 0/1 face mask (equivalent sizes). Participants were asked to use pressure sufficient to achieve ‘appropriate’ chest rise. Leak, applied load, airway pressure and tidal volume were measured continuously. Participants were unaware that load was being recorded. Results There was no difference in mask leak between resuscitation devices. Leak was significantly lower when the snap-fit mask was used with all resuscitation devices, compared with the standard mask (14% vs 37% leak, P<0.01). The snap-fit mask was preferred by 83% of participants. The device-mask combinations had no significant effect on applied load. Conclusions The Laerdal Upright Resuscitator resulted in similar leak to the other resuscitation devices studied, and did not exert additional load to the face and head. The snap-fit mask significantly reduced overall leak with all resuscitation devices and was the mask preferred by participants.

[1]  A. Axelin,et al.  Scaling Up Safer Birth Bundle Through Quality Improvement in Nepal (SUSTAIN)—a stepped wedge cluster randomized controlled trial in public hospitals , 2019, Implementation Science.

[2]  J. Eilevstjønn,et al.  Born not breathing: A randomised trial comparing two self-inflating bag-masks during newborn resuscitation in Tanzania. , 2017, Resuscitation.

[3]  J. Eilevstjønn,et al.  Randomised comparison of two neonatal resuscitation bags in manikin ventilation , 2015, Archives of Disease in Childhood: Fetal and Neonatal Edition.

[4]  Indira Narayanan,et al.  Performance and Acceptability of Two Self-Inflating Bag-Mask Neonatal Resuscitator Designs , 2015, Respiratory Care.

[5]  W. Keenan,et al.  Helping babies breathe: Lessons learned guiding the way forward. A 5-year report from the HBB Global Development Alliance. , 2015 .

[6]  New Zealand. Maternal mortality in 1990-2013 WHO, UNICEF, UNFPA, The World Bank, and United Nations Population Division Maternal Mortality Estimation Inter-Agency Group , 2013 .

[7]  F. Walther,et al.  Compressive force applied to a manikin's head during mask ventilation , 2011, Archives of Disease in Childhood: Fetal and Neonatal Edition.

[8]  M. Hazinski,et al.  Resuscitation and Emergency Cardiovascular Care Science With Treatment , 2010 .

[9]  Kenneth Hill,et al.  Levels and trends in child mortality, 1990–2009 , 2010, The Lancet.

[10]  C. Morley,et al.  Leak and obstruction with mask ventilation during simulated neonatal resuscitation , 2010, Archives of Disease in Childhood: Fetal and Neonatal Edition.

[11]  M. Kelm,et al.  Equipment and operator training denote manual ventilation performance in neonatal resuscitation. , 2010, American journal of perinatology.

[12]  Peter G Davis,et al.  Reducing lung injury during neonatal resuscitation of preterm infants. , 2008, The Journal of pediatrics.

[13]  P. Davis,et al.  From liquid to air: breathing after birth. , 2008, The Journal of pediatrics.

[14]  P. Davis,et al.  A respiratory function monitor improves mask ventilation , 2008, Archives of Disease in Childhood Fetal and Neonatal Edition.

[15]  C. O. Kamlin,et al.  Improved techniques reduce face mask leak during simulated neonatal resuscitation: study 2 , 2007, Archives of Disease in Childhood Fetal and Neonatal Edition.

[16]  J. Pillow,et al.  Brief, large tidal volume ventilation initiates lung injury and a systemic response in fetal sheep. , 2007, American journal of respiratory and critical care medicine.

[17]  N. Finer,et al.  Comparison of methods of bag and mask ventilation for neonatal resuscitation. , 2001, Resuscitation.

[18]  K. Peevy,et al.  Chest wall restriction limits high airway pressure-induced lung injury in young rabbits. , 1989, Journal of applied physiology.

[19]  P. Soler,et al.  Intermittent positive-pressure hyperventilation with high inflation pressures produces pulmonary microvascular injury in rats. , 2015, The American review of respiratory disease.