Surface Damage on Open Box Posterior-Stabilized Polyethylene Tibial Inserts

Twenty retrieved Scorpio® posterior-stabilized implants were available for analysis. The mean implantation time was 22 months (range, 2 days–42 months). Favorable types and amounts of surface damage were seen on the tibiofemoral and backsides of these modular PE liners that had been packaged in an inert environment and then sterilized by gamma irradiation. Delamination represented only 0.1% of the total surface damage. Off-axis loading (varus malalignment) was associated with tibial component loosening but there was no evidence of peripheral damage of PE caused by edge loading. With this open-box design, hyperextension marks on the anterior aspect of the posterior-stabilized post from femoral component impingement occurred in 11 of 20 cases and was related to sagittal component positioning: excess tibial slope or increased tibial slope combined with a flexed femoral component. Unique coarse abrasions occurred in 16 of 20 cases and were the result of cement extrusion into the open box, especially with varus malalignment. These observations provide guidance for optimizing the surgical technique and the design of posterior-stabilized total knee components.

[1]  John J. Callaghan,et al.  Current Concepts Review - Wear in Total Hip and Knee Replacements* , 1999 .

[2]  O. Mahoney,et al.  Loosening and Osteolysis with the Press-Fit Condylar Posterior-Cruciate-Substituting Total Knee Replacement , 2001, The Journal of bone and joint surgery. American volume.

[3]  G. Engh,et al.  The Coventry Award. Modular tibial insert micromotion. A concern with contemporary knee implants. , 1998 .

[4]  G. Engh,et al.  In Vivo Deterioration of Tibial Baseplate Locking Mechanisms in Contemporary Modular Total Knee Components , 2001, The Journal of bone and joint surgery. American volume.

[5]  J J Callaghan,et al.  Wear in total hip and knee replacements. , 2000, The Journal of bone and joint surgery. American volume.

[6]  K. Saum,et al.  Impact of gamma sterilization on clinical performance of polyethylene in the knee. , 1996, The Journal of arthroplasty.

[7]  O. Mahoney,et al.  The effect of total knee arthroplasty design on extensor mechanism function. , 2002, The Journal of arthroplasty.

[8]  F. C. Ewald The Knee Society total knee arthroplasty roentgenographic evaluation and scoring system. , 1989, Clinical orthopaedics and related research.

[9]  R B Bourne,et al.  Tibial Post Wear in Posterior Stabilized Total Knee Arthroplasty: An Unrecognized Source of Polyethylene Debris , 2001, The Journal of bone and joint surgery. American volume.

[10]  A Sarmiento,et al.  The origin of submicron polyethylene wear debris in total hip arthroplasty. , 1995, Clinical orthopaedics and related research.

[11]  T. Andriacchi,et al.  Surface Damage in Machined Ram-Extruded and Net-Shape Molded Retrieved Polyethylene Tibial Inserts of Total Knee Replacements , 2002, The Journal of bone and joint surgery. American volume.

[12]  P. Walker,et al.  The conflicting requirements of laxity and conformity in total knee replacement. , 1999, Journal of biomechanics.