Switching Monopolar Radiofrequency Ablation Technique Using Multiple, Internally Cooled Electrodes and a Multichannel Generator: Ex Vivo and In Vivo Pilot Study

Objective:We sought to determine the optimal switching time and interprobe distance for creating a large coagulation zone in the liver by switching monopolar radiofrequency ablation (RFA) technique using a prototype multichannel radiofrequency generator and multiple electrodes. Materials and Methods:Using 3, 17-gauge, internally cooled electrodes and a prototype multichannel radiofrequency (RF) generator to allow automatic switching of RF energy among 3 electrodes according to their impedance changes, RF energy was applied in a switching monopolar mode to explanted bovine livers or to the livers of 7 dogs. In the ex vivo study, we evaluated the effect of the switching time (5, 15, 30, and 60 seconds) and the interprobe distance (3 cm and 4 cm) on the mean volume of the ablation zone using the switching RFA; we then compared the ablation volume of the switching RFA group (12 minutes) with that of the overlapping RFA group (12 minutes × 3 = 36 minutes). In our preliminary in vivo experiments using 2 dogs, the electrodes were placed in a triangular array (spacing 2 cm, 3 cm, 3.5 cm, or 4 cm) and were activated for 12 minutes or 20 minutes in a monopolar mode (power output 200 W). In our primary experiments using 5 dogs, RFA in a switching mode created 10 coagulation zones at a 2- to 3-cm interprobe distance. The duration of the RFA was 12 minutes. The size and geometry of the coagulation zone also were assessed. Results:In this ex vivo study, switching monopolar RFA at a 3-cm interprobe distance and at a 30-second switching time allowed the creation of a large, confluent ablation zone. There was no significant difference in the mean ablation volume between the overlapping (72.9 ± 12.6 cm3) and the switching RFA groups (65.7 ± 12.6 cm3; P > 0.05). In the preliminary in vivo experiments, RFA created spherical coagulations at interprobe distances of 2 cm and 3 cm but at larger distances than 3 cm, RFA created partially confluent coagulation zones. In our principal experiments, switching RFA created areas of well-defined coagulation, ie, at a 2-cm interprobe distance, the volume and short-axis coagulation diameter were 35.5± 5.7 cm3 and 4.6 ± 0.5 cm, respectively, whereas at 3 cm, they were 40.7 ± 12.8 cm3 and 4.8 ± 0.8 cm, respectively. The mean values of the circularity (isometric ratio) of the coagulation at the 2- and 3-cm interprobe distances were 0.95 ± 0.02, and 0.85± 0.06, respectively. Conclusions:Our study demonstrated that switching monopolar RFA using the multichannel RF system at a 2- or 3-cm interprobe distance and at a 30-second switching time can create a large, confluent coagulation zone in the liver within a clinically acceptable time frame. We believe that this technology will provide a useful tool for the treatment of large liver tumors.

[1]  G. Dodd,et al.  Radiofrequency thermal ablation: computer analysis of the size of the thermal injury created by overlapping ablations. , 2001, AJR. American journal of roentgenology.

[2]  R. Paczynski,et al.  Automated measurement of infarct size with scanned images of triphenyltetrazolium chloride-stained rat brains. , 1996, Stroke.

[3]  A. Gillams Radiofrequency ablation in the management of liver tumours. , 2003, European journal of surgical oncology : the journal of the European Society of Surgical Oncology and the British Association of Surgical Oncology.

[4]  S Nahum Goldberg,et al.  Microwave ablation: results with a 2.45-GHz applicator in ex vivo bovine and in vivo porcine liver. , 2006, Radiology.

[5]  Christopher L Brace,et al.  Multiple-electrode radiofrequency ablation creates confluent areas of necrosis: in vivo porcine liver results. , 2006, Radiology.

[6]  Dieter Haemmerich,et al.  Multiple probe radiofrequency ablation: pilot study in an animal model. , 2003, Journal of vascular and interventional radiology : JVIR.

[7]  John G. Webster,et al.  Finite-element analysis of hepatic multiple probe radio-frequency ablation , 2002, IEEE Transactions on Biomedical Engineering.

[8]  S. Mulier,et al.  A review of the general aspects of radiofrequency ablation , 2005, Abdominal Imaging.

[9]  D E Dupuy,et al.  Image-guided radiofrequency tumor ablation: challenges and opportunities--part II. , 2001, Journal of vascular and interventional radiology : JVIR.

[10]  Jeong Won Lee,et al.  The Potential Contribution of a Computer-Aided Detection System for Lung Nodule Detection in Multidetector Row Computed Tomography , 2004, Investigative radiology.

[11]  G. Gazelle,et al.  Hepatocellular carcinoma: radio-frequency ablation of medium and large lesions. , 2000, Radiology.

[12]  L Solbiati,et al.  Large-volume tissue ablation with radio frequency by using a clustered, internally cooled electrode technique: laboratory and clinical experience in liver metastases. , 1998, Radiology.

[13]  D. Haemmerich,et al.  Multiple applicator approaches for radiofrequency and microwave ablation , 2005, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[14]  T. Winter,et al.  Effect of vascular occlusion on radiofrequency ablation of the liver: results in a porcine model. , 2001, AJR. American journal of roentgenology.

[15]  Philippe L Pereira,et al.  Multipolar radiofrequency ablation with internally cooled electrodes: experimental study in ex vivo bovine liver with mathematic modeling. , 2006, Radiology.

[16]  G. Bachar,et al.  Radiofrequency ablation for the management of liver tumors. , 2003, The Israel Medical Association journal : IMAJ.

[17]  Thomas J Vogl,et al.  Image-guided tumor ablation: proposal for standardization of terms and reporting criteria. , 2003, Radiology.

[18]  R W Günther,et al.  Multipolar radiofrequency ablation: first clinical results. , 2004, RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin.

[19]  K. Tranberg Percutaneous ablation of liver tumours. , 2004, Best practice & research. Clinical gastroenterology.

[20]  G. Dodd,et al.  Radiofrequency thermal ablation of liver tumors , 1999, Journal of clinical ultrasound : JCU.

[21]  E F Halpern,et al.  Percutaneous radio-frequency ablation of hepatic metastases from colorectal cancer: long-term results in 117 patients. , 2001, Radiology.

[22]  A. Siriwardena,et al.  Radiofrequency ablation of liver tumours: systematic review. , 2004, The Lancet Oncology.

[23]  J. Webster,et al.  Large-volume radiofrequency ablation of ex vivo bovine liver with multiple cooled cluster electrodes. , 2005, Radiology.

[24]  Philippe L Pereira,et al.  Radiofrequency ablation: in vivo comparison of four commercially available devices in pig livers. , 2004, Radiology.

[25]  Jae Young Lee,et al.  Hepatic Radiofrequency Ablation Using Multiple Probes: Ex Vivo and In Vivo Comparative Studies of Monopolar versus Multipolar Modes , 2006, Korean journal of radiology.

[26]  Clarisse Dromain,et al.  Usefulness of Guiding Needles for Radiofrequency Ablative Treatment of Liver Tumors , 2005, CardioVascular and Interventional Radiology.

[27]  G. Gazelle,et al.  Thermal ablation therapy for focal malignancy: a unified approach to underlying principles, techniques, and diagnostic imaging guidance. , 2000, AJR. American journal of roentgenology.

[28]  B. Choi,et al.  Dual-Probe Radiofrequency Ablation: An In Vitro Experimental Study in Bovine Liver , 2004, Investigative radiology.

[29]  G. Gazelle,et al.  Variables affecting proper system grounding for radiofrequency ablation in an animal model. , 2000, Journal of vascular and interventional radiology : JVIR.

[30]  D. Haemmerich,et al.  Multiple-electrode radiofrequency ablation: simultaneous production of separate zones of coagulation in an in vivo porcine liver model. , 2005, Journal of vascular and interventional radiology : JVIR.

[31]  D E Dupuy,et al.  Image-guided radiofrequency tumor ablation: challenges and opportunities--part I. , 2001, Journal of vascular and interventional radiology : JVIR.

[32]  B. Choi,et al.  Ex Vivo Experiment of Saline-Enhanced Hepatic Bipolar Radiofrequency Ablation with a Perfused Needle Electrode: Comparison with Conventional Monopolar and Simultaneous Monopolar Modes , 2005, CardioVascular and Interventional Radiology.

[33]  B. Choi,et al.  A Comparative Experimental Study of the In-vitro Efficiency of Hypertonic Saline-Enhanced Hepatic Bipolar and Monopolar Radiofrequency Ablation , 2003, Korean journal of radiology.

[34]  Jae Young Lee,et al.  Radiofrequency ablation of the porcine liver in vivo: increased coagulation with an internally cooled perfusion electrode. , 2006, Academic radiology.

[35]  Wei Yang,et al.  Large liver tumors: protocol for radiofrequency ablation and its clinical application in 110 patients--mathematic model, overlapping mode, and electrode placement process. , 2004, Radiology.

[36]  Steven A. Curley,et al.  Radiofrequency Ablation of Hepatocellular Cancer in 110 Patients With Cirrhosis , 2000, Annals of surgery.

[37]  Thomas Albrecht,et al.  Multipolar radiofrequency ablation of hepatic tumors: initial experience. , 2005, Radiology.

[38]  P R Mueller,et al.  Treatment of intrahepatic malignancy with radiofrequency ablation , 2000, Cancer.

[39]  T. Helmberger,et al.  Effects of Vascular Perfusion on Coagulation Size in Radiofrequency Ablation of Ex Vivo Perfused Bovine Livers , 2006, Investigative radiology.