High‐performance radiofrequency coils for 23Na MRI: brain and musculoskeletal applications

23Na RF coil design for brain and MSK applications presents a number of challenges, including poor coil loading for arrays of small coils and SNR penalties associated with providing 1H capability with the same coil. The basics of RF coil design are described, as well as a review of historical approaches to dual tuning. There follows a review of published high performance coil designs for MSK and brain imaging. Several coil designs have been demonstrated at 7T and 3T which incorporate close‐fitting receive arrays and in some cases design features which provide 1H imaging with little penalty to 23Na sensitivity. The “nested coplanar loop” approach is examined, in which small transmit‐receive 1H elements are placed within each 23Na loop, presenting only a small perturbation to 23Na performance and minimizing RF shielding issues. Other designs incorporating transmit‐receive arrays for 23Na and 1H are discussed including a 9.4 T 23Na / 1H brain coil. Great gains in 23Na SNR have been made with many of these designs, but simultaneously achieving high performance for 1H remains elusive. Copyright © 2015 John Wiley & Sons, Ltd.

[1]  G B Matson,et al.  A practical double‐tuned 1H/31P quadrature birdcage headcoil optimized for 31P operation , 1999, Magnetic resonance in medicine.

[2]  Nicholas R Zwart,et al.  A new design and rationale for 3D orthogonally oversampled k‐space trajectories , 2011, Magnetic resonance in medicine.

[3]  P. Roemer,et al.  The NMR phased array , 1990, Magnetic resonance in medicine.

[4]  Jullie W Pan,et al.  High frequency volume coils for clinical NMR imaging and spectroscopy , 1994, Magnetic resonance in medicine.

[5]  C Hawryszko,et al.  Design and evaluation of a novel dual-tuned resonator for spectroscopic imaging , 1990 .

[6]  J. Fitzsimmons,et al.  Double resonant quadrature birdcage , 1993, Magnetic resonance in medicine.

[7]  Britton Chance,et al.  A new double-tuned probed for concurrent 1H and 31P NMR☆ , 1985 .

[8]  Vladimir Juras,et al.  Sodium Magnetic Resonance Imaging of Ankle Joint in Cadaver Specimens, Volunteers, and Patients After Different Cartilage Repair Techniques at 7 T: Initial Results , 2015, Investigative radiology.

[9]  Tiejun Zhao,et al.  Multichannel transceiver dual-tuned RF coil for proton/sodium MR imaging of knee cartilage at 3 T. , 2012, Magnetic resonance imaging.

[10]  C N Chen,et al.  Quadrature detection in the laboratory frame , 1984, Magnetic resonance in medicine.

[11]  T R Brown,et al.  Two configurations of the four-ring birdcage coil for 1H imaging and 1H-decoupled 31P spectroscopy of the human head. , 1994, Journal of magnetic resonance. Series B.

[12]  D. Sodickson,et al.  Ideal current patterns yielding optimal signal‐to‐noise ratio and specific absorption rate in magnetic resonance imaging: Computational methods and physical insights , 2012, Magnetic resonance in medicine.

[13]  J. Schenck,et al.  An efficient, highly homogeneous radiofrequency coil for whole-body NMR imaging at 1.5 T , 1985 .

[14]  M Alecci,et al.  Practical design of a 4 Tesla double-tuned RF surface coil for interleaved 1H and 23Na MRI of rat brain. , 2006, Journal of magnetic resonance.

[15]  D. W. Alderman,et al.  An efficient decoupler coil design which reduces heating in conductive samples in superconducting spectrometers , 1979 .

[16]  Robert E. Lenkinski,et al.  A design for a double-tuned birdcage coil for use in an integrated MRI/MRS examination , 1990 .

[17]  Tiejun Zhao,et al.  Quantitative 23Na MRI of human knee cartilage using dual‐tuned 1H/23Na transceiver array radiofrequency coil at 7 tesla , 2013, Journal of magnetic resonance imaging : JMRI.

[18]  J T Vaughan,et al.  Evaluation of 31P metabolite differences in human cerebral gray and white matter , 1998, Magnetic resonance in medicine.

[19]  Riccardo Lattanzi,et al.  The Electric Dipole Array : An Attempt to Match the Ideal Current Pattern for Central SNR at 7 Tesla , 2011 .

[20]  R S Balaban,et al.  Radiofrequency shielding of surface coils at 4.0 t , 1995, Journal of magnetic resonance imaging : JMRI.

[21]  D. Twieg,et al.  Is the intracellular pH different from normal in the epileptic focus of patients with temporal lobe epilepsy? , 1996, Neurology.

[22]  Boguslaw Tomanek,et al.  Double-frequency birdcage volume coils for 4.7T and 7T , 2005 .

[23]  Daniel Sodickson,et al.  4572 A 7 T 8 channel transmit-receive dipole array for head imaging : dipole element and coil evaluation , 2013 .

[24]  R. Brown,et al.  A Nested Dual Frequency Birdcage / Stripline Coil for Sodium / Proton Brain Imaging at 7 T , 2009 .

[25]  J R Fitzsimmons,et al.  A comparison of double‐tuned surface coils , 1989, Magnetic resonance in medicine.

[26]  Ravinder R Regatte,et al.  Design of a nested eight‐channel sodium and four‐channel proton coil for 7T knee imaging , 2013, Magnetic resonance in medicine.

[27]  W. Barber,et al.  Comparison of linear and circular polarization for magnetic resonance imaging , 1985 .

[28]  K. Uğurbil,et al.  B1-insensitive heteronuclear adiabatic polarization transfer for signal enhancement☆ , 1992 .

[29]  C A T van den Berg,et al.  Design of a radiative surface coil array element at 7 T: The single‐side adapted dipole antenna , 2011, Magnetic resonance in medicine.

[30]  Jens Bornemann,et al.  Multiple tuning of birdcage resonators , 1997, Magnetic resonance in medicine.

[31]  D. Sodickson,et al.  SNR Degradation in Receive Arrays Due to Preamplifier Noise Coupling and A Method for Mitigation , 2011 .

[32]  L. Wald,et al.  Theory and application of array coils in MR spectroscopy , 1997, NMR in biomedicine.

[33]  K. Scheffler,et al.  Combination of a sodium birdcage coil with a tunable patch antenna for B0 shimming and anatomical localization at 9.4 T , 2013 .

[34]  Armin M Nagel,et al.  Evaluation of adaptive combination of 30‐channel head receive coil array data in 23Na MR imaging , 2016, Magnetic resonance in medicine.

[35]  Fernando Boada,et al.  224 An 8 Channel Transmit Receive Sodium & Nested 8 Channel Transmit Receive Proton Coil for 3 . 0 T Brain Imaging , 2013 .

[36]  C J Hardy,et al.  Proton‐decoupled, overhauser‐enhanced, spatially localized carbon‐13 spectroscopy in humans , 1989, Magnetic resonance in medicine.

[37]  Peter Kellman,et al.  Image reconstruction in SNR units: A general method for SNR measurement † , 2005, Magnetic resonance in medicine.

[38]  Klaus Scheffler,et al.  Three‐layered radio frequency coil arrangement for sodium MRI of the human brain at 9.4 Tesla , 2016, Magnetic resonance in medicine.

[39]  Aiming Lu,et al.  Quantitative sodium imaging with a flexible twisted projection pulse sequence , 2010, Magnetic resonance in medicine.

[40]  P M Joseph,et al.  A technique for double resonant operation of birdcage imaging coils. , 1989, IEEE transactions on medical imaging.

[41]  Lawrence L Wald,et al.  Sodium imaging of human brain at 7 T with 15‐channel array coil , 2012, Magnetic resonance in medicine.

[42]  Alan R. Rath Design and performance of a double-tuned bird-cage coil , 1990 .

[43]  L Axel,et al.  Noise performance of surface coils for magnetic resonance imaging at 1.5 T. , 1985, Medical physics.

[44]  R. Kuzniecky,et al.  Lateralization of human temporal lobe epilepsy by 31P NMR spectroscopic imaging at 4.1 T , 1998, Neurology.

[45]  K. Thulborn,et al.  Dual‐frequency, dual‐quadrature, birdcage RF coil design with identical b1 pattern for sodium and proton imaging of the human brain at 1.5 T , 1997, Magnetic resonance in medicine.

[46]  G. Adriany,et al.  A half-volume coil for efficient proton decoupling in humans at 4 tesla. , 1997, Journal of magnetic resonance.

[47]  P. Boesiger,et al.  Advances in sensitivity encoding with arbitrary k‐space trajectories , 2001, Magnetic resonance in medicine.

[48]  Xiaoyu Yang,et al.  A 3D Parallel Imaging Capable Transmit and 15-Channel Receive Array Knee Coil at 3T. , 2008, Proceedings of the International Society for Magnetic Resonance in Medicine ... Scientific Meeting and Exhibition. International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition.

[49]  Klaus Scheffler,et al.  High‐resolution quantitative sodium imaging at 9.4 tesla , 2015, Magnetic resonance in medicine.

[50]  Mary P. McDougall,et al.  Trap design for insertable second-nuclei radiofrequency coils for magnetic resonance imaging and spectroscopy , 2009 .