Design of High Efficiency Differential Electromagnetic Type Transducer for Implantable Middle Ear System

The differential electromagnetic transducer for IME(implantable middle ear) system, which have two small permanent magnets glued with the same pole facing each other in the coil, has high vibration efficiency and does not influenced by external magnetic field. In this paper, using finite element method, highly efficient structure of the transducer was proposed and vibration force of the transducer was calculated by electromagnetic theory. And the necessary vibration force of transducer to transmit the sound signal to inner ear when the transducer is attached at stapes was calculated and the design parameters of the transducer were investigated. Using the parameters, the differential electromagnetic transducer was manufactured in small size to implant in confined human middle ear. And it was examined by unloaded and loaded vibration experiment using temporal bone sampled from cadaver.

[1]  T. G. Dietz,et al.  Partially implantable vibrating ossicular prosthesis , 1997, Proceedings of International Solid State Sensors and Actuators Conference (Transducers '97).

[2]  Y. Onchi Mechanism of the Middle Ear , 1961 .

[3]  Jozef J. Zwislocki,et al.  Analysis of the Middle‐Ear Function. Part I: Input Impedance , 1962 .

[4]  M Chasin,et al.  Current Trends in Implantable Hearing Aids , 1997, Trends in amplification.